Hinco Instruments Pty Ltd

Site

Hinco Instruments Pty Ltd

Accreditation No.

546

Site No.

539

Date of Accreditation

Address

Unit 5/52 Vinnicombe Drive
Canning Vale, WA 6155
Australia

hinco.com.au

Contact

Mr John French P: +61 (08) 94556777

Availability

Services available to external clients

Scope of Accreditation

Withdrawn

ISO/IEC 17025 (2005)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Withdrawn
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators DC voltage

0.01 V to 100 V
Process calibrators only

Capability

0.05% reading  

DC current

0.1 mA to 100 mA
Process calibrators only

Capability

0.05% reading from 0.1 to 100 mA

DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters AC current

10 mA to 100 A

Capability

0.1% reading from 10 mA to 20 A at 50 and 60 Hz
0.5% reading from 20 A to 1000 A for Toroidal Current Clamps [using multiple-turn technique]
1.0% reading from 20 A to 1000 A for Non-Toroidal Current Clamps [using multiple-turn technique]

Ohm meters Resistance

1Ω to 1 GΩ

Capability

0.5% of reading from 1Ω to 100 kΩ
2% of reading above 100 kΩ to 10 MΩ
5% reading above 10 MΩ to 100 MΩ
10%reading above 100 MΩ to 1 GΩ

Voltmeters DC voltage

0.01 V to 1000 V

Capability

0.05% of reading

AC voltage

0.5 V to 700 V at 50 and 60 Hz

Capability

0.1% of reading

In-situ calibration of pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gaugesNULL Gauge pressure To be determined AS 1349 (pressure and vacuum gauges)
MSA Test Method 2 (pressure and vacuum gauges)
MSA Test Method 1 (pressure transducers, pressure recorders and industrial grade, self indicating and electrical transducers with dc voltage, dc current or resistance outputs)

-95 kPa to 100 MPa

Capability

With Calibration and Measurement Capability of-

0.2 kPa from -95 to 0 kPa;

0.05% of reading or 0.3 kPa (whichever is greater) from 0 kPa up to 4 MPa;

0.05% of reading or 3 kPa (whichever is greater) from 4 to 100 MPa

Manometers Gauge pressure To be determined

MSA Test Method 1

-20 kPa to 20 kPa

Capability

0.02 kPa from -20 kPa to 20 kPa

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Withdrawn
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators DC current; DC voltage; Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

Process calibrators only
DC current
0.05% reading from 0.1 mA to 100 mA  
DC voltage
0.05% reading from 0.01 V to 100 V
Resistance

0.1% reading ± 1 mΩ from 0.1 Ω to 10 kΩ 


DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Current clamps; Ohm meters; Voltmeters AC current; AC voltage; DC current; DC voltage; Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

AC current
0.1% reading from 10 mA to 20 A at 50 and 60 Hz
0.5% reading from 20 A to 1000 A for Toroidal Current Clamps [using multiple-turn technique]
1.0% reading from 20 A to 1000 A for Non-Toroidal Current Clamps [using multiple-turn technique]
DC current

0.05% of reading from 0.1 mA to 20 A
0.9% of reading from 20 A to 1000 A for Current Clamps [using multiple-turn technique]

AC voltage
0.1% of reading  from 0.5 V to 700 V at 50 Hz and 60 Hz
DC voltage
0.05% of reading from 0.01 V to 1000 V
Resistance
0.5% of reading from 1 Ω to 100 kΩ
2% of reading above 100 kΩ to 10 MΩ
5% of reading above 10 MΩ to 100 MΩ
10% of reading above 100 MΩ to 1 GΩ
DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

0.5% of reading + 1 mΩ from 0.001 Ω to 0.1 Ω
0.1% of reading + 1 mΩ from 0.1 Ω to 10 kΩ
0.5% of reading from 10 kΩ  to 100 kΩ
2% of reading from 100 kΩ to 10 MΩ

Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations Manometers Gauge pressure By pressure calibrator

MSA Test Method 1

Capability

Including on site calibrations

with Calibration and Measurement Capability of-

0.02 kPa from -20 kPa to 20 kPa

Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure By pressure calibrator

AS 1349 (pressure and vacuum gauges)
MSA Test Method 2 (pressure and vacuum gauges)
MSA Test Method 1 (pressure transducers, pressure recorders and industrial grade, self indicating and electrical transducers with dc voltage, dc current or resistance outputs)



Capability

Including on site calibrations

with Calibration and Measurement Capability of-

0.2 kPa from -95 to 0 kPa;

0.05% of reading or 0.3 kPa (whichever is greater) from 0 kPa up to 4 MPa;

0.05% of reading or 3 kPa (whichever is greater) from 4 MPa to 100 MPa

Temperature metrology - Ancillary temperature measuring equipment Digital voltmeters (DVM) and digital multimeters (DMM); Process calibrators Resistance to temperature conversion; Volt to temperature conversion Direct measurement by electrical input

Capability

with Calibration and Measurement Capability of -

Temperature calibrator instruments
by electrical simulation [measure and source mode]

 0.8 °C + 0.1% of Temperature for Type K thermocouples from -100 °C to +1100 °C
 0.8 °C + 0.1% of Temperature for Type T thermocouples from -100 °C to +390 °C
 0.8 °C + 0.1% of Temperature for Type N thermocouples from +100 °C to +1200 °C
 0.3 °C for RTD PT100 from -100 °C to 400 °C

Temperature metrology - Temperature measuring equipment Digital temperature measuring systems; Temperature loggers - Single and multi-channel Temperature Measurement against reference standard

Capability

with Calibration and Measurement Capability of -
0.5°C from -50°C to 0°C
0.2 °C above 0°C to 150°C
0.3 °C above 150°C to 200°C

Analogue chart recorders
0.5°C from -50°C to 200°C

Temperature metrology - Temperature standards and reference equipment Dry block calibrators Temperature Direct measurement using a reference standard

Capability

with Calibration and Measurement Capability of -

0.5 °C from 0°C to 500°C

Temperature metrology - Verification of controlled enclosuresincluding on site calibrations Furnaces; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement

Capability

including on site calibrations

with Calibration and Measurement Capability of-

1°C from 0°C to 200 °C
 2.5°C above 200°C to 400°C
 5°C above 400°C to 1100°C

Medical refrigeration equipment Spatial uniformity; Temperature Direct temperature measurement

AS 3864.2 sections 3.7 and 3.8.3

Capability

Including on site calibrations

with Calibration and Measurement Capability of-

AS 3864.2 section 3.7 - Annual tasks
0.8°C from -50°C to 0°C
0.3°C above 0°C to 10°C
AS 3864.2 section 3.8.3 - Spatial temperature distribution check
0.8°C from -50°C to 10°C

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.