Air-Met Scientific Pty Ltd

Site

Air-Met Scientific Pty Ltd

Accreditation No.

14926

Site No.

21449

Date of Accreditation

20 Aug 2015

Address

7-11 Ceylon Street
Nunawading, VIC 3131
Australia

airmet.com.au

Contact

Mr Jonathan Alexander P: 1800000744

Availability

Services available to external clients

Scope of Accreditation

Air-Met Scientific Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Flow metrology - Environmental flow measuring devices and systems Air sampling pumps and meters Flow rate Comparison with a reference meter; Volumetric transfer WI 802.10 (Air Sampling Pump Calibration - Back Pressure Test and Leak Test only) ;
WI 803.10 (Flow Meter Calibration using the Bubble Flow meter as reference)

Capability


With Calibration and Measurement Capability of -
Calibration of gas flow meters and pumps using air at atmospheric conditions
as the test fluid with a minimum test volume of 1000 mL
with a Calibration and Measurement Capability of -
1.0% of flow rate from 200 mL/min to 2500 mL/min
(2.0% of flow rate minus 25 mL/min) above 2500 mL/min to 5000 mL/min

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.