Kellip Enterprises Pty Ltd t/a Sharp Conditioning Solutions

Site

Calibration Activities

Accreditation No.

17088

Site No.

21995

Date of Accreditation

06 May 2014

Address

59 Hobart Street
Riverstone, NSW 2765
Australia

sharpcs.com.au

Contact

Mr Phillip Hall P: +61 1800474277

Availability

Services available to external clients

Scope of Accreditation

Calibration Activities

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Temperature metrology - Temperature measuring equipment Digital temperature measuring systems Temperature Measurement against reference standard

Capability

With Calibration and Measurement Capability of-
0.66 ºC from -90 ºC to below -40 ºC
0.36 ºC from -40 ºC to below 0 ºC
0.15 ºC from 0 ºC to below 60 ºC
0.7 ºC from 60 ºC to below 200 ºC
0.8 ºC from 200 ºC to below 300 ºC
1.4 ºC from 300 ºC to 600 ºC
Temperature metrology - Verification of controlled enclosuresIncluding onsite calibrations Environmental chambers - Temperature; Freezers; Incubators; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement IEC 60068-3-5
IEC 60068-3-6
Excluding rate of change and heating tests
In-house procedure SWM-PRO-0075 Spatial Mapping for Chambers Procedure

Capability

Including on site calibrations
with calibration and measurement capability of -
2.4 ºC from -80 ºC to below 0 ºC
0.7 ºC at 0 ºC
 2.8 ºC above 0 ºC to 40 ºC
Temperature metrology - Verification of controlled enclosuresincluding on site calibrations Medical refrigeration equipment Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement By the methodology of AS 3864.2

Capability

Temperature verifications and calibrations of medical refrigeration equipment;
Including on site calibration and
Including the following enclosure types:
Blood fridges (upright, single and multi-door and under-bench) and Plasma freezers (ultra-low, upright and chest)
As per the methodology of AS 3864.2
Spatial uniformity
With Calibration and Measurement Capability of-
0.5 ºC from -40 ºC to below 0 ºC
0.3 ºC from 0 ºC to 15 ºC
Annual Tasks
With Calibration and Measurement Capability of-
0.4 ºC from -40 ºC to below 0 ºC
0.3 ºC from 0 ºC to 15 ºC

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.