ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
| Service | Product | Determinant | Technique | Procedure | Limitations |
|---|---|---|---|---|---|
| Evaluation of air control equipment | Biological safety cabinets | Inward air velocity | Cold DOP |
AS 1807 clause 4.8, |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), |
| Biological safety cabinets; Cytotoxic drug safety cabinets | Sound level | Not applicable |
AS 1807 clause 4.7, |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), |
|
| Air velocity and uniformity of air velocity | Cold DOP |
AS 1807 clause 4.1, |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), |
||
| Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
| Illuminance | Not applicable |
AS 1807 clause 4.5, |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), |
||
| Integrity of HEPA filter installations | Cold DOP |
AS 1807.7, AS 1807 clause 4.4, |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets), AS 4260 clause 5.2, |
||
| Air barrier containment | Cold DOP |
AS 1807 clause 4.9 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
| Work zone integrity | Cold DOP |
AS 1807 clause 4.3 |
for compliance with: AS 2252.1 (2002), AS 2252.2 (Biological safety cabinets), AS 2252.5 (Cytotoxic safety cabinets) |
||
| Clean rooms, pharmaceutical isolators and work stations | Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1 |
for compliance with: AS 2252.6 (Clean Workstation), AS 4273 clause 9.4 (Pharmaceutical isolators) |
|
| Sound level | Not applicable |
AS 1807 Clause 4.7 AS 1807.16 |
for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
| Integrity of HEPA filter installations | Cold DOP |
AS 1807.7, AS 1807 clause 4.4, AS ISO 14644.3 Method B.7 |
for compliance with: AS 2252.6 (Clean Workstation), AS 4260 clause 5.2, AS 4273 clause 9.3 (Pharmaceutical isolators), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
| Illuminance | Not applicable | AS 1807 clause 4.5 |
for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
| Work zone integrity | Cold DOP | AS 1807 clause 4.3 |
for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS ISO 14644.3 (Clean Room) |
||
| Clean rooms, pharmaceutical isolators and work stations; Cytotoxic drug safety cabinets | Airborne particle count | Automatic particle counter |
AS ISO 14644.1 |
For Compliance with: AS ISO 14644.1, AS 2252.5, AS 2252.6 |
|
| Fume cupboards and recirculating fume cabinets | Extraction performance | Face velocity |
AS/NZS 2243.8 Appendix B, AS/NZS 2243.9 Appendix B |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
|
| Sound level | Not applicable | AS 1807 clause 4.7 |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
||
| Extraction performance | Smoke visualisation |
AS/NZS 2243.8 Appendix A, AS/NZS 2243.9 Appendix A |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
||
| Illuminance | Not applicable | AS 1807 clause 4.5 |
for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
The Product column may also describe an Item or Material.
The Limitations column may also describe the Range.
Accreditation No.
20596
Site No.
24688
Print date
25 Aug 2026