Scope of Accreditation

Withdrawn 21 Sep 2026

Voluntary

Adelaide Laboratory

ISO/IEC 17025 (2017)

Calibration

Withdrawn
Service Product Determinant Technique Procedure Limitations
Mass - Determination of mass and calibration of weighing devices Industrial weighing devices; Laboratory weighing devices; Precision laboratory balances Mass Measurement against a reference standard. In house (Method 02-28)

Capability

including on-site calibrations
with Calibration and Measurement Capability of
Precision laboratory balances
1 in 105 + 200 μg, from 1 g to 5000 g
Industrial balances
3 in 105 + 1 mg, from 1 g to 5000 g
Pressure metrology - Pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure By pressure calibrator; Comparison with reference instrument

Capability

including on-site calibrations
with Calibration and Measurement Capability of
0.05 kPa from - 95 kPa to below -60 kPa
0.015 kPa + 0.07% of reading from -60 kPa to below 0 kPa
0.025 kPa from 0 kPa to 200 kPa
 0.5 kPa above 200 kPa to 700 kPa
0.8 kPa above 700 kPa to 2000 kPa
1.0 kPa above 2000 kPa to 4000 kPa
Temperature metrology - Ancillary temperature measuring equipment Multi-channel thermocouple data recorders; Process calibrators Resistance to temperature conversion; Volt to temperature conversion Direct measurement by electrical input

Capability

Including on site calibrations
With Calibration and Measurement Capability of-
 0.2ºC + 0.01% of reading from -200ºC to 1300ºC
Temperature metrology - Temperature measuring equipment Digital temperature measuring systems; Temperature loggers - Single and multi-channel Temperature Measurement against reference standard In house (Method 02-07, 02-08, 02-10, 02-05, 02-01)

Capability

With Calibration and Measurement Capability of-
In the range from -196ºC to 1300ºC
Including on site calibrations

0.25°C from -196°C to below -40°C
0.005°C at 0°C (using RTD sensors)
0.05°C from -40°C to 300°C
0.25°C above 300°C to 650°C
3°C above 650°C to 1300°C

Temperature metrology - Verification of controlled enclosures Autoclaves and sterilising ovens; Baths; Cryogenic storage tanks; Dry block heaters; Environmental chambers - Temperature; Freezers; Furnaces; Incubators; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement AS 2853 - 1986
IEC 60068 (3-5)

Capability

Including on site calibrations

With Calibration and Measurement Capability of-

In the range from -196ºC to 1300ºC

0.5ºC from -196ºC to -40ºC
0.25ºC above -40ºC to 0ºC
0.5ºC above 0ºC to 500ºC
2.5ºC above 500ºC to 1300ºC

Temperature metrology - Verification of controlled enclosures Autoclaves and sterilising ovens; Baths; Cryogenic storage tanks; Dry block heaters; Environmental chambers - Temperature; Freezers; Furnaces; Incubators; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement AS 2853 - 1986
IEC 60068 (3-5)

Capability

Including on site calibrations

With Calibration and Measurement Capability of-

In the range from -196ºC to 1300ºC

0.5ºC from -196ºC to -40ºC
0.25ºC above -40ºC to 0ºC
0.5ºC above 0ºC to 500ºC
2.5ºC above 500ºC to 1300ºC

Temperature metrology - Verification of controlled enclosures Environmental chambers - Humidity Humidity; Spatial uniformity By methods of AS/IEC 60068.3.6 IEC 60068-3-6

Capability

Including on site calibrations
With Calibration and Measurement Capability of-
In the range from 10% rh to 95% rh from 10ºC to 90ºC
2.5% rh from 10% h to 95% rh from 10ºC to 90ºC
Temperature metrology - Verification of controlled enclosures Environmental chambers - Humidity Humidity; Spatial uniformity By temperature and humidity measurement

Capability

Including on site calibrations

With Calibration and Measurement Capability of-

In the range from 10% rh to 95% rh from 10ºC to 90ºC

2.5% rh from 10% h to 95% rh from 10ºC to 90ºC

Temperature metrology - Verification of controlled enclosures Environmental chambers - Humidity Humidity; Spatial uniformity By temperature and humidity measurement WHO TRS 961, Annex 9: Model guidance for the storage and transport of time and temperature sensitive pharmaceutical products

Capability

Including on-site calibrations
with Calibration and Measurement Capability of:
 2.5%rh from 10% to 95%rh, from 10ºC to 90ºC
Temperature metrology - Verification of controlled enclosures Environmental chambers - Humidity Humidity; Spatial uniformity By temperature and humidity measurement

Capability

Including on site calibrations

With Calibration and Measurement Capability of-

In the range from 10% rh to 95% rh from 10ºC to 90ºC

2.5% rh from 10% h to 95% rh from 10ºC to 90ºC

Temperature metrology - Verification of controlled enclosures Environmental chambers - Temperature; Freezers Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement WHO TRS 961, Annex 9: Model guidance for the storage and transport of time and temperature sensitive pharmaceutical products

Capability

Including on-site calibrations
with Calibration and Measurement Capability of:
 0.5°C In the range from -40°C to 90ºC
Temperature metrology - Verification of controlled enclosures Medical refrigeration equipment Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement AS 3864.2 (excluding clause 3.6.4)

Capability

Including on site calibrations

For the following enclosure types
Blood fridges (upright, single & multi-door, under bench and walk-in rooms) ;
Plasma freezers (ultra-low, upright, chest and walk-in rooms)

With Calibration and Measurement Capability of-

In the range from -196ºC to 20ºC

0.5ºC from -196ºC to -40ºC
0.25ºC above -40ºC to 0ºC
0.5ºC above 0ºC to 20ºC

Temperature metrology - Verification of controlled enclosures Medical refrigeration equipment Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement AS 3864.2 (excluding clause 3.6.4)

Capability

Including on site calibrations

For the following enclosure types
Blood fridges (upright, single & multi-door, under bench and walk-in rooms) ;
Plasma freezers (ultra-low, upright, chest and walk-in rooms)

With Calibration and Measurement Capability of-

In the range from -196ºC to 20ºC

0.5ºC from -196ºC to -40ºC
0.25ºC above -40ºC to 0ºC
0.5ºC above 0ºC to 20ºC