Active Environmental Solutions Australia

Site

Active Environmental Solutions Australia Victorian Laboratory

Accreditation No.

21576

Site No.

25985

Date of Accreditation

20 Oct 2025

Address

2 Merchant Avenue
Thomastown, VIC 3074
Australia

aesolutions.com.au

Contact

Availability

Services available to external clients

Scope of Accreditation

Active Environmental Solutions Australia Victorian Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Flow metrology - Environmental flow measuring devices and systems Air sampling pumps and meters Flow rate Comparison with a reference meter; Volumetric transfer AES – LAB–TM01 (Air
Sampling Pump Calibration -
Back Pressure
Test and Leak
Test only)
AES – LAB–TM01 (Flow Meter Calibration using the Bubble Flow meter as reference)

Capability



Calibration of gas flow meters and pumps using air at atmospheric conditions as the test fluid with a minimum test volume of 100 mL at 100 mL/min up to 1 L/min and 1 L at 1 L/min to 5 L/min
with a Calibration and Measurement Capability of -
Flow meters
0.66% of flow rate from 100 mL/min up to 1 L/min
0.34% of flow rate at 1 L/min to 5 L/min
Pumps
0.65% of change of flow rate from 200 mL/min up to 1 L/min
0.25% of change of flow rate at 1 L/min to 4 L/min



The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

The Product column may also describe an Item or Material.

The Limitations column may also describe the Range.