National Association of Testing Authorities, Australia
Scope of Accreditation
LAF Technologies Pty Ltd
Site
LAF Technologies Qld
Accreditation No.
14775
Site No.
18352
Date of Accreditation
29 Apr 2009
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.2 |
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 Appendix A | for compliance with: AS 2252.2 | ||
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS 2252.2 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), | for compliance with: AS 2252.2 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 | for compliance with: AS 2252.2 | ||
Work zone integrity | Not applicable | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.2 | ||
Inward air velocity | Cold DOP | AS 1807.21 (2000), AS 1807 clause 4.8 | for compliance with: AS 2252.2 | ||
Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with: AS 2252.2 | ||
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS 2252.2 | ||
Clean rooms | Recovery times | Not applicable | AS 1807.24 (2000), AS ISO 14644.3 Method B.4 | for compliance with: AS ISO 14644.1 | |
Sound level | Not applicable | AS 1807.16 | for compliance with: AS ISO 14644.1 | ||
Clean rooms, pharmaceutical isolators and work stations | Air pressure | Not applicable | AS 1807.10(2000), AS ISO 14644.3 Method B.1 | for compliance with: AS 2252.6, AS ISO 14644.1 | |
Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4, | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 Appendix A, AS/NZS 14644.3 Appendix B.1, B.2, B.3 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Work zone integrity | Cold DOP | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5, | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.3 (2000), AS ISO 14644.3 Method B.4 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Clean workstations; Pharmaceutical isolators | Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS 2252.6 | |
Cytotoxic drug safety cabinets | Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS 2252.5 | |
Air velocity and uniformity of air velocity | Cold DOP | AS 1807.1 (2000), AS 1807 clause 4.1 | for compliance with: AS 2252.5 | ||
Work zone integrity | Not applicable | AS 1807.5 (2000), AS 1807 clause 4.3 | for compliance with: AS 2252.2 | ||
Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS 2252.5 | ||
Ultraviolet intensity | Not applicable | AS 1807.23 (2000), AS 1807 clause 4.10 | for compliance with: AS 2252.5 | ||
Inward air velocity | Cold DOP | AS 1807.21 (2000), AS 1807 clause 4.8 | for compliance with: AS 2252.5 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807.6 (2000), AS 1807.7, AS 1807 clause 4.4 | for compliance with: AS 2252.5 | ||
Particle count in work zone | Automatic particle counter | AS 1807.8 (2000), AS ISO 14644.1 Appendix A | for compliance with: AS 2252.2 | ||
Air barrier containment | Cold DOP | AS 1807.22 (2000), AS 1807 clause 4.9 | for compliance with: AS 2252.5 | ||
Fume cupboards and recirculating fume cabinets | Illuminance | Not applicable | AS 1807.15 (2000), AS 1807 clause 4.5 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | |
Extraction performance | Smoke visualisation | AS/NZS 2243.8 Appendix A AS/NZS 2243.9 Appendix A | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Sound level | Not applicable | AS 1807.20 (2000), AS 1807 clause 4.7 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Extraction performance | Face velocity | AS/NZS 2243.8 Appendix B, AS/NZS 2243.9 Appendix B | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.