Scope of Accreditation
JST Biosafety Australia
Scope of Accreditation
JST Biosafety
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
| Service | Product | Determinant | Technique | Procedure | Limitations |
|---|---|---|---|---|---|
| Evaluation of air control equipment | Biological safety cabinets | Air flow; Air flow direction and visualization | Not applicable | NSF ANSI 49 N5.2, N5.3, N5.4 | for compliance with: NSF ANSI 49 |
| Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4, NSF ANSI 49 N5.5 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
| Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1, NSF ANSI 49 N5.2 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
| Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7, NSF ANSI 49 N5.11 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
| Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5, NSF ANSI 49 N5.9 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
| Recovery times | Not applicable | AS 1807.24 (2000), AS/NZS 14644.3 Annex B4 | for compliance with: AS 2252.1, AS 2252.2 | ||
| Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 | for compliance with: AS 2252.1, AS 2252.2 | ||
| Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.1, AS 2252.2 | ||
| Work zone integrity | Not applicable | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.1, AS 2252.2 | ||
| Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with: AS 2252.1, AS 2252.2 | ||
| Gas tightness | Not applicable | AS 1807.25 (2004), AS 1807 clause 4.11 | for compliance with: AS 2252.1, AS 2252.2 | ||
| Inward air velocity | Cold DOP | AS 1807.21 (2000), AS 1807 clause 4.8, NSF ANSI 49 N5.3 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
| Clean rooms | Air pressure | Not applicable | AS ISO 14644.3 Method B.1 | for compliance with: AS ISO 14644.3 | |
| Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS ISO 14644.3 | ||
| Sound level | Not applicable | AS 1807.16 | for compliance with: AS ISO 14644.3 | ||
| Recovery times | Not applicable | AS ISO 14644.3 Method B.4 | for compliance with: AS ISO 14644.3 | ||
| Air velocity and uniformity of air velocity | Cold DOP | AS ISO 14644.3 Methods B.2 and B.3 | for compliance with: AS ISO 14644.3 | ||
| Integrity of HEPA filter installations | Cold DOP | AS ISO 14644.3 Method B.7 | for compliance with: AS ISO 14644.3 | ||
| Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Appendix A | for compliance with: AS ISO 14644.1 | ||
| Clean workstations | Work zone integrity | Cold DOP | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.6 | |
| Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS 2252.6 | ||
| Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS 2252.6 | ||
| Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.6 | ||
| Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1 | for compliance with: AS 2252.6 | ||
| Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 | for compliance with: AS 2252.1, AS 2252.2, AS 2252.6 | ||
| Cytotoxic drug safety cabinets | Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 |
for compliance with: AS 2252.5 |
|
| Sound level | Not applicable | AS 1807.16 (2000) |
for compliance with: AS 2252.5 |
||
| Gas tightness | Not applicable | AS 1807.25 (2004), AS 1807 clause 4.11 |
for compliance with: AS 2252.5 |
||
| Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 |
for compliance with: AS 2252.5 |
||
| Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 |
for compliance with: AS 2252.5 |
||
| Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 |
for compliance with: AS 2252.1, AS 2252.2, AS 2252.5 |
||
| Particle count in work zone | Automatic particle counter |
AS 1807.8 (2000), AS ISO 14644.1 |
for compliance with: AS 2252.5 |
||
| Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1 |
for compliance with: AS 2252.5 |
||
| Air pressure | Not applicable | AS 1807.10 (2000), AS/NZS 14644.3 Annex B1 |
for compliance with: AS 2252.5 |
||
| Work zone integrity | Not applicable | AS 1807.5 (2000), AS 1807 clause 4.3 |
for compliance with: AS 2252.5 |
||
| Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 |
for compliance with: AS 2252.5 |
||
| Fume cupboards and recirculating fume cabinets | Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS/NZS 2243.8, AS 2243.9 | |
| Extraction performance | Smoke visualisation |
AS/NZS 2243.8 Appendix A AS 2243.9 Appendix A |
for compliance with: AS/NZS 2243.8, AS 2243.9 | ||
| Extraction performance | Face velocity |
AS/NZS 2243.8 Appendix B AS 2243.9 Appendix B |
for compliance with: AS/NZS 2243.8, AS 2243.9 | ||
| Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS/NZS 2243.8, AS 2243.9 | ||
| Pharmaceutical isolators | Air pressure | Not applicable | AS 1807.10 (2000), AS ISO 14644.3 Method B.1 | for compliance with: AS 4273 | |
| Recovery times | Not applicable | AS 1807.24 (2000), AS ISO 14644.3 Method B.4 | for compliance with: AS 4273 | ||
| Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 Appendix A | for compliance with: AS 4273 | ||
| Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1 | for compliance with: AS 4273 | ||
| Work zone integrity | Cold DOP | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 4273 | ||
| Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 | for compliance with: AS 4273 | ||
| Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS 4273 | ||
| Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS 4273 |
The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
The Product column may also describe an Item or Material.
The Limitations column may also describe the Range.
Accreditation No.
20802
Site No.
24959
Print date
23 Aug 2026
END OF SCOPE