IPAC Solutions Pty Ltd

Site

IPAC Solutions Pty Ltd

Accreditation No.

15808

Site No.

17043

Date of Accreditation

19 Mar 2009

Address

F14, 64-86 Beresford Road
Lilydale, VIC 3140
Australia

ipacsolutions.com.au

Contact

Mr Robert Thomson P: +61 (03) 97358888

Availability

Services available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Mass - Determination of mass and calibration of weighing devices Industrial weighing devices Mass Gravimetric measurement against reference mass EUROMET Calibration Guide 18
NMI Monograph 4 Section 6
IPAC Solutions 6-7-12-01 Calibration of a Weighing Instrument

Capability

Including on site calibrations
With Calibration and Measurement Capability of -
1 in 104 or 100 mg (whichever is greater) up to 300 kg
Laboratory weighing devices; Precision laboratory balances Mass Gravimetric measurement against reference mass EUROMET Calibration Guide 18
NMI Monograph 4 Section 6
IPAC Solutions 6-7-12-01 Calibration of a Weighing Instrument

Capability

Including on site calibrations
with Calibration and Measurement Capability of -
 6 μg from 1 mg up to and including 500 mg
1 in 106 or 0.01 mg (whichever is greater) above 500 mg up to and including 300 g
1 in 106 or 6 mg (whichever is greater) above 300 g up to and including 10 kg
5 in 105 or 1 mg (whichever is greater) above 10 kg up to 40 kg
Pressure metrology - Pressure and vacuum measuring equipment Manometers; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Absolute pressure Comparison with dead weight tester; Comparison with reference instrument

As defined in MSA Test Method 1 or 2, as appropriate.







Capability

with Calibration and Measurement Capability of -
Pneumatic Instruments
0.006% of reading or 0.020 kPa (whichever is greater) from 1 kPa to 3500 kPa (absolute)
Hydraulic Instruments
0.012 % of reading or 0.05 kPa (whichever is greater) from 0 to below 4 MPa
0.010% of reading at 4 MPa to below 60 MPa
0.008 % of reading at 60 MPa to 120 MPa

in-situ Pneumatic Instruments
0.03% of reading or 0.06 kPa (whichever is greater) from 1 kPa to 3500 kPa (absolute)


Pressure metrology - Pressure and vacuum measuring equipmentNULL Manometers; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure Comparison with dead weight tester; Comparison with reference instrument

As defined in MSA Test Method 1 or 2, as appropriate.

For liquid manometers as defined in in-house method 3_5_06 Master Calibration - Standard Operating Procedures for Liquid Manometers

Capability

with Calibration and Measurement Capability of -

Pneumatic instruments
0.006% of reading or 0.8 Pa (whichever is greater) from -100 kPa  to -1 kPa and
1 kPa to 3500 kPa
0.8 Pa from -1 kPa to 1 kPa
Hydraulic instruments
0.012 % of reading or 0.05 kPa (whichever is greater) from 0  MPa to below 4 MPa
0.010% of reading at 4 MPa to below 60 MPa
0.008 % of reading at 60 MPa to 120 MPa

in-situ Pneumatic instruments
1 Pa from -1 to 1 kPa
0.03% of reading or 0.06 kPa (whichever is greater)

Pressure metrology - Pressure standards Pressure calibrators - Non-dead weight Absolute pressure Comparison with dead weight tester; Comparison with reference instrument

As defined in MSA Test Method 1 .


Pneumatic instruments 1 kPa to 3500 kPa (abs)

Hydraulic instruments 500 kPa to 120 MPa (abs)

Capability

with Calibration and Measurement Capability of -
Pneumatic Instruments
0.006% of reading or 0.008 kPa (whichever is greater) from 1 kPa to 3500 kPa (absolute)
Hydraulic Instruments
0.008% of reading from 3500 kPa to 120 MPa  (absolute)

in-situ Pneumatic Instruments
0.03% of reading or 0.06 kPa (whichever is greater) from 1 kPa to 3500 kPa (absolute)


Pressure metrology - Pressure standardsNULL Pressure calibrators - Non-dead weight Gauge pressure Comparison with dead weight tester; Comparison with reference instrument As defined in MSA Test Method 1 

Capability

with Calibration and Measurement Capability of -

Pneumatic instruments
0.006% of reading or 1 Pa (whichever is greater) from -100 kPa to -1 kPa and 1 kPa to 3500 kPa
0.8 Pa from -1 kPa to 1 kPa
 Hydraulic instruments
0.008% of reading or 0.05 kPa (whichever is greater)

Temperature metrology - Humidity measuring equipment Hygrometers; Relative humidity sensors Relative humidity (RH) Comparison with a reference standard

Capability

including on-site calibration

with Calibration and Measurement Capability of -

Temperature
Calibration of temperature sensors for RH sensors

0.23 °C at 5 °C
 0.16 °C at 10 °C
 0.08 °C at 20 °C
 0.08 °C at 30 °C
 0.11 °C at 40 °C
 0.18 °C at 50 °C
 0.24 °C at 60 °C

Relative Humidity

Calibration of relative humidity sensors

all values are in % relative humidity













5  %

10 %

   20 %
30 %
40 %
50 %
60 %
70 %
80 %
90 %

95 %

5 °C

-

0.50

0.50
0.70
0.80
1.00
1.20
1.30
1.50
1.60

-

15 °C

0.40

0.50

0.60
0.60
0.65
0.70
0.80
0.85
0.95
1.00

1.05

21 °C

0.50

0.50

0.55
0.60
0.70
0.75
0.80
0.85
0.95
1.00

1.00

25 °C

0.50

0.50

0.55
0.65
0.70
0.75
0.80
0.85
0.95
1.00

1.05

44 °C

   0.45
   0.50
0.55
0.65
0.65
0.75
0.80
0.85
0.95
1.00
   1.05
  60 °C
     -
   0.50
0.50
0.60
0.70
0.80
0.90
1.00
1.20
1.20
(at 88 % RH)
     -


Temperature metrology - Temperature measuring equipment Digital temperature measuring systems Temperature Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

In the laboratory
0.025°C from - 90°C to below - 80°C
0.016°C from - 80°C to 250°C
0.005°C at 0°C

On site calibration
0.03°C from −90°C to below 150°C
0.05°C from 150°C to 250°C
0.01°C at 0°C

Temperature metrology - Verification of controlled enclosures Environmental chambers - Temperature; Freezers; Incubators; Ovens Spatial uniformity; Temperature Direct temperature measurement

AS2853-1986 and IPAC in-house method 4_7_84_02_Temp Lab SOP - Temperature Controlled Enclosure

Capability

with Calibration and Measurement Capability of -

including on-site calibrations

Freezers, ovens , environmental chambers 

0.5 °C from -90°C to  150 °C

Incubators

0.3 °C from 0 °C to 50 °C (thermistor sensors)

0.5 °C from 0 °C to 150 °C (thermocouple sensors)


Medical refrigeration equipment Spatial uniformity; Temperature Direct temperature measurement

AS2853-1986 and IPAC in-house method 4_7_84_02_Temp Lab SOP - Temperature Controlled Enclosure

Six-monthly tasks and annual tasks by the methods of AS3864.2 clauses 3.6 and 3.7 excluding clause 3.6.4 and IPAC in-house method 4_7_84_15_Temp Lab SOP – Medical

Spatial temperature distribution verification by the methods of AS3864.2 clause 3.8 and IPAC in-house method 4_7_84_15_Temp Lab SOP – Medical

Capability

with Calibration and Measurement Capability of -

including on-site calibrations

0.3 °C from -90 °C to 25 °C (thermistor sensors)

0.5 °C from -80 °C to 25 °C (thermocouple sensors)

Six-monthly tasks and annual tasks

0.1 °C from -90 °C to 25 °C


The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.