Air-Met Scientific Pty Ltd

Site

Air-Met Scientific Pty Ltd – Mackay Site

Accreditation No.

14926

Site No.

24703

Date of Accreditation

01 Feb 2021

Address

135 Sydney Street
Mackay, QLD 4740
Australia

airmet.com.au

Contact

Mr Jonathan Alexander P: +61 0458867178

Availability

Services available to external clients

Scope of Accreditation

Air-Met Scientific Pty Ltd – Mackay Site

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers; Mine safety and confined space equipment Concentration Comparison measurement against a certified reference material; Direct measurement by certified reference material in-house methods B; C; S Including on-site calibrations

Capability

Including on site calibrations -
With a calibration and measurement capability of -

Ammonia: Electrochemical detector with reading in parts per million (ppm).
1x10-6 %(v/v) at zero instrument reading
15.5% of the reading over the range of 9.9x10-6 to 178.2 x 10-6 %(v/v) of ammonia.

Butene: Photo ionisation detector with reading in parts per million (ppm).
1x10-6 %(v/v) at zero instrument reading
5.0% of the reading over the range of 9.7x10-6 to 175x10-6 %(v/v) of butene

Carbon dioxide: Infrared/Electrochemical with reading in % (volume/volume).
1x10-6 %(v/v) at zero instrument reading
5.6% of the reading over the range of 1x10-6 to 18x10-6 %(v/v) of carbon dioxide.

Carbon monoxide: Electrochemical detector with reading in parts per million reading (ppm).
1x10-6 %(v/v) at zero instrument
5.0% of the reading over the range of 10.1x10-6 to 182x10-6 %(v/v) of carbon monoxide.

Hexane: Catalytic combustion detector with reading in % Lower Explosive Limit (LEL).
1x10-6 %(v/v) at zero instrument reading
5.6% of the reading over the range of 2.5x10-6 to 45x10-6 %(v/v) of hexane.

Hydrogen: Catalytic combustion detector with reading in % Lower Explosive Limit (LEL).
1x10-6 %(v/v) at zero instrument reading
6.5% of the reading over the range of 5x10-6 to 90x10-6 %(v/v) of hydrogen.

Hydrogen : Electrochemical detector with reading in parts per million (ppm).
1x10-6 %(v/v) at zero instrument reading
6.5% of the reading over the range of 48x10-6 to 864x10-6 %(v/v) of hydrogen.

Hydrogen sulfide : Electrochemical detector with reading in parts per million (ppm).
0.1x10-6 %(v/v) at zero instrument reading
9.9% of the reading over the range of 2.5x10-6 to 45x10-6 %(v/v) of hydrogen sulfide.

Methane : Catalytic combustion detector with reading in % (volume/volume).
0.01x10-6 %(v/v) at zero instrument reading
2.3% of the reading over the range of 0.3x10-6 to 4.5x10-6 %(v/v) of methane.

Methane : Catalytic combustion detector with reading in % Lower Explosive Limit (LEL).
1x10-6 %(v/v) at zero instrument reading
5.6% of the reading over the range of 5.0x10-6 to 90.0x10-6 %(v/v) of methane.

Methane : Flame ionisation detector with reading in % (volume/volume).
0.01x10-6 %(v/v) at zero instrument reading
4.5% of the reading over the range of 0.3x10-6 to 4.5x10-6 %(v/v) of methane.

Methane : Infrared detector with reading in % (volume/volume).
1x10-6 at zero instrument reading
5.5% of the reading over the range of 0.8x10-6 to 14.2x10-6 %(v/v) of methane.

Methane : Solid-state semiconductor detector with reading in parts per million (ppm).
1x10-6 %(v/v) at zero instrument reading
5.6% of the reading over the range of 0.8x10-6 to 14.2x10-6 %(v/v) of methane.

Methane : Thermal conductivity detector with reading in % Lower Explosive Limit (LEL).
1x10-6 %(v/v) at zero instrument reading
6.1% of the reading over the range of 0.3x10-6 to 4.5x10-6 %(v/v) of methane.

Nitric oxide : Electrochemical detector with reading in parts per million (ppm).
1x10-6 %(v/v) at zero instrument reading
4.5% of the reading over the range 5.4x10-6 to 97.2x10-6 %(v/v) of nitric oxide.

Nitrogen dioxide : Electrochemical detector with reading parts per million (ppm).
0.1x10-6 %(v/v) at zero instrument reading
5.2x10-6 to 93.6x10-6 %(v/v) of nitrogen dioxide.

Oxygen : Electrochemical detector with reading in % (volume/volume).
0.1x10-6 %(v/v) at zero instrument reading
1.7% of the reading over the range 1.8x10-6 to 31.9x10-6 %(v/v) of oxygen.

Oxygen : Electrochemical detector with reading in parts per million (ppm).
1x10-6 %(v/v) at zero instrument reading
8.1% of the reading over the range of 100x10-6 to 1800x10-6 %(v/v) of oxygen.

Pentane : Catalytic combustion detector with reading in % Lower Explosive Limit (LEL).
1x10-6 %(v/v) at zero instrument reading
3.6% of the reading over the range of 2.50x10-6 to 45x10-6 %(v/v) of pentane.

Sulfur dioxide : Electrochemical detector with reading in parts per million (ppm).
0.1x10-6 %(v/v) at zero instrument reading
5.5% of the reading over the range of 4.9x10-6 to 88.2x10-6 %(v/v) of sulfur dioxide.