Civil Calibration Services Pty Limited

Site

Calibration Laboratory

Accreditation No.

4935

Site No.

4928

Date of Accreditation

31 Oct 1995

Address

2/42 Beaumont Road
Mount Kuring-gai, NSW 2080
Australia

Contact

Availability

Services available to external clients

Scope of Accreditation

Calibration Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Dimensional metrology - Engineering equipment and precision instruments Dial gauges; Electronic length gauges Length measurements Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

2.9 μm from 0 mm to 25 mm

Force metrology - Force measuring and testing equipment Compression and universal machines in compression Force in compression Comparison measurement with reference load cell Calibration to Class A requirements of AS 2193




Capability

Including on site calibrations

with Calibration and Measurement Capability of -

0.25% of reading above 0.5 kN to 100 kN
0.5% of reading above 100 kN to 200 kN
0.25% of reading above 200 kN to 2000 kN

Elastic force measuring devices Force in compression Comparison measurement with reference load cell; Gravimetric measurement with reference mass Class A requirements of AS 2193

Capability

Including on site calibration of load cells

Calibration of devices such as load measuring rings and load cells with digital displays for CBR and UCS machines in compression

with Calibration and Measurement Capability of -

0.5% of reading from 0.05 kN to 0.5 kN
0.25% of reading above 0.5 kN to 100 kN
0.5% of reading above 100 kN to 200 kN
0.25% of reading above 200 kN to 2000 kN


Force metrology - Hardness standards and equipment Rubber bases Rubber hardness Comparison with a reference standard

Capability

Rubber hardness measurement 

With Calibration and Measurement Capability of -

1 IRHD from 86 IRHD to 94 IRHD

Mass - Determination of mass and calibration of weighing devices Industrial mass standards Mass Gravimetric measurement against reference mass

Capability

with Calibration ad Measurement Capability of -

0.12 g from 100 g to 2 kg
0.17 g from 5 kg to 10 kg

Industrial weighing devices; Laboratory weighing devices Mass Gravimetric measurement against reference mass

Capability

Including on site calibrations

with Calibration and Measurement Capability of -

1 in 105 or 0.15 g (whichever is greater) from 100 g to 20 kg
1 in 105 or 0.45 g (whichever is greater) above 20 kg to 30 kg


Vibration metrology - Vibration equipment Vibrating tables Displacement sensitivity; Frequency Direct measurement by electrical input

Verification of vibrating tables and tools to the requirements of AS1289 5.5.1
Verification of vibrating tables and tools to the requirements of AS1012 3.3 Vebe
Verification of vibrating tables and tools to the requirements of AS1012 8.1, 8.2 and 8.3
Verification of vibrating tables and tools to the requirements of RMS (NSW) Method T304
Verification of internal vibrating tools to the requirements of AS1012 8.1, 8.2 and 8.3
Verification of internal vibrating tools to the requirements of RMS (NSW) Method T304

TfNSW TS 02800.05

Capability

with Calibration and Measurement Capability of -


4.5% of reading from 30 Hz to 40 Hz
3.5%of reading above 40 Hz to 50 Hz
2.7% of reading above 50 Hz to 150 Hz
1% of reading above 150 Hz to 300 Hz
7% of reading from 0.05 mm to 0.1 mm
3.6% of reading above 0.1 mm to 0.2 mm
1.9% of reading above 0.2 mm to 0.4 mm
1% of reading above 0.4 mm to 1 mm


Volume and density - Determination of density Standard nuclear gauge density blocks kg/m3 Gravimetric AS 1289.5.8.5

Capability


Calibration of standard density blocks 
with Calibration and Measurement Capability of-
0.02% of reading from 1000 kg/m3  to 3000 kg/m
Volume and density - Nuclear density gauges Nuclear moisture/density gauges Density; Moisture Comparison with a reference standard AS1289.5.8.4, AS/NZS 2891.14.4

Capability

Calibration of nuclear gauges using Type A density blocks for density and moisture in direct transmission and backscatter mode

with Calibration and Measurement Capability of -

By direct transmission 9 kg/m3 for density range  of 1727 kg/m to 2713 kg/m3

By backscatter mode 30 kg/mfor density in the range  of 1727 kg/m to 2713 kg/m3

9 kg/m3 for water content in the range of 0 to 588 kg/m3

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.