EnerMech Pty Ltd

Site

Calibration Laboratory

Accreditation No.

18347

Site No.

22708

Date of Accreditation

Address

931 Cockburn Road
Henderson, WA 6166
Australia

enermech.com

Contact

Mr M Sarubin P: +61 (08) 65556560

Availability

Services available to external clients

Scope of Accreditation

Withdrawn

Calibration Laboratory

ISO/IEC 17025 (2005)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Withdrawn
Service Product Determinant Technique Procedure Limitations
In-situ calibration of pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure To be determined AS 1349 and MSA Test Method 2 (Pressure and vacuum gauges including test gauges)
MSA Test Method 1 (Self indicating pressure transducers and electrical transducers with electrical output)
In-house method ENM-AUS-QHSE-WI-601 (Mechanical chart recorders)

On-Site calibration (in Environmentally Controlled Mobile Container Only)

-95 kPa to 280 MPa

Capability

with Calibration and Measurement Capability of -

0.02 kPa from -95 kPa to -1.3 kPa;

0.02 kPa from 1.3 kPa to 100 kPa;

0.02% of reading from 100 kPa to 300 kPa;

0.05% of reading from 300 kPa to 280 MPa;

Pressure metrology - Pressure and vacuum measuring equipmentincluding on-site calibrations Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure By pressure calibrator MSA Test Method 2 (Pressure and vacuum gauges including test gauges)
MSA Test Method 1
In-house method ENM-AUS-QHSE-WI-601 (Mechanical chart recorders)

-95 kPa to 280 MPa

Capability

with Calibration and Measurement Capability of -

0.02 kPa from -95 kPa to -1.3 kPa;

0.02 kPa from 1.3 kPa to 100 kPa;

0.02% of reading from 100 kPa to 300 kPa;

0.05% of reading from 300 kPa to 280 MPa;

in-situ calibrations

with Calibration and Measurement Capability of -

0.02 kPa from -95 kPa to -1.3 kPa;

0.02 kPa from 1.3 kPa to 100 kPa;

0.02% of reading from 100 kPa to 300 kPa;

0.05% of reading from 300 kPa to 280 MPa;


Temperature metrology - Temperature measuring equipmentincluding on-site calibrations Bimetallic systems; Digital temperature measuring systems; Filled metal systems Temperature Measurement against reference standard

Calibration in the laboratory
From -20ºC to 500ºC 


Capability

with Calibration and Measurement Capability of -

Bimetallic systems and filled metal systems

1ºC from -20ºC to 500ºC

Digital temperature indicator systems

0.2ºC from -20ºC to 200ºC
0.3ºC above 200ºC to 300ºC
0.4ºC above 300ºC to 450ºC
0.5ºC above 450ºC to 500ºC

Torque - Torque measuring and testing equipmentincluding on-site calibrations Torque wrenches Torque Comparison with a reference standard  ISO 6789

5 Nm to 1 500 Nm

Capability

with Calibration and Measurement Capability of -

1.0% of reading from 5 Nm to 1 500 Nm

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.