Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material In-house 02-414-30 and manufacturers specifications Calibration in the laboratory or in the field (field does not include Hâ‚‚S)
CH4 (Catalytic): 0.01-5%
CH4 (NDIR): 0.01-5%
CO (Electrochemical): 0.1-1000 ppm

CO2 (NDIR): 0.01-5%
H2S (Electrochemical): 0.1-200 ppm
O2 (Electrochemical): 0.1-30%

Capability

With Calibration and Measurement Capability of-
CH4 (Catalytic): 2.7% in the range 0.01-5%
CH4 (NDIR): 2.8% in the range 0.01-5%
CO (Electrochemical): 2.9% in the range 0.1-1000 ppm

CO2 (NDIR): 4.9% in the range 0.01-5%
H2S (Electrochemical): 4.2% in the range 0.1-200 ppm
O2 (Electrochemical): 2.7 % in the range 0.1-30%