National Association of Testing Authorities, Australia
Scope of Accreditation
The Australian Gas Association (AGA)
Site
The AGA Laboratory
Accreditation No.
17379
Site No.
23492
Date of Accreditation
Contact
Availability
Services available to external clients
Scope of Accreditation
Withdrawn
ISO/IEC 17025 (2005)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Chemical metrology - Gas analysers | Gas analysers; Mine safety and confined space equipment | Concentration | Direct measurement by certified reference material | in-house method QWI 738 |
CO in air: NDIR - 3 ppm to 3000 ppm CO v/v CO2 in air: NDIR - 0.1% to 20% CO2 v/v |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability - 2.5% of reading over the range 3 ppm to 3000 ppm CO v/v 2.5% of reading over the range.of 0.1% to 20% CO2 v/v |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Concentration | Direct measurement by certified reference material | in-house method QWI 739 |
O2 in air: Paramagnetic - 0.4% to 21% O2 v/v |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability - 2.4% of reading over the range 0.4% to 21% O2 v/v |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Concentration | Direct measurement by certified reference material | in-house method QWI 736 |
NO in air: Chemiluminescent - 0.1 ppm to 10 ppm NO v/v NO2 in air: Chemiluminescent - 0.1 ppm to 10 ppm NO2 v/v NOXin air: Chemiluminescent - 0.1 ppm to 10 ppm NOx v/v |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability - 4.3% of reading over the range 0.1 ppm to 10 ppm NO v/v 4.3% of reading over the range 0.1 ppm to 10 ppm NO2 v/v 4.3% of reading over the range 0.1 ppm to 10 ppm NOx v/v |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Concentration | Direct measurement by certified reference material | in-house method QWI 737 |
CO in air: Electrochemical - 10 ppm to 3000 ppm CO v/v O2 in air: Electrochemical - 0.4% to 21% O2 v/v |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With calibration and measurement capability - 4% of reading over the range of 10 ppm to 3000 ppm CO v/v 4% of reading over the range 0.4% to 21% O2 v/v |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.