Clean Air Control

Site

Clean Air Control

Accreditation No.

1650

Site No.

1643

Date of Accreditation

25 Oct 2017

Address

Unit 7, 1 Merino Entrance
Cockburn Central, WA 6164
Australia

cleanaircontrol.com.au

Contact

Mr Andrew Plackett P: +61 (08) 94174112

Availability

Services available to external clients

Scope of Accreditation

Clean Air Control

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Air barrier containment Not applicable AS 1807 clause 4.9 For compliance with AS 2252.2
Integrity of HEPA filter installations Not applicable

AS 1807.7, AS 1807 clause 4.4

For compliance with AS 2252.1 (2002) and AS 2252.2
Air velocity and uniformity of air velocity Not applicable AS 1807 clause 4.1 For compliance with AS 2252.2
Illuminance Not applicable AS 1807 clause 4.5 For compliance with AS 2252.1 (2002) and AS 2252.2
Work zone integrity Not applicable AS 1807 clause 4.3 For compliance with AS 2252.2
Sound level Not applicable AS 1807 clause 4.7 For compliance with AS 2252.1 (2002) and AS 2252.2
Inward air velocity Not applicable AS 1807 clause 4.8 For compliance with AS 2252.1 (2002)
Ultraviolet intensity Not applicable AS 1807 clause 4.10 For compliance with AS 2252.1 (2002) and 2252.2
Biological safety cabinets; Clean rooms, pharmaceutical isolators and work stations; Cytotoxic drug safety cabinets Particle count in work zone Automatic particle counter AS ISO 14644.1 Appendix A For compliance with AS 4273 (pharmaceutical isolators), AS 2252.5 (cytotoxic drug cabinets)

Clean rooms, pharmaceutical isolators and work stations Air velocity and uniformity of air velocity Not applicable AS 1807 clause 4.1 For compliance with AS 2252.6
Air pressure Not applicable AS ISO 14644.3 Method B.1 For compliance with AS 4273 (pharmaceutical isolators)
Illuminance Not applicable AS 1807 clause 4.5 For compliance with AS 2205.6
Integrity of HEPA filter installations Not applicable AS 1807.7, AS 1807 clause 4.4 For compliance with AS 2252.6
Clean workstations; Pharmaceutical isolators Sound level Not applicable AS 1807 clause 4.7 For compliance with AS 2252.6 (work stations) and AS 4273 (pharmaceutical isolators)
Cytotoxic drug safety cabinets Air velocity and uniformity of air velocity Not applicable AS 1807 clause 4.1 For compliance with AS 2252.5
Illuminance Not applicable AS 1807 clause 4.5 For compliance with 2252.5
Sound level Not applicable AS 1807 clause 4.7
Ultraviolet intensity Not applicable AS 1807 clause 4.10 For compliance with AS 2252.5
Integrity of HEPA filter installations Not applicable AS 1807 clause 4.4 For compliance with AS 2252.5
Air barrier containment Not applicable AS 1807 clause 4.9 For compliance with AS 2252.5
Integrity of HEPA filter installations Not applicable AS 1807.7
Work zone integrity Not applicable AS 1807 clause 4.3 For compliance with AS 2252.5
Fume cupboards and recirculating fume cabinets Sound level Not applicable AS 1807 clause 4.7, AS 1807.20 (2000) For compliance with AS/NZS 2243.8 and AS/NZS 2243.9
Illuminance Not applicable AS 1807 clause 4.5, AS 1807.15 (2000) For compliance with AS/NZS 2243.8 and AS/NZS 2243.9
Extraction performance Smoke visualisation AS/NZS 2243.8 Appendix A, AS/NZS 2243.9 Appendix A For compliance with AS/NZS 2243.8 and AS/NZS 2243.9
Extraction performance Face velocity

 AS/NZS 2243.8 Appendix B; AS/NZS 2243.9 Appendix B

For compliance with AS/NZS 2243.8 and AS/NZS 2243.9