JST Biosafety Australia

Site

JST Biosafety

Accreditation No.

20802

Site No.

24959

Date of Accreditation

04 Sep 2020

Address

88 Dorville Road
Carseldine, QLD 4034
Australia

Contact

Availability

Services available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Air barrier containment; Air flow; Air pressure; Air velocity and uniformity of air velocity; Gas tightness; Illuminance; Integrity of HEPA filter installations; Inward air velocity; Particle count in work zone; Recovery times; Relative humidity (RH); Sound level; Ultraviolet intensity; Work zone integrity Not applicable

AS 1807.1, 3, .5, .8, .10, .13, .15, .16, .20, .21, .22, .23, .24, .25

AS ISO 14644.1, AS/NZS 14644.3 Annex B Sections 4.2.2, 4.2.3, 5, 6.4 (leak test method only), B.7, B.8, B.9, B.12, and 13.2.2  
NSF/ANSI 49 Annex 5, Sections N-5.2 (all), N-5.3 (all), N-5.4 (all), N-5.5 (all), N-5.9, N-5.11
for the purpose of determining compliance with AS 2252.1, AS 2252.2 and NSF/ANSI 49
Clean rooms, pharmaceutical isolators and work stations Air barrier containment; Air pressure; Air velocity and uniformity of air velocity; Illuminance; Integrity of HEPA filter installations; Integrity of the refuge chamber under differential atmospheric pressure; Inward air velocity; Particle count in work zone; Recovery times; Relative humidity (RH); Sound level; Ultraviolet intensity; Work zone integrity Automatic particle counter; Cold DOP; Face velocity; Smoke visualisation

AS 1807.1, 3, .5, .6, .7, .8, .10, .13, .15, .16, .20, .21, .22, .23, .24

AS ISO 14644.1

AS NZS 14644.3 Annex B Sections 4.2.2, 4.2.3.5, 6.4 (leak test method only), B.7, B.8, B.9, B.12 (excluding Annex 12.3.2) and 13.2.2


for the purpose of determining compliance with AS 2252.6 and AS ISO 14644.1 Annex B
Cytotoxic drug safety cabinets Air barrier containment; Air change rate; Air pressure; Air velocity and uniformity of air velocity; Illuminance; Integrity of HEPA filter installations; Inward air velocity; Particle count in work zone; Relative humidity (RH); Sound level; Ultraviolet intensity; Work zone integrity Automatic particle counter; Cold DOP; Face velocity; Smoke visualisation AS 1807.1, 3, .5, .6, .7, .8, .10, .13, .15, .16, .20, .21, .22, .23, .25 for the purpose of determining compliance with AS 2252.5


Fume cupboards and recirculating fume cabinets Extraction performance Face velocity; Smoke visualisation

AS/NZS 2243.8 Appendices A and B

AS/NZS 2243.9 Appendices A and B

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.