National Association of Testing Authorities, Australia
Scope of Accreditation
ENERGY QUEENSLAND LIMITED
Site
Instrument Calibration Laboratory
Accreditation No.
74
Site No.
20791
Date of Accreditation
Contact
Availability
Services available to external clients
Scope of Accreditation
Withdrawn
ISO/IEC 17025 (2005)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DC and low frequency electrical metrology - Electrical measurement and test equipmentIncluding on site calibrations in-situ | Ammeters; Digital multimeters (DMM) | AC current; DC current | Comparison with a reference standard |
D.C. ammeters At 0 mA and from 1 mA to 10 A A.C. ammeters At 0 mA and from 5 mA to 10 A at 45 Hz to 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of - D.C. ammeters 0.025% + 0.05 μA + 1 digit for current range of 0 mA and 1 mA to <3.3 mA A.C. ammeters 0.2% + 1 digit for current range of 0 mA and 5 mA to <2.2 A at 45 Hz to 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital multimeters (DMM); Ohm meters | DC voltage; Resistance | Comparison with a reference standard; Direct measurement against a reference standard |
At 0 Ω and from 1 Ω to 100 MΩ Insulation testers up to 1 GΩ and output voltages up to 10 kV |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of - 0.02% + 0.015 Ω + 1 digit for resistance range of 0 Ω and 1 Ω to <330 Ω Insulation testers |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital multimeters (DMM); Voltmeters | AC voltage; DC voltage | Comparison with a reference standard |
D.C. voltage At 0 mV and from 1 mV to 1000 V A.C. voltage At 0 mV and from 1 mV to 1000 V at 45 Hz to 1 kHz |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of - D.C. voltmeters 0.01% + 3 μV + 1 digit for voltage range of 0 mV and 1 mV to <330 mV A.C. voltmeters 0.4% + 20 μV + 1 digit for voltage range of 0 mV and 1 mV to <100 mV at 45 Hz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
In-situ calibration of electrical measurement and test equipment | Ammeters | AC current; DC current | Comparison method |
D.C. ammeters At 0 mA and from 1 mA to 10 A A.C. ammeters At 0 mA and from 5 mA to 10 A at 45 Hz to 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - D.C. ammeters 0.025% + 0.05 μA + 1 digit for current range of 0 mA and 1 mA to <3.3 mA A.C. ammeters 0.2% + 1 digit for current range of 0 mA and 5 mA to <2.2 A at 45 Hz to 1 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Ohm meters | DC voltage; Resistance | Comparison method; Direct measurement against a reference standard |
At 0 Ω and from 1 Ω to 100 MΩ Insulation testers up to 1 GΩ and output voltages up to 10 kV |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.02% + 0.015 Ω + 1 digit for resistance range of 0 Ω and 1 Ω to <330 Ω
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Voltmeters | AC voltage; DC voltage | Comparison method |
D.C. voltage At 0 mV and from 1 mV to 1000 V A.C. voltage At 0 mV and from 1 mV to 1000 V at 45 Hz to 1 kHz |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - D.C. voltmeters 0.01% + 3 μV + 1 digit for voltage range of 0 mV and 1 mV to <330 mV A.C. voltmeters 0.4% + 20 μV + 1 digit for voltage range of 0 mV and 1 mV to <100 mV at 45 Hz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
In-situ calibration of frequency, time and waveform measuring equipment | Frequency meters | Frequency | To be determined |
From 10 Hz to 100 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.02% + 1 mHz + 1 digit for frequency range of 10 Hz to <10 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency, time and waveform measuring equipmentIncluding on site calibrations in-situ | Frequency meters | Frequency | Measurement against reference standard |
From 10 Hz to 100 kHz |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of - 0.02% + 1 mHz + 1 digit for frequency range of 10 Hz to <10 kHz |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.