The Australian Gas Association (AGA)

Site

The AGA Laboratory

Accreditation No.

17379

Site No.

23491

Date of Accreditation

01 Jun 2017

Address

66 Malcolm Road
Braeside , VIC 3195
Australia

aga.asn.au

Contact

Mr Dennis Karafotias P: +61 (03) 95804500

Availability

Services available to external clients

Scope of Accreditation

Withdrawn 01 Jul 2022

Voluntary

The AGA Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Withdrawn
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material in-house method QWI 737 

Capability

With calibration and measurement capability - 

CO in air: Electrochemical - 10 ppm to 3000 ppm CO v/v

4% of reading over the range of 10 ppm to 3000 ppm CO v/v

O2 in air: Electrochemical - 0.4% to 21% O2 v/v

4% of reading over the range 0.4% to 21% O2 v/v

Concentration Direct measurement by certified reference material in-house method QWI 736

Capability

With calibration and measurement capability - 

NO in air: Chemiluminescent - 0.1 ppm to 10 ppm NO v/v

4.3% of reading over the range 0.1 ppm to 10 ppm NO v/v

NOin air: Chemiluminescent - 0.1 ppm to 10 ppm NOv/v

4.3% of reading over the range 0.1 ppm to 10 ppm NOv/v

NOX in air: Chemiluminescent - 0.1 ppm to 10 ppm NOv/v

4.3% of reading over the range 0.1 ppm to 10 ppm NOv/v

Concentration Direct measurement by certified reference material in-house method QWI 738 

Capability

 With calibration and measurement capability - 

CO in air: NDIR  - 3 ppm to 3000 ppm CO v/v

 2.5% of reading over the range 3 ppm to 3000 ppm CO v/v

CO2 in air: NDIR -  0.1% to 20% COv/v

2.5% of reading over the range.of 0.1% to 20% COv/v

Concentration Direct measurement by certified reference material in-house method QWI 739

Capability

 With calibration and measurement capability - 

O2 in air: Paramagnetic -  0.4% to 21% Ov/v

2.4% of reading over the range 0.4% to 21% Ov/v

Flow metrology - Flow measuring devices and systems
Pressure metrology - Pressure and vacuum measuring equipment Manometers Gauge pressure Comparison with reference instrument

Capability

with Calibration and Measurement Capability of -
0.03 kPa from 0 kPa to 10 kPa

Pressure gauges; Pressure transducers Gauge pressure Comparison with reference instrument

Capability

with Calibration and Measurement Capability of -
0.03 kPa from 0 kPa to 10 kPa
3 kPa above 10 kPa to 4 MPa
Temperature metrology - Temperature measuring equipment

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.