Scope of Accreditation

AES Environmental

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Sound level Not applicable AS 1807.20 (2000),
AS 1807 clause 4.7
for compliance with:
AS 2252.2
Inward air velocity Cold DOP AS 1807.21 (2000),
AS 1807 clause 4.8
for compliance with:
AS 2252.2
Ultraviolet intensity Not applicable AS 1807.23 (2000),
AS 1807 clause 4.10
for compliance with:
AS 2252.2
Work zone integrity Not applicable AS 1807.5 (2000),
AS 1807 clause 4.3
for compliance with:
AS 2252.2
Air barrier containment Cold DOP AS 1807.22 (2000),
AS 1807 clause 4.9
for compliance with:
AS 2252.2
Particle count in work zone Automatic particle counter AS 1807.8 (2000),
AS ISO 14644.1 Appendix A
for compliance with:
AS 2252.2
Illuminance Not applicable AS 1807.15 (2000),
AS 1807 clause 4.5
for compliance with:
AS 2252.2
Integrity of HEPA filter installations Cold DOP AS 1807.6 (2000),
AS 1807.7,  
AS 1807 clause 4.4
for compliance with:
AS 2252.2
Air velocity and uniformity of air velocity Cold DOP

AS 1807.1 (2000),
AS 1807 clause 4.1

for compliance with:
AS 2252.2
Clean rooms Sound level Not applicable AS 1807.16

for compliance with:
AS ISO 14644.1
Recovery times Not applicable AS 1807.24 (2000),
AS ISO 14644.3 Method B.4
for compliance with:
AS ISO 14644.1
Clean rooms, pharmaceutical isolators and work stations Air velocity and uniformity of air velocity Cold DOP

AS 1807.1 (2000),
AS 1807 clause 4.1

for compliance with:
AS 2252.6,
AS ISO 14644.1
Integrity of HEPA filter installations Cold DOP AS 1807.6 (2000),
AS 1807.7,
AS 1807 clause 4.4,
for compliance with:
AS 2252.6,
AS ISO 14644.1
Air velocity and uniformity of air velocity Cold DOP AS 1807.3 (2000),
AS ISO 14644.3 Method B.4
for compliance with:
AS 2252.6,
AS ISO 14644.1
Ultraviolet intensity Not applicable AS 1807.23 (2000),
AS 1807 clause 4.10
for compliance with:
AS 2252.6,
AS ISO 14644.1
Particle count in work zone Automatic particle counter AS 1807.8 (2000),
AS ISO 14644.1 Appendix A,
AS/NZS 14644.3 Appendix B.1, B.2, B.3
for compliance with:
AS 2252.6,
AS ISO 14644.1
Work zone integrity Cold DOP AS 1807.5 (2000),
AS 1807 clause 4.3
for compliance with:
AS 2252.6,
AS ISO 14644.1
Air pressure Not applicable AS 1807.10(2000),
AS ISO 14644.3 Method B.1
for compliance with:
AS 2252.6,
AS ISO 14644.1
Illuminance Not applicable AS 1807.15 (2000),
AS 1807 clause 4.5,
for compliance with:
AS 2252.6,
AS ISO 14644.1
Clean workstations; Pharmaceutical isolators Sound level Not applicable AS 1807.20 (2000),
AS 1807 clause 4.7
for compliance with:
AS 2252.6
Fume cupboards and recirculating fume cabinets Extraction performance Face velocity; Not applicable AS/NZS 2243.8 Appendix  B,
AS/NZS 2243.9 Appendix  B
for  compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9
Extraction performance Smoke visualisation

AS/NZS 2243.8 Appendix A
AS/NZS 2243.9 Appendix A

for  compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9
Illuminance Not applicable AS 1807.15 (2000),
AS 1807 clause 4.5
for  compliance with:
AS/NZS 2243.8,
AS/NZS 2243.9