National Association of Testing Authorities, Australia
Scope of Accreditation
Critical Certification (VIC) Pty Ltd
Site
Critical Certification (VIC) Pty Ltd
Accreditation No.
18235
Site No.
20049
Date of Accreditation
22 Mar 2011
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Air velocity and uniformity of air velocity | Cold DOP |
AS 1807.1 (2000) AS 1807 clause 4.1 |
For determining compliance with AS 2252.2 |
Inward air velocity | Cold DOP |
AS 1807.21 (2000) AS 1807 clause 4.8 |
For determining compliance with AS 2252.1 (2002) | ||
Air barrier containment | Cold DOP |
AS 1807.22 (2000) AS 1807 clause 4.9 |
For determining compliance with AS 2252.2 | ||
Illuminance | Not applicable |
AS 1807.15 (2000) AS 1807 clause 4.5 |
For determining compliance with: AS 2252.1 (2002) AS 2252.2 |
||
Integrity of HEPA filter installations | Cold DOP | AS 1807.7 (2000) |
For determining compliance with: AS 2252.1 (2002) AS 2252.2 |
||
Integrity of HEPA filter installations | Cold DOP |
AS 1807.6 (2000) AS 1807 clause 4.4 |
For determining compliance with: AS 2252.1 (2002) AS 2252.2 |
||
Ultraviolet intensity | Not applicable |
AS 1807.23 (2000) AS 1807 clause 4.10 |
For determining compliance with: AS 2252.1 (2002) AS 2252.2 |
||
Work zone integrity | Cold DOP |
AS 1807.5 (2000) AS 1807 clause 4.3 |
For determining compliance with AS 2252.2 | ||
Sound level | Not applicable |
AS 1807.20 (2000) AS 1807 clause 4.7 |
For determining compliance with: AS 2252.1 (2002) AS 2252.2 |
||
Clean rooms | Recovery times | Cold DOP |
AS 1807.24 (2000) AS/1SO 14644.3 Section B.4 |
For determining compliance with the performance requirements of AS 2252.6 (excluding clause 3.3.4), AS/NZS 2243.3 clause 5.4.4(k), AS 4273, AS ISO 14644.1 Annex B; 14644.3 Annex B clauses B4.2.2, B4.2.3, B4.2.4, B12.3.1 |
|
Clean rooms, pharmaceutical isolators and work stations | Ultraviolet intensity | Not applicable |
AS 1807.23 (2000) AS 1807 clause 4.10 |
For determining compliance with AS 2252.6 | |
Work zone integrity | Cold DOP |
AS 1807.5 (2000) AS 1807 clause 4.3 |
For determining compliance with AS 2252.6 | ||
Illuminance | Not applicable |
AS 1807.15 (2000) AS 1807 clause 4.5 |
For determining compliance with AS 2252.6 | ||
Sound level | Not applicable |
AS 1807.20 (2000) AS 1807 clause 4.7 |
For determining compliance with AS 2252.6 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.7 (2000) | For determining compliance with AS 2252.6 | ||
Particle count in work zone | Automatic particle counter |
AS 1807.8 (2000) AS ISO 14644 Part 1 Appendix A |
|||
Integrity of HEPA filter installations | Cold DOP |
AS 1807.6 (2000) AS 1807 clause 4.4 |
For determining compliance with AS 2252.6 | ||
Air pressure | Not applicable |
AS 1807.10 (2000) AS ISO 14644 Part 3 Method B.1 |
|||
Air velocity and uniformity of air velocity | Not applicable |
AS 1807.3 (2000) AS ISO 14644 Part 3 Method B.2 |
|||
Air velocity and uniformity of air velocity | Cold DOP |
AS 1807.1 (2000) AS 1807 clause 4.1 |
For determining compliance with AS 2252.6 | ||
Sound level | Not applicable | AS 1807.16 (2000) | |||
Cytotoxic drug safety cabinets | Integrity of HEPA filter installations | Cold DOP | AS 1807.7 (2000) | For determining compliance with AS 2252.5 | |
Work zone integrity | Cold DOP |
AS 1807.5 (2000) AS 1807 clause 4.3 |
For determining compliance with AS 2252.5 | ||
Air barrier containment | Cold DOP |
AS 1807.22 (2000) AS 1807 clause 4.9 |
For determining compliance with AS 2252.5 | ||
Ultraviolet intensity | Not applicable |
AS 1807.23 (2000) AS 1807 clause 4.10 |
For determining compliance with AS 2252.5 | ||
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | For determining compliance with AS 2252.5 | ||
Particle count in work zone | Automatic particle counter |
AS 1807.8 (2000) AS ISO 14644 Part 1 Appendix A |
For determining compliance with AS 2252.5 | ||
Air velocity and uniformity of air velocity | Cold DOP |
AS 1807.1 (2000) AS 1807 clause 4.1 |
For determining compliance with AS 2252.5 | ||
Illuminance | Not applicable |
AS 1807.15 (2000) AS 1807 clause 4.5 |
For determining compliance with AS 2252.5 | ||
Integrity of HEPA filter installations | Cold DOP |
AS 1807.6 (2000) AS 1807 clause 4.4 |
For determining compliance with AS 2252.5 | ||
Fume cupboards and recirculating fume cabinets | Sound level | Not applicable | AS 1807.20 (2000) AS 1807 clause 4.7 |
For determining compliance with: AS/NZS 2243.8 AS/NZS 2243.9 |
|
Extraction performance | Smoke visualisation |
AS/NZS 2243.9 Appendix A & B |
For determining compliance with the performance requirements of EN12469 |
||
Illuminance | Not applicable | AS 1807.15 (2000) AS 1807 clause 4.5 |
For determining compliance with: AS/NZS 2243.8 AS/NZS 2243.9 |
||
Extraction performance | Face velocity |
AS/NZS 2243.8 section Appendix B AS/NZS 2243.9 section Appendix B |
For determining compliance with the performance requirements of AS 2252.5 | ||
Extraction performance | Smoke visualisation |
AS/NZS 2243.8 Appendix A AS/NZS 2243.9 Appendix A & B |
For determining compliance with the performance requirements of AS 2252.1, AS 2252.2, NSF/ANSI 49 Annex F Methods F2, 3, 4, 5, 9 and 11 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
18235
Site No.
20049
Print date
29 Jun 2022