Enerven Energy Infrastructure Pty Ltd

Site

Calibration & Testing Services

Accreditation No.

179

Site No.

172

Date of Accreditation

16 Aug 1955

Address

41 - 55 Barnes Avenue
Marleston, SA 5033
Australia

enerven.com.au

Contact

Mr Khang Vu P: +61 (08) 82920121

Availability

Services available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

AC current

with Calibration and Measurement Capability of -

0.02% from 10 mA to 5 A at 50 Hz
0.1% from 20 A to 4000 A at 50 Hz
0.03% from 10 mA to 5 A and 50 Hz to 1 kHz
0.5% from 10 mA to 5 A and 1 kHz to 5 kHz
0.05% + 0.1 μA up to 20 A and 50 Hz to 1 kHz
1.0% + 0.1 μA up to 20 A and 1 kHz to 5 kHz

DC current

with Calibration and Measurement Capability of -

50 μA/A up to 5 A;
0.02% from 5 A to 50 A; 0.05% from 50 A to 500 A;
0.1% from 500 A to 1 000 A

AC voltage

with Calibration and Measurement Capability of-

0.005% from 1 V to 300 V and 50 Hz to 10 kHz
0.01% from 100 mV to 1000 V and 30 Hz to 10 kHz
0.01% from 1 V to 100 V and 10 Hz to 100 kHz
0.02% from 30 mV to 100 mV and 400 Hz to 10 kHz
0.02% from 100 mV to 1 V and 10 kHz to 50 kHz
0.02% from 300 mV to 1 V and 50 kHz to 200 kHz
0.02% from 1 V to 30 V and 100 kHz to 200 Khz
0.02% from 100 V to 300 V and 10 kHz to 100 kHz
0.02% from 300 V to 1000 V and 10 kHz to 50 kHz
0.05% from 30 mV to 1000 V and 10 kHz to 30 Hz
0.05% from 30 mV to 100 mV and 30 kHz to 400 Hz
0.05% from 30 to 100 mV and 10 to 200 kHz
0.05% from 10 to 30 mV and 400 Hz to 10 kHz
0.05% from 100 to 300 mV and 50 to 500 kHz
0.05% from 100 mV to 10 V and 200 to 500 kHz
0.1% from 10 to 30 mV and 10 Hz to 10 kHz
0.2% from 10 to 30 mV and 10 to 500 kHz
0.2% from 30 to 100 mV and 200 to 500 kHz
0.2% from 100 mV to 10 V and 500 kHz to 1 MHz
0.5% from 3 mV to 10 mV and 10 Hz to 200 kHz
0.5% from 10 mV to 100 mV and 500 kHz to 1 Mhz
1.0% from 3 mV to 10 mV and 200 kHz to 1 MHz
0.1% from 1 kV to 36 kV at 50 Hz

DC Voltage

with Calibration and Measurement Capability of-

3 μV/V + 0.2 μV up to 200 V
7 μV/V from 200 V to 1.1 kV
0.3% from 1.1 to 30 kV (40 kV measure only)

Resistance

with Calibration and Measurement Capability of -

1% or 0.1 μΩ (whichever is greater) up to 100 μΩ;
0.1% from 0.1 mΩ to 1 mΩ;
0.01% from 1 mΩ to 100 mΩ;
5 μΩ/Ω from 0.1Ω to 10 kΩ;
10 μΩ/Ω from 10 kΩ to 100 kΩ;
50 μΩ/Ω from 100 kΩ to 1 MΩ;
0.01% from 1 MΩ to 10 MΩ
0.1% from 10 MΩ to 100 MΩ
0.5% from 100 MΩ to 1 GΩ
0.5% from 1 MΩ to 10 GΩ 

DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Galvanometers and null detectors; Residual current circuit breaker testers AC current; DC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of-

DC ammeters including clamp-on meters

50 μA/A up to 5 A;
0.02% from 5 A to 50 A

0.05% from 50 A to 500 A;
0.1% from 500 A to 1 000 A

AC ammeters

0.02% from 10 mA to 5 A at 50 Hz
0.1% from 20 A to 4000 A at 50 Hz
0.03% from 10 mA to 5 A and 50 Hz to 1 kHz
0.5% from 10 mA to 5 A and 1 kHz to 5 kHz
0.05% + 0.1 μA up to 20 A and 50 Hz to 1 kHz
1.0% + 0.1 μA up to 20 A and 1 kHz to 5 kHz

Electricity meters; Varmeters; Watt meters Active power; Phase angle; Power factor; Reactive power Comparison with a reference standard Calibration and verification of single phase Class 0.2 and three phase Class 0.2 direct connected and transformer operated production watt-hour meters
including compliance with NMI NITP 14.

Capability

with Calibration and Measurement Capability of -
Watthour meters
(0.02/cos Φ)% from 50 to 500 V and 50 mA to 120 A at 50 Hz (single phase)
(0.05/cos Φ)% from 50 to 80 V and 0.1 to 10 A at 50 Hz (three phase)
(0.1/cos Φ)% from 100 to 250 V and 0.1 to 100 A at 50 Hz (three phase)
Varhour meters
(0.2/sin Φ)% from 50 to 80 V and 0.1 to 10 A at 50 Hz (three phase)

Ohm meters Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -
including megaohm meters up to 10 GΩ

1% or 0.1 μΩ (whichever is greater) up to 100 μΩ;
0.1% from 0.1 mΩ to 1 mΩ
0.01% from 1 mΩ to 100 mΩ
5 μΩ/Ω from 0.1Ω to 10 kΩ
10 μΩ/Ω from 10 to 100 kΩ
50 μΩ/Ω from 100 kΩ to 1 MΩ
0.01% from 1 MΩ to 10 MΩ
0.1% from 10 MΩ to 100 MΩ
0.5% from 100 MΩ to 1 GΩ
0.5% from 1 GΩ to 10 GΩ 

Phase angle meters Phase angle Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of-

0.1º at 50 Hz 

Power factor meters Power factor Comparison with a reference standard

Capability

with Calibration and Measurement Capability of-

0.002

Varmeters Reactive power Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of-

(0.1/sin Φ)% up to 65 V and 5 A at 50 Hz (single phase)
(0.1/sin Φ)% from 65 to 1 200 V and up to 120 A at 50 Hz (single phase)
(0.2/sin Φ)% up to 130 V and 5 A at 50 Hz (three phase)

Voltmeters AC voltage; DC voltage Comparison with a reference standard

Capability

DC voltmeters

with Calibration and Measurement Capability of-

3 μV/V + 0.2 μV up to 200 V
7 μV/V from 200 V to 1.1 kV
0.3% from 1.1 to 30 kV (40 kV measure only)

AC voltmeters

with Calibration and Measurement Capability of-

0.005% from 1 V to 300 V and 50 Hz to 10 kHz
0.01% from 100 mV to 1000 V and 30 Hz to 10 kHz
0.01% from 1 V to 100 V and 10 kHz to 100 kHz
0.02% from 30 mV to 100 mV and 400 Hz to 10 kHz
0.02% from 100 mV to 1 V and 10 kHz to 50 kHz
0.02% from 300 mV to 1 V and 50 kHz to 200 kHz
0.02% from 1 V to 30 V and 100 kHz to 200 Khz
0.02% from 100 V to 300 V and 10 kHz to 100 kHz
0.02% from 300 V to 1000 V and 10 kHz to 50 kHz
0.05% from 30 mV to 1000 V and 10 kHz to 30 Hz
0.05% from 30 mV to 100 mV and 30 Hz to 400 Hz
0.05% from 30 mV to 100 mV and 10 kHz to 200 kHz
0.05% from 10 mV to 30 mV and 400 Hz to 10 kHz
0.05% from 100 mV to 300 mV and 50 kHz to 500 kHz
0.05% from 100 mV to 10 V and 200 kHz to 500 kHz
0.1% from 10 mV to 30 mV and 10 Hz to 10 kHz
0.2% from 10 mV to 30 mV and 10 kHz to 500 kHz
0.2% from 30 mV to 100 mV and 200 kHz to 500 kHz
0.2% from 100 mV to 10 V and 500 kHz to 1 MHz
0.5% from 3 mV to 10 mV and 10 Hz to 200 kHz
0.5% from 10 mV to 100 mV and 500 kHz to 1 Mhz
1.0% from 3 mV to 10 mV and 200 kHz to 1 MHz
0.1% from 1 kV to 36 kV at 50 Hz

Watt meters Active power Comparison with a reference standard

Capability

with Calibration and Measurement Capability of-

0.02% up to 1 100 V and 10 A on dc
0.1% up to 1 100 V and 300 A on dc
(0.02/cos Φ)% up to 500 V and 120 A at 50 Hz
(0.05/cos Φ)% from 500 to 1 200 V and up to
120 A at 50 Hz (single phase)

DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

1% or 0.1 μΩ (whichever is greater) up to 100 μΩ;
0.1% from 0.1 mΩ to 1 mΩ
0.01% from 1 mΩ to 100 mΩ
5 μΩ/Ω from 0.1Ω to 10 kΩ
10 μΩ/Ω from 10 to 100 kΩ
50 μΩ/Ω from 100 kΩ to 1 MΩ
0.01% from 1 MΩ to 10 MΩ
0.1% from 10 MΩ to 100 MΩ
0.5% from 100 MΩ to 1 GΩ
0.5% from 1 GΩ to 10 GΩ 

Current shunts Current Direct measurement against a reference standard

Capability


with Calibration and Measurement Capability of -

DC shunts

0.05% up to 1 000 A

AC shunts

0.02% from 10 mA to 5 A at 50 Hz
0.1% from 20 A to 4000 A at 50 Hz
0.03% from 10 mA to 5 A and 50 Hz to 1 kHz
0.5% from 10 mA to 5 A and 1 kHz to 5 kHz
0.05% + 0.1 μA up to 20 A and 50 Hz to 1 kHz
1.0% + 0.1 μA up to 20 A and 1 kHz to 5 kHz

DC bridges; DC potentiometers; Ratiometers Current ratio; Resistance; Voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of-

DC potentiometers

5 μV/V or 0.2 μV (whichever is the greatest)

Ratiometers

0.05%

Inductors Inductance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

1.5% reading at frequencies up to 1 000 Hz from 1 mH to 10 H

Potential dividers; Voltage ratio boxes Voltage ratio Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of-

10 μV/V for applied voltages up to 1 100 V d.c.

Precision capacitors Capacitance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

Two terminal capacitors from 50 Hz to 1592 Hz

0.1% + 1 pF from 100 pF to 11 μF
0.5% from 11 μF to 100 μF


Three terminal capacitors from 50 Hz to 1592 Hz
0.1% from 1 pF to 11 μF;
0.5% from 11 μF to 100 μF


Loss tangent
0.1% + 0.0001

Voltage standards - E.M.F. reference devices; Voltage standards - Standard cells Voltage; Voltage ratio Direct measurement against a reference standard

Capability

Standard cells

with Calibration and Measurement Capability of -

3 μV

Electronic e.m.f.reference devices

with Calibration and Measurement Capability of -

3 μV/V at 1 V, 1.018 V and 10 V

Voltage Comparison with a reference standard

Capability

Standard cells

with Calibration and Measurement Capability of-

3 μV

E.M.F. reference devices

3 μV/V at 1 V, 1.018 V and 10 V

Magnetism - Magnetic instruments and equipment Flux density meters Magnetic flux density Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

3% of reading + 10 nT up to 2 mT r.m.s from DC to 400 Hz

Time and frequency metrology - Frequency, time and waveform measuring equipment Clocks and timers; Counters; Frequency meters; Stroboscopes; Tachometers; Time interval meters Frequency; Rotational speed; Time interval Comparison with a reference standard

Frequency meters


Capability

with Calibration and Measurement Capability of -

Frequency meters, counters up to 10 MHz

1 in 109

Time interval meters

1 in 105 + 1 count

Clocks and watches

5 in 106

Stroboscopes 

1 in 106

Tachometers

Optical tachometers

0.017 Hz from 0.17 Hz to 16.7 Hz (1 rpm from 10 to 1 000 rpm)

0.025 Hz above 16.7 Hz to 1 500 Hz (1.5 rpm from above 1 000 to 90 000 rpm)

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.