National Association of Testing Authorities, Australia
Scope of Accreditation
AMT Group Australasia (Melbourne)
Site
AMT – Melbourne
Accreditation No.
20596
Site No.
24688
Date of Accreditation
07 Feb 2020
Contact
Mr Cameron Welch P: +61 0431237038
Availability
Services available to external clients
Scope of Accreditation
AMT – Melbourne
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Inward air velocity | Cold DOP | AS 1807 clause 4.8, | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), |
Biological safety cabinets; Cytotoxic drug safety cabinets | Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1, | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), | |
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) | ||
Air barrier containment | Cold DOP | AS 1807 clause 4.9 | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) | ||
Sound level | Not applicable | AS 1807 clause 4.7, | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), | ||
Work zone integrity | Cold DOP | AS 1807 clause 4.3 | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.7, AS 1807 clause 4.4, | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), AS 4260 clause 5.2, | ||
Illuminance | Not applicable | AS 1807 clause 4.5, | for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), | ||
Clean rooms, pharmaceutical isolators and work stations | Work zone integrity | Cold DOP | AS 1807 clause 4.3 | for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) | |
Sound level | Not applicable | AS 1807 Clause 4.7 AS 1807.16 | for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.7, AS 1807 clause 4.4, AS ISO 14644.3 Method B.7 | for compliance with: AS 2252.6 (Clean Workstation), AS 4260 clause 5.2, AS 4273 clause 9.3 (Pharmaceutical isolators), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1 | for compliance with: AS 2252.6 (Clean Workstation), AS 4273 clause 9.4 (Pharmaceutical isolators) | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) | ||
Fume cupboards and recirculating fume cabinets | Extraction performance | Face velocity | AS/NZS 2243.8 Appendix B, AS/NZS 2243.9 Appendix B | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | |
Extraction performance | Smoke visualisation | AS/NZS 2243.8 Appendix A, AS/NZS 2243.9 Appendix A | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Sound level | Not applicable | AS 1807 clause 4.7 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
20596
Site No.
24688
Print date
26 Nov 2024
END OF SCOPE