National Association of Testing Authorities, Australia
Scope of Accreditation
JST Biosafety Australia
Scope of Accreditation
JST Biosafety
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5, NSF ANSI 49 N5.9 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 |
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7, NSF ANSI 49 N5.11 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4, NSF ANSI 49 N5.5 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
Air flow; Air flow direction and visualization | Not applicable | NSF ANSI 49 N5.2, N5.3, N5.4 | for compliance with: NSF ANSI 49 | ||
Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.1, AS 2252.2 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1, NSF ANSI 49 N5.2 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
Recovery times | Not applicable | AS 1807.24 (2000), AS/NZS 14644.3 Annex B4 | for compliance with: AS 2252.1, AS 2252.2 | ||
Work zone integrity | Not applicable | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.1, AS 2252.2 | ||
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 | for compliance with: AS 2252.1, AS 2252.2 | ||
Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with: AS 2252.1, AS 2252.2 | ||
Gas tightness | Not applicable | AS 1807.25 (2004), AS 1807 clause 4.11 | for compliance with: AS 2252.1, AS 2252.2 | ||
Inward air velocity | Cold DOP | AS 1807.21 (2000), AS 1807 clause 4.8, NSF ANSI 49 N5.3 | for compliance with: AS 2252.1, AS 2252.2, NSF ANSI 49 | ||
Clean rooms, pharmaceutical isolators and work stations | Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | |
Inward air velocity | Not applicable | AS 1807.21 (2000), AS 1807 clause 4.8 | for compliance with: AS 2252.6, AS ISO 14644.1 (Clean Room) | ||
Air barrier containment | Not applicable | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with: AS 2252.6, AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Sound level | Not applicable | AS 1807.7 (2000), AS 1807 clause 4.7 | for compliance with: AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Recovery times | Not applicable | AS 1807.24 (2000), AS/NZS 14644.3 Annex B4 | for compliance with: AS 2252.6, AS/NZS 14664.3 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.3 (2000), AS ISO 14644.3 clause 4.2.2, 4.2.3, Methods B.2 | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1 | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Work zone integrity | Cold DOP | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4, AS/NZS 14644.3 Clause 4.2.7, Method B7 | for compliance with: AS 2252.1, AS 2252.2, AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Sound level | Not applicable | AS 1807.16 (2000) | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room) | ||
Air pressure | Not applicable | AS 1807.10 (2000), AS/NZS 14644.3 Annex B1 | for compliance with: AS 2252.6 (Clean Workstation), AS/NZS 14644.3 (Clean Room), AS 4273 | ||
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 Appendix A | for compliance with: AS 2252.6 (Clean Workstation), AS ISO 14644.1 (Clean Room), AS 4273 | ||
Cytotoxic drug safety cabinets | Air pressure | Not applicable | AS 1807.10 (2000), AS/NZS 14644.3 Annex B1 | for compliance with: AS 2252.5 | |
Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 | for compliance with: AS 2252.1, AS 2252.2, AS 2252.5 | ||
Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with: AS 2252.5 | ||
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS 2252.5 | ||
Sound level | Not applicable | AS 1807.16 (2000) | for compliance with: AS 2252.5 | ||
Gas tightness | Not applicable | AS 1807.25 (2004), AS 1807 clause 4.11 | for compliance with: AS 2252.5 | ||
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 | for compliance with: AS 2252.5 | ||
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS 2252.5 | ||
Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.5 | ||
Work zone integrity | Not applicable | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.5 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1 | for compliance with: AS 2252.5 | ||
Fume cupboards and recirculating fume cabinets | Extraction performance | Smoke visualisation | AS/NZS 2243.8 Appendix A AS/NZS 2243.9 Appendix A | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | |
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Extraction performance | Face velocity | AS/NZS 2243.8 Appendix B AS/NZS 2243.9 Appendix B | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
20802
Site No.
24959
Print date
07 Dec 2024
END OF SCOPE