Australian Federal Police

Site

Breath Analyser Calibration Laboratory

Accreditation No.

10821

Site No.

15787

Date of Accreditation

17 Nov 2005

Address

Traffic Operations Centre, 24 Sawmill Street
Hume, ACT 2620
Australia

afp.gov.au

Contact

Availability

Services conditionally available to external clients

Scope of Accreditation

Breath Analyser Calibration Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Breath alcohol instruments Breath alcohol instruments; Drager 9510 Alcohol content Direct measurement against a reference standard

in-house methods TI_79; WI_42

Capability

Drager Alcotest 9510

With Calibration and Measurement Capability of -
0.0023 g/210L for above 0.000 up to 0.050 g/210L
0.0031 g/210L for above 0.050 up to 0.100 g/210L
 3.5% for above 0.100 up to 0.500 g/210L


The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.