Site

Melbourne Laboratory

Accreditation No.

5473

Site No.

5466

Date of Accreditation

07 Nov 1994

Address

26 Business Park Drive
Notting Hill, VIC 3168
Australia

ecefast.com.au

Contact

Mr Phong Nguyen P: +61 (03) 95388169

Availability

Services available to external clients

Scope of Accreditation

Melbourne Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Pressure metrology - Pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges Gauge pressure By pressure calibrator; Comparison with reference instrument

Capability

including on-site calibrations
with Calibration and Measurement Capability of
0.5 kPa from - 95 kPa to below -1.25 kPa
1.3 Pa from -1.25 kPa to 1.25 kPa
0.5 kPa above 1.25 kPa to 700 kPa
0.8 kPa above 700 kPa to 2000 kPa
 1.0 kPa above 2000 kPa to 4000 kPa
Temperature metrology - Ancillary temperature measuring equipment Multi-channel thermocouple data recorders; Process calibrators Resistance to temperature conversion; Volt to temperature conversion Direct measurement by electrical input

Capability

including on site calibrations
with Calibration and Measurement Capability of-
 0.2 ºC + 0.01% of reading from -200 ºC to 1300 ºC
Temperature metrology - Ancillary temperature measuring equipment Type J extension wire; Type K extension wire; Type N extension wire; Type R extension wire; Type S extension wire; Type T extension wire Volt to temperature conversion Direct measurement by electrical input

Capability

with Calibration and Measurement Capability of-

0.2 ºC + 0.1% of reading from 0 ºC to 50 ºC
Temperature metrology - Humidity measuring equipment Hygrometers; Psychrometers; Relative humidity sensors Relative humidity (RH) Comparison with a reference standard In-house (Method 02-16)

Capability

with Calibration and Measurement Capability of-
In the range from 10 %rh to 95 %rh from 20 ºC to 90 ºC
2.0 %rh from 20 ºC to 50 ºC
2.8 %rh above 50 ºC to 90 ºC

Temperature metrology - Non-contact temperature measuring equipment Radiation pyrometers; Thermal imaging systems Temperature Comparison with a reference standard; Direct measurement against a black body source

Capability

with Calibration and Measurement Capability of

For temperatures from -30 °C to 900°C - emissivity 0.95 to 1
above 900 °C to 1200 °C, - emissivity 1 only.

1.5 ºC from -30 ºC to 300 ºC
2.5 ºC above 300 ºC to 500 ºC
3.5 ºC above 500 ºC to 1200 ºC

Temperature metrology - Temperature measuring equipment Base metal thermocouples Temperature Measurement against reference standard


 

Capability

with Calibration and Measurement Capability of-

0.5 ºC + 0.1% of reading  from -196 ºC to below -40 ºC
0.2 ºC + 0.1% of reading from-40 ºC to 650 ºC
0.2 ºC + 0.2% of reading above 650 ºC to 1300 ºC

Temperature metrology - Temperature measuring equipment Bimetallic systems Temperature Measurement against reference standard

Capability

with Calibration and Measurement Capability of-
0.5 ºC from 0 ºC to 250 ºC

Temperature metrology - Temperature measuring equipment Digital temperature measuring systems; Temperature loggers - Single and multi-channel Temperature Measurement against reference standard In house (Method 02-07, 02-08, 02-10, 02-05, 02-01)

Capability

including on site calibrations

with Calibration and Measurement Capability of-

0.25 °C from -196 °C to below -40 °C
0.005 °C at 0 °C (using RTD sensors)
0.05 °C from -40 °C to 300 °C
0.25 °C above 300 °C to 650 °C
3 °C above 650 °C to 1300 °C

Temperature metrology - Temperature measuring equipment Liquid in glass (LIG) thermometers Temperature Measurement against reference standard

Capability

including on site calibrations

with Calibration and Measurement Capability of-

0.01 ºC at 0 ºC (Ice point only)

Temperature metrology - Temperature measuring equipment Metallic resistance thermometers Temperature Measurement against reference standard In-house (Method 02-05)

Capability

including onsite calibrations

with Calibration and Measurement Capability of-
0.20 ºC at -196 ºC
0.20 ºC at -100 ºC
0.05 ºC above -100 ºC to below 0 ºC
0.01 ºC at 0 ºC
0.05 ºC above 0 ºC to 300 ºC
0.25 ºC above 300 ºC to 650ºC 

Temperature metrology - Temperature measuring equipment Surface probes Temperature Measurement against reference standard

Capability

with Calibration and Measurement Capability of-
2.9 ºC from 0 ºC to 650 ºC

Temperature metrology - Temperature standards and reference equipment Dry block calibrators Temperature Direct measurement using a reference standard In-house ( Method 02-11)

Capability

including on site calibrations

with Calibration and Measurement Capability of-

0.2 ºC from -90 ºC to below 140 ºC
0.5 ºC above 140 ºC to 650 ºC
2.5 ºC above 650 ºC to 1200 ºC

Temperature metrology - Temperature standards and reference equipment Rare metal thermocouples mV per oC Direct measurement using a reference standard

Capability

with Calibration and Measurement Capability of-
 0.2 ºC + 0.1% of reading from 0 ºC to 1300 ºC
Temperature metrology - Verification of controlled enclosures Autoclaves and sterilising ovens; Baths; Cryogenic storage tanks; Dry block heaters; Environmental chambers - Temperature; Freezers; Furnaces; Incubators; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement AS 2853 - 1986
IEC 60068 (3-5)

Capability

including on site calibrations

with Calibration and Measurement Capability of-

0.5 ºC from -196 ºC to -40 ºC
0.25 ºC above -40 ºC to 0 ºC
0.5 ºC above 0 ºC to 500 ºC
2.5 ºC above 500 ºC to 1300 ºC

Temperature metrology - Verification of controlled enclosures Environmental chambers - Humidity Humidity; Spatial uniformity By temperature and humidity measurement

Capability

including on site calibrations

with Calibration and Measurement Capability of-

2.5 %rh from 10 %rh to 95 %rh from 10 ºC to 90 ºC
Temperature metrology - Verification of controlled enclosures Medical refrigeration equipment Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement AS 3864.2 (excluding clause 3.6.4)

Capability

Including on site calibrations

For the following enclosure types
Blood fridges (upright, single & multi-door, under bench and walk-in rooms) ;
Plasma freezers (ultra-low, upright, chest and walk-in rooms)

with Calibration and Measurement Capability of-

0.5 ºC from -196 ºC to -40 ºC
0.25 ºC above -40 ºC to 0 ºC
0.5 ºC above 0 ºC to 20 ºC

The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

The Product column may also describe an Item or Material.

The Limitations column may also describe the Range.