National Association of Testing Authorities, Australia
Scope of Accreditation
ASG Exical Pty Ltd as trustee for the ASG Exical Unit Trust
Site
ASG-X-Technologies
Accreditation No.
15464
Site No.
16363
Date of Accreditation
01 Aug 2006
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DC and low frequency electrical metrology - Electrical instrument calibrators | Instrument calibrators | AC current; AC voltage; DC current; DC voltage; Resistance | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - DC Voltage 0.0017% of reading + 1.4 μV from 0 V to 0.1 V 0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz 0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz 1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz 0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz 0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz 0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz 0.04% of reading from 10 V to 100 V at 10 Hz 0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz 0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz 0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz 0.07% of reading from 100 V to 700 V at 40 Hz 0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz 0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz DC Current 0.004% of reading + 0.002 μA from 0 μA to 100 μA 0.003% of reading + 0.14 μA above 100 μA to 10 mA 0.005% of reading + 1.9 μA above 10 mA to 100 mA 0.014% of reading + 30 μA above 100 mA to 1 A 0.14% of reading + 0.75 mA above 1 A to 3A 0.18% of reading + 0.22 mA above 3 A to 10 A AC Current 0.6% of reading from 10 μA to 10 A at 10 Hz 0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 0.003% of reading + 2 mΩ above 10 Ω to 100 Ω 0.002% of reading + 1.5 Ω above 100 Ω to 100 kΩ 0.003% of reading + 24 Ω above 100 kΩ to 1 MΩ 0.01% of reading + 730 Ω above 1 MΩ to 10 MΩ 0.085% of reading + 0.06 MΩ above 10 MΩ to 100 MΩ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - Electrical measurement and test equipment | Ammeters; Digital multimeters (DMM); Ohm meters; Voltmeters | AC current; AC voltage; DC current; DC voltage; Resistance | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - DC Voltage 0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz 0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz 1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz 0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz 0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz 0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz 0.04% of reading from 10 V to 100 V at 10 Hz 0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz 0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz 0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz 0.07% of reading from 100 V to 700 V at 40 Hz 0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz 0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz DC Current 0.004% of reading + 0.002 μA from 0 μA to 100 μA 0.003% of reading + 0.14 μA above100 μA to 10 mA 0.005% of reading + 1.9 μA above 10 mA to 100 mA 0.014% of reading + 30 μA above 100 mA to 1 A 0.14% of reading + 0.75 mA above 1 A to 3A 0.18% of reading + 0.22 mA above 3 A to 10 A AC Current 0.6% of reading from 10 μA to 10 A at 10 Hz 0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 0.003% of reading + 2 mΩ above 10 Ω to 100 Ω 0.002% of reading + 1.5 Ω above 100 Ω to 100 kΩ 0.003% of reading + 24 Ω above 100 kΩ to 1 MΩ 0.01% of reading + 730 Ω above 1 MΩ to 10 MΩ 0.085% of reading + 0.06 MΩ above 10 MΩ to 100 MΩ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
DC and low frequency electrical metrology - Electrical standards | Conductance boxes; Precision resistors; Resistance boxes | Resistance | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 0.003% of reading + 2 mΩ above 10 Ω to 100 Ω 0.002% of reading + 1.5 Ω above 100 Ω to 100 kΩ 0.003% of reading + 24 Ω above 100 kΩ to 1 MΩ 0.01% of reading + 730 Ω above 1 MΩ to 10 MΩ 0.085% of reading + 0.06 MΩ above 10 MΩ to 100 MΩ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure and vacuum measuring equipmentincluding on site calibrations | Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges | Gauge pressure; High vacuum | Comparison with dead weight tester; Comparison with reference instrument | Electrical transducers with dc voltage, dc current or resistance outputs | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Gauge pressure calibration with Calibration and Measurement Capability of- 0.016% reading or 0.016 kPa (whichever is greater) from -95 kPa to 100 kPa 0.016% reading above 100 kPa to 120 MPa On-site calibration with Calibration and Measurement Capability of- 0.13% of reading or 0.13 kPa (whichever is greater) from -95 kPa to 120 MPa | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pressure metrology - Pressure standards | Pressure calibrators - Non-dead weight | Gauge pressure | Comparison with dead weight tester; Comparison with reference instrument | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Time and frequency metrology - Frequency, time and waveform measuring equipment | Counters; Stroboscopes; Tachometers; Time interval meters | Frequency; Rotational speed; Time interval | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - Counters0.045% of reading from 1 Hz to 40 Hz 0.045% of reading from 60 rpm to 2400 rpm 0.015% of reading from 2400 rpm to 90,000 rpm Stroboscopes 0.045% of reading from 1 Hz to 40 Hz 0.015% of reading above 40 Hz to 1.5 kHz Time interval meters 0.015 % of reading from 100 ns to 1 s |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
15464
Site No.
16363
Print date
07 Dec 2024