National Association of Testing Authorities, Australia
Scope of Accreditation
ECEFast
Site
Fastlab Calibration Laboratory
Accreditation No.
5473
Site No.
5466
Date of Accreditation
07 Nov 1994
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Pressure metrology - Pressure and vacuum measuring equipmentincluding on-site calibrations | Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges | Gauge pressure | By pressure calibrator; Comparison with reference instrument | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on-site calibrationswith Calibration and Measurement Capability of 0.5 kPa from - 95 kPa to below -1.25 kPa 1.3 Pa from -1.25 kPa to 1.25 kPa 0.5 kPa above 1.25 kPa to 700 kPa 0.8 kPa above 700 kPa to 2000 kPa 1.0 kPa above 2000 kPa to 4000 kPa | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Ancillary temperature measuring equipment | Process calibrators | Resistance to temperature conversion; Volt to temperature conversion | Direct measurement by electrical input | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrationsWith Calibration and Measurement Capability of- 0.2ºC + 0.01% of reading from -200ºC to 1300ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Type J extension wire; Type K extension wire; Type N extension wire; Type R extension wire; Type S extension wire; Type T extension wire | Volt to temperature conversion | Direct measurement by electrical input | 0ºC to 50ºC | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.2ºC + 0.1% T in the range from 0ºC to 50ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Humidity measuring equipment | Hygrometers; Psychrometers; Relative humidity sensors | Relative humidity (RH) | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 2.0%RH from 20°C to 50°C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Non-contact temperature measuring equipment | Radiation pyrometers; Thermal imaging systems | Temperature | Comparison with a reference standard; Direct measurement against a black body source | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 1.5ºC from -30ºC to 300ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature measuring equipment | Base metal thermocouples | Temperature | Measurement against reference standard |
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.5ºC + 0.1% T from -196ºC to below -40ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Bimetallic systems | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital temperature measuring systems; Temperature loggers - Single and multi-channel | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.25°C from -196°C to below -40°C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Resistance over oC range; mV over oC range | Direct temperature and electrical measurement against reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability In the range -196ºC to 1300ºCIncluding on site calibrations With Calibration and Measurement Capability of- 0.2ºC + 0.01% of reading from -196ºC to 1300ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Liquid in glass (LIG) thermometers | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.01ºC at 0ºC (Ice point only) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Metallic resistance thermometers | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.20°C from -80°C to below -20°C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Surface probes | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature standards and reference equipment | Dry block calibrators | Temperature | Direct measurement using a reference standard | Including the methods of - | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- 0.25°C from -90°C to -30°C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Rare metal thermocouples | mV per oC | Direct measurement using a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of-In the range from 0ºC to 1300ºC 0.2ºC + 0.1% T in the range from 0ºC to 1300ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Verification of controlled enclosuresincluding on site calibrations | Autoclaves and sterilising ovens; Baths; Cryogenic storage tanks; Environmental chambers - Temperature; Freezers; Furnaces; Incubators; Ovens | Spatial uniformity; Temperature; Temporal uniformity | Direct temperature measurement | AS 2853 - 1986 IEC 60068 (3-5) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and Measurement Capability of- In the range from -196ºC to 1300ºC 0.5ºC from -196ºC to -40ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Environmental chambers - Humidity | Humidity; Spatial uniformity | By temperature and humidity measurement | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and Measurement Capability of- In the range from 10% rh to 95% rh from 10ºC to 90ºC2.5% rh from 10% h to 95% rh from 10ºC to 90ºC | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Medical refrigeration equipment | Spatial uniformity; Temperature; Temporal uniformity | Direct temperature measurement | AS 3864.2 (excluding clause 3.6.4) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations For the following enclosure typesBlood fridges (upright, single & multi-door, under bench and walk-in rooms) ; Plasma freezers (ultra-low, upright, chest and walk-in rooms) With Calibration and Measurement Capability of- In the range from -196ºC to 20ºC 0.5ºC from -196ºC to -40ºC |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.