National Measurement Institute (NMI)

Site

Trade Measurement Laboratory, Hobart

Accreditation No.

1

Site No.

14817

Date of Accreditation

Address

15A Lampton Avenue
Derwent Park, TAS 7009
Australia

measurement.gov.au

Contact

Availability

Services conditionally available to external clients

Scope of Accreditation

Withdrawn

ISO/IEC 17025 (2005)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Withdrawn
Service Product Determinant Technique Procedure Limitations
Flow metrology - Flow measuring devices and systems Liquid meters Flow rate

Master meters for water
for flow rates from 50 L/min up to and including 1000 L/min

Capability

with Calibration and Measurement Capability of-

0.05% of volume with a minimum 500 L test volume

Mass - Determination of mass and calibration of weighing devices Industrial mass standards Mass Gravimetric measurement against reference mass

From 100 g to 1200 kg


Capability

with Calibration and Measurement Capability of-

1 in 105 from 100 g up to 1200 kg

Mass

2 in 105 from 1 g up to 1200 kg

 
Industrial weighing devices Mass Gravimetric measurement against reference mass; Substitution gravimetric measurement

by the method and specification of National Instrument Test Procedures (NITP) 6.1–6.4

Capability

with Calibration and Measurement Capability of-

1 in 104 from 1 kg up to and including 30 t for direct loading and 150 t using the substitution method


Laboratory weighing devices; Precision laboratory balances Mass Gravimetric measurement against reference mass

From 1 mg to 1500 kg

Capability

with Calibration and Measurement Capability of-

Precision balances and laboratory weighing devices

3 μg from 1 mg up to 50 mg;
4 μg above 50 mg up to and including 0.2 g;
8 μg above 0.2 g up to and including 2 g;
15 μg above 2 g up to and including 10 g;
30 μg above 10 g up to and including 100 g;
4 in 107 above 100 g up to and including 40 kg

4 in 106 above 40 kg up to and including 500 kg

1 in 105 above 500 kg up to and including 1500 kg 


 

Mass standards Mass Gravimetric measurement against reference mass

From 1 mg to 50 kg

Capability

with Calibation and Measurement Capability of -

3 μg from 1 mg up to and including 10 mg;
5 μg above 10 mg up to and including 200 mg;
10 μg above 200 mg up to and including 2 g;
2 in 106 above 2 g up to and including 50 g;
1 in 106 above 50 g up to and including 25 kg;
6 in 106 above 25 kg up to and including 50 kg

Volume and density - Determination of density Density of liquids kg/m3 To be determined

500 kg/m3 to 1200 kg/m3

Capability

with Calibration and Measurement Capability of-

0.1%

Volume and density - Industrial volumetric equipment and bulk storage Bulk storage tanks; Industrial volumetric proving measures; Standard measures - > 25 L Volume

Proving tanks by standard measures
Proving tanks by gravimetric methods
From 1 L up to 1000 L

Standard measures 

From 2 L to 1000L

Industrial storage tanks of capacity greater than 500 L

 

Capability

with Calibration and Measurement Capability of-

Proving tanks by standard measures
1 in 104 from 1 L up to 5000 L


Proving tanks by gravimetric methods
10 in 104 from 1 L up to 1000 L

Standard Measures

7.5 in 105 from 2 L up to 1000 L

Industrial storage tanks
0.1%

Volume and density - Laboratory volumetric glassware and measures Standard measures - < 50 L; Volumetric glassware and measures Volume Gravimetric

Volumetric glassware and standard measures

From 0.1 mL to 2 L

Capability

with Calibration and Measurement Capability 0f-

2 in 103 or 0.001 mL (whichever is greater) from 0.1 mL up to 1 mL;
1 in 104 or 0.004 mL (whichever is greater) above 1 mL up to 2 L


The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.