National Association of Testing Authorities, Australia
Scope of Accreditation
Alcoa World Alumina Australia
Site
Flow Test Rig – Pinjarra Refinery
Accreditation No.
14117
Site No.
14110
Date of Accreditation
26 Oct 1998
Contact
Availability
Services not available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Flow metrology - Flow measuring devices and systems | Liquid meters | Flow rate | Comparison with a reference meter | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of- Calibration of flow meters from 150 mm to 350 mm using a master meter method 1 % of reading with a minimum 21000 L test volume and flow rates from 185 m³/h to 1000 m³/h using water as the transfer medium |