National Association of Testing Authorities, Australia
Scope of Accreditation
ENERGY QUEENSLAND LIMITED
Site
Meter Field Services
Accreditation No.
74
Site No.
20790
Date of Accreditation
Contact
Availability
Services available to external clients
Scope of Accreditation
Withdrawn
ISO/IEC 17025 (2005)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DC and low frequency electrical metrology - Electrical measurement and test equipment | Electricity meters | Active energy; Reactive energy | Direct measurement against a reference standard |
Watthour meters At 50 Hz, 63.5 V to 240 V, 0.2 A to 20 A Varhour meters At 50 Hz, 63.5 V to 240 V, 0.5 A to 20 A |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - Watthour meters 0.06%/cosΦ, three phase Varhour meters 0.5% from zero to 0.5 power factor lag |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - High-voltage/high-current standards and equipmentIncluding on site calibrations in-situ | Current transformers; Voltage transformers | Current; Phase displacement; Voltage | Comparison with a reference standard |
Current transformers (CTs) at 50 Hz Voltage transformers (VTs) at 50 Hz and test points specified in AS 60044 Parts 2 and 3, AS 1243 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations with Calibration and Measurement Capability of - Current transformers (CTs) at 50 Hz with rated primary currents from 1 A to 10 000 Awith rated secondary currents of 1 A or 5 A and at 0.5 lagging to unity power-factor burden 0.04% for current error Voltage transformers (VTs) at 50 Hz with rated primary voltages from 110/√ 3 V to 132/√ 3 kVand rated secondary voltages of 110/√ 3 V or 110 V and at 0.1 lagging to unity power-factor burden 0.04% for voltage error 0.04 crad for phase angle error |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.