DC and low frequency electrical metrology - Electrical instrument calibrators | Instrument calibrators | Resistance | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability With Calibration and Measurement Capability of -
| Expanded Uncertainty / (µΩ/Ω) | ≤ 1 Ω to ≤ 10 kΩ | 1.0 | < 10 kΩ to ≤ 100 kΩ | 1.5 | < 100 kΩ to ≤ 1 MΩ | 2 | < 1 MΩ to ≤ 10 MΩ | 4.0 | < 10 MΩ to ≤ 100 MΩ | 10 | < 100 MΩ to ≤ 1 GΩ | 100 |
|
| | AC current; AC voltage; DC current; DC voltage | Comparison with a primary realisation; Comparison with a reference standard | |
|
Capability with Calibration and Measurement Capability of -
DC Voltage 0 V to 1100 V Measurements with Reference standard JVS presently not available
| Expanded Uncertainty | Voltage | Reference standard: Nanovoltmeter | Reference standard: Calibrator | Reference standard: JVS | 0 V to 10 nV | 0.1 µV | - | 0.05 µV | > 10 nV to 100 nV | 0.2 µV | - | 0.05 µV | > 100 nV to < 100 mV | 0.3 µV | - | 0.05 µV | 100 mV to 220 mV | 2.2 µV | 11 µV/V | 1.4 µV/V | 220 mV to 10 V | - | 2.9 µV/V | 0.8 µV/V |
| Expanded Uncertainty / (µV/V) | Voltage | Reference standard: Calibrator | Reference standard: JAVS | 10 V to 22 V | 2.3 | 1.1 | 22 V to 220 V | 2.1 | 1.2 | 220 V to 1100 V | 2.8 | 1.2 |
AC Voltage from 0.6 mV to 1000 V and 1 Hz to 1 MHz
| Expanded Uncertainty / (µV/V) |
| Voltage Measurement Standard | Frequency | 10 Hz to < 20 Hz | 20 Hz to < 30 Hz | 30 Hz to < 40 Hz | 40 Hz to < 57 Hz | 57 Hz to < 100 Hz | 100 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | > 1 kHz to 5 kHz | 0.6 mV to < 1 mV | 250 | 200 | 210 | 190 | 180 | 190 | 170 | 180 | 170 | 1 mV to < 2.2 mV | 120 | 90 | 90 | 110 | 90 | 110 | 90 | 80 | 90 | 2.2 mV to < 6 mV | 40 | 40 | 70 | 44 | 40 | 40 | 40 | 40 | 41 | 6 mV to < 22 mV | 59 | 50 | 38 | 81 | 39 | 52 | 42 | 38 | 38 | 22 mV to < 60 mV | 44 | 34 | 31 | 27 | 27 | 29 | 32 | 27 | 27 | 60 mV to < 220 mV | 130 | 110 | 98 | 97 | 96 | 96 | 96 | 96 | 96 | 0.22 V to < 0.6 V | 46 | 29 | 20 | 13 | 12 | 11 | 11 | 11 | 15 | 0.6 V to 1 V | 33 | 20 | 13 | 9 | 10 | 10 | 8 | 8 | 9 | > 1 V to 6 V | 33 | 21 | 13 | 10 | 10 | 10 | 10 | 10 | 10 | > 6V to 22 V | 33 | 21 | 14 | 13 | 12 | 13 | 13 | 12 | 12 | > 22 V to 60 V | 41 | 24 | 20 | 18 | 17 | 18 | 17 | 20 | 19 | > 60 V to 100 V | 36 | 23 | 18 | 18 | 18 | 18 | 18 | 17 | 18 | > 100 V to 250 V | 35 | 23 | 18 | 18 | 18 | 18 | 18 | 18 | 18 | > 250 V to 300 V | - | - | - | 20 | 20 | 20 | 20 | 20 | 20 | > 300 V to 750 V | - | - | - | 20 | 20 | 20 | 20 | 20 | 24 | > 750 V to 1000 V | - | - | - | 21 | 21 | 21 | 21 | 21 | 26 |
| Expanded Uncertainty / (µV/V) |
| Voltage Measurement Standard | Frequency | > 5 kHz to 10 kHz | > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | > 30 kHz to 50 kHz | > 50 kHz to 100 kHz | > 100 kHz to 200 kHz | > 200 kHz to 300 kHz | > 300 kHz to 500 kHz | > 500 kHz to 800 kHz | > 800 kHz to 1 MHz | 0.6 mV to < 1 mV | 170 | 160 | 150 | 250 | 690 | 1100 | 1300 | 1400 | 1900 | 2000 | 1 mV to < 2.2 mV
| 80 | 80 | 90 | 120 | 140 | 310 | 310 | 490 | 710 | 690 | 2.2 mV to < 6 mV | 40 | 40 | 40 | 41 | 53 | 110 | 140 | 230 | 270 | 300 | 6 mV to < 22 mV | 38 | 40 | 38 | 45 | 47 | 69 | 84 | 150 | 210 | 230 | 22 mV to < 60 mV | 27 | 34 | 34 | 34 | 44 | 54 | 67 | 89 | 160 | 180 | 60 mV to < 220 mV | 98 | 98 | 98 | 99 | 110 | 110 | 120 | 140 | 200 | 220 | 0.22 V to < 0.6 V | 13 | 14 | 18 | 20 | 22 | 26 | 33 | 58 | 82 | 110 | 0.6 V to 1 V | 9 | 10 | 12 | 14 | 18 | 22 | 27 | 41 | 74 | 99 | > 1 V to 6 V | 10 | 12 | 12 | 14 | 19 | 26 | 31 | 49 | 95 | 140 | > 6 V to 22 V | 12 | 14 | 14 | 16 | 24 | 32 | 34 | 51 | 110 | 160 | > 22 V to 60 V | 19 | 30 | 26 | 43 | 35 | 43 | 54 | - | - | - | > 60 V to 100 V | 18 | 18 | 18 | 21 | 28 | 42 | - | - | - | - | > 100 V to 250 V | 18 | 18 | 18 | 21 | 29 | - | - | - | - | - | > 250 V to 300 V | 20 | 20 | 23 | 23 | 33 | - | - | - | - | - | > 300 V to 750 V | 24 | 24 | 24 | 34 | 66 | - | - | - | - | - | > 750 V to 1000 V | 26 | 26 | 32 | - | - | - | - | - | - | - |
DC Current from 0 A up to 10 A
| Expanded Uncertainty
| 0 A to 50 µA | 1.5 nA | 50 µA to < 100 µA | 30 µA/A | 100 µA to < 1 mA | 14 µA/A | 1 mA to < 10 mA | 12 µA/A | 10 mA to < 100 mA | 7 µA/A | 100 mA to < 1A | 13 µA/A | 1 A to 10 A | 12 µA/A |
AC Current from 10 μA to 100 A and 10 Hz to 30 kHz
| Expanded Uncertainty / (µA/A) |
| 10 Hz to < 57 Hz | 57 Hz | > 57 Hz to 400 Hz | > 400 Hz to 1 kHz | > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | > 10 kHz to 30 kHz | 10 µA to < 0.1 mA | 230 | 90 | 90 | 92 | 140 | 600 | - | 0.1 mA to < 1 mA | 120 | 45 | 45 | 47 | 67 | 300 | 620 | 1 mA to < 10 mA | 120 | 44 | 44 | 46 | 63 | 300 | 150 | 10 mA to < 100 mA | 120 | 44 | 44 | 45 | 63 | 300 | 170 | 100 mA | 65 | 26 | 26 | 26 | 40 | 180 | 130 | > 100 mA to 1 A | 100 | 39 | 56 | 200 | 250 | 300 | - | > 1 A to 10 A | 200 | 78 | 120 | 400 | - | - | - | > 10 A to 20 A | 250 | 100 | 150 | 500 | - | - | - | > 20 A to 100 A
| - | 40 | - | - | - | - | - |
|
DC and low frequency electrical metrology - Electrical measurement and test equipment | Ammeters; Digital multimeters (DMM); Electricity meters; Ohm meters; Power factor meters; Varmeters; Voltmeters; Watt meters | AC current; AC voltage; Apparent power; DC current; DC voltage; Power factor; Resistance | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability with Calibration and Measurement Capability of - DC Voltage from 1 mV up to 1100 V Measurements with Reference standard JVS presently available only at voltages from 1 mV to 10 V
| Expanded Uncertainty | Voltage | Reference standard: Calibrator | Reference standard: JVS | 0 V to 1 mV
| 5 μV
| 0.5 μV
| 1 mV to 10 mV | 500 μV/V | 50 μV/V | 10 mV to 100 mV | 50 μV/V | 8.6 μV/V | 100 mV to 1 V | 8 μV/V | 0.5 μV/V | 1 V to 10 V | 3 μV/V | 0.1 μV/V | 10 V to 100 V | 2 μV/V | 0.1 μV/V | 100 V to 1000 V | 5 μV/V | 1.0 μV/V |
AC Voltage from 3 mV to 1000 V and 1 Hz to 1 MHz For Digital Multimeter
| Expanded Uncertainty / (μV/V) |
| Digital Sampling Voltmeter | Digital Sampling Voltmeter | Voltage Measurement Standard | Frequency | 1 Hz to < 5 Hz | 5 Hz to < 10 Hz | 10 Hz to < 20 Hz | 20 Hz to < 30 Hz | 30 Hz to < 40 Hz | 40 Hz to < 57 Hz | 57 Hz to < 100 Hz | 100 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | 3 mV to < 10 mV | 60 | 30 | 790 | 690 | 670 | 670 | 670 | 620 | 620 | 620 | 10 mV to < 30 mV | 60 | 20 | 670 | 560 | 540 | 530 | 530 | 460 | 460 | 460 | 30 mV to < 100 mV | 60 | 20 | 440 | 430 | 420 | 420 | 420 | 380 | 380 | 380 | 100 mV to < 0.3 V | 60 | 20 | 130 | 100 | 80 | 80 | 80 | 80 | 80 | 80 | 0.3 V to < 1 V | 60 | 20 | 100 | 70 | 50 | 50 | 50 | 50 | 50 | 50 | 1 V to 3 V | 60 | 20 | 80 | 60 | 40 | 40 | 40 | 40 | 40 | 40 | > 3 V to 10 V | 60 | 20 | 80 | 60 | 40 | 40 | 40 | 40 | 40 | 40 | > 10 V to 30 V | 60 | 20 | 80 | 60 | 40 | 40 | 40 | 40 | 40 | 40 | > 30 V to 100 V | 60 | 20 | 80 | 60 | 40 | 40 | 40 | 40 | 40 | 40 | > 100 V to 700 V | 60 | 20 | 80 | 60 | 60 | 60 | 40 | 40 | 40 | 40 | > 700 V to 1000 V | - | - | 100 | 80 | 70 | 70 | 70 | 70 | 70 | 70 |
| Expanded Uncertainty / (μV/V)
|
| Voltage Measurement Standard
| Frequency | > 1 kHz to 5 kHz | > 5 kHz to 10 kHz
| > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | > 30 kHz to 50 kHz | > 50 kHz to 100 kHz | > 100 kHz to 200 kHz | > 200 kHz to 300 kHz | > 300 kHz to 500 kHz | > 500 kHz to 1 MHz | 3 mV to < 10 mV
| 620 | 620 | 620 | 620 | 720 | 750 | 850 | 1700 | 2400 | 5000 | 10 mV to < 30 mV | 460 | 460 | 470 | 480 | 600 | 640 | 750 | 1600 | 2400 | 5000 | 30 mV to < 100 mV | 380 | 380 | 380 | 390 | 470 | 500 | 620 | 1100 | 1200 | 2500 | 100 mV to < 0.3 V | 80 | 80 | 90 | 100 | 190 | 270 | 450 | 480 | 670 | 1500 | 0.3 V to < 1 V | 50 | 50 | 60 | 80 | 110 | 210 | 420 | 450 | 650 | 1400 | 1 V to 3 V | 40 | 50 | 60 | 70 | 110 | 210 | 410 | 450 | 650 | 1400 | > 3 V to 10 V | 40 | 50 | 60 | 70 | 110 | 210 | 410 | 450 | 650 | 1400 | > 10 V to 30 V | 40 | 50 | 60 | 70 | 110 | 210 | 410 | 450 | 650 | 1400 | > 30 V to 100 V | 40 | 50 | 60 | 70 | 110 | 210 | 410 | 450 | 640 | - | > 100 V to 700 V | 40 | 50 | 60 | 80 | 110 | 210 | 410 | 450 | - | - | > 700 V to 1000 V | 70 | 70 | 80 | 130 | 160 | 290 | - | - | - | - |
DC Current (including Picoameters) from 10 pA up to 20 A
| Expanded Uncertainty / (μA/A) | 10 pA
| 1400
| 100 pA
| 700
| 1 nA
| 530
| 10 nA
| 530
| 100 nA
| 530
| 1 μA
| 210
| 10 μA
| 210
| 100 μA
| 120
| 100 μA to < 1 mA | 14 | 1 mA to < 10 mA | 12 | 10 mA to 100 mA | 7 | > 100 mA to 1 A | 13 | > 1 A to 10 A | 12 | > 10 A to 20 A | 26 |
AC Current from 10 μA to 20 A and 10 Hz to 30 kHz
| Expanded Uncertainty / (μA/A) |
| 10 Hz to < 57 Hz | 57 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | 10 μA to < 50 μA | 740 | 430 | 450 | 440 | - | - | - | - | 50 μA to < 100 μA | 470 | 140 | 140 | 140 | - | - | - | - | 0.1 mA to < 1 mA | 120 | 59 | 59 | 69 | 88 | 110 | 340 | 620 | 1 mA to < 10 mA | 120 | 59 | 59 | 59 | 73 | 87 | 100 | 150 | 10 mA to < 100 mA | 120 | 58 | 58 | 58 | 62 | 71 | 120 | 170 | 100 mA | 86 | 29 | 29 | 29 | 29 | 45 | 83 | 130 | > 100 mA to 1 A | 68 | 30 | 30 | 30 | 45 | 83 | 120 | 200 | > 1 A to 10 A | 110 | 72 | 72 | 95 | 300 | 500 | 700 | - | > 10 A to 20 A | 170 | 110 | 110 | 150 | 450 | 750 | 1100 | - |
Ohmmeters from 0 mΩ to 1 GΩ
| Expanded Uncertainty | 0 mΩ to <100 mΩ
| 0.2 μΩ
| 100 mΩ to < 10 Ω | 2 µΩ/Ω | 10 Ω to 10 kΩ | 1 µΩ/Ω | > 10 kΩ to 100 kΩ | 2 µΩ/Ω | > 100 kΩ to 1 MΩ | 3 µΩ/Ω | > 1 MΩ to 100 MΩ | 10 µΩ/Ω | > 100 MΩ to 1 GΩ | 100 µΩ/Ω |
Wattmeters and Electricity Meters (sinewave from 1 V to 1000 V and 0.01 A to 200 A and 47 Hz to 60 Hz) Mains Frequency Power
| Expanded Uncertainty / (μW/(VA)) | Voltage | Frequency | Power Factor | 0.01 A to 0.1 A | 20 | 1 V to 1000 V | 47 Hz to 60 Hz | 1 to 0 L/C | > 0.1 A to 100 A > 100 A to 200 A | 10 30 | 1 V to 1000 V 1 V to 1000 V | 47 Hz to 60 Hz 47 Hz to 60 Hz | 1 to 0 L/C 1 to 0 L/C |
Power with Distorted Voltage or Current Distorted voltage or current from 60 V to 240 V and 0.001 A to 10 A and 47 Hz to 60 Hz Fundamental frequency and voltage or current harmonics up to 200 kHz
|
| Expanded Uncertainty / (μW/(VA)) | Voltage | Frequency | Voltage or Current Harmonics maximum frequency / kHz | Power Factor |
| 0.001 A to 0.01 A | 100 | 60 V to 240 V | 47 Hz to 60 Hz | 200 | 1 to 0 L/C |
| > 0.01 A to 10 A
| 50 | 60 V to 240 V | 47 Hz to 60 Hz
| 200 | 1 to 0 L/C |
Varmeters 0.004% of Apparent Power from 60 V to 240 V and 0.1A to 120 A at 47 Hz to 60 Hz
Power Factor Meters 0.00005 power factor from 60 V to 480 V and 0.1 A to 100 A and 47 Hz to 60 Hz |
| Multi-purpose AC measurement and transfer standard | AC current; AC voltage; DC current; DC voltage | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability with Calibration and Measurement Capability of -
DC Voltage from 0 mV up to 1100 V Measurements with Reference standard JVS presently available only at voltages from 1 mV to 10 V
| Expanded Uncertainty | Voltage | Reference standard: Calibrator | Reference standard: JVS | 0 V to <1 mV
| 0.5 μV
| 0.05 μV
| 1 mV to 10 mV | 500 μV/V | 50 μV/V | 10 mV to 100 mV | 50 μV/V | 8.6 μV/V | 100 mV to 1 V | 8 μV/V | 0.5 μV/V | 1 V to 10 V | 3 μV/V | 0.1 μV/V | 10 V to 100 V | 2 μV/V | 0.1 μV/V | 100 V to 1000 V | 5 μV/V | 1.0 μV/V |
AC Voltage from 1 mV to 1000 V and 10 Hz to 1 MHz For AC Measurement Standard
| Expanded Uncertainty / (μV/V) | Frequency | 10 Hz to < 20 Hz | 20 Hz to < 30 Hz | 30 Hz to < 40 Hz | 40 Hz to < 57 Hz | 57 Hz to < 100 Hz | 100 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | 1 mV to < 2 mV | 200 | 150 | 120 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 2 mV to < 6 mV | 130 | 100 | 70 | 70 | 50 | 50 | 50 | 50 | 50 | 50 | 6 mV to < 20 mV | 90 | 70 | 45 | 35 | 35 | 35 | 35 | 35 | 35 | 35 | 20 mV to < 60 mV | 70 | 50 | 40 | 35 | 35 | 35 | 35 | 35 | 35 | 35 | 60 mV to < 200 mV | 65 | 45 | 35 | 35 | 35 | 35 | 35 | 35 | 35 | 35 | 200 mV to < 600 mV | 60 | 36 | 30 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 600 mV to < 2 V
| 10 | 9 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 2V
| 9 | 8 | 7 | 7 | 7 | 7 | 7 | 7 | 7 | 7 | > 2 V to 6 V
| 11 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | > 6 V to 20 V
| 11 | 10 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | > 20 V to 60 V
| 15 | 12 | 11 | 11 | 11 | 11 | 11 | 11 | 11 | 11 | > 60 V to 200 V
| 20 | 15 | 15 | 15 | 15 | 15 | 15 | 15 | 15 | 15 | > 200 V to 600 V | 20 | 18 | 17 | 17 | 17 | 17 | 17 | 17 | 21 | 21 | > 600 V to 1000 V | 41 | 27 | 23 | 23 | 23 | 23 | 23 | 23 | 23 | 25 |
| Expanded Uncertainty / (μV/V) | Frequency | > 10 kHz to 20 kHz | > 20 kHz to 30 kHz
| > 30 kHz to 50 kHz
| > 50 kHz to 100 kHz
| > 100 kHz to 200 kHz
| > 200 kHz to 300 kHz
| > 300 kHz to 500 kHz
| > 500 kHz to 800 kHz
| > 800 kHz to 1 MHz
| 1 mV to < 2 mV | 100 | 100 | 100 | 200 | 800 | 1100 | 1200 | 1500 | 1600 | 2 mV to < 6 mV | 50 | 50 | 50 | 130 | 180 | 200 | 220 | 550 | 800 | 6 mV to < 20 mV | 35 | 35 | 35 | 70 | 80 | 110 | 200 | 400 | 600 | 20 mV to < 60 mV | 35 | 35 | 35 | 55 | 65 | 90 | 140 | 260 | 300 | 60 mV to < 200 mV | 35 | 35 | 35 | 50 | 60 | 80 | 130 | 200 | 240 | 200 mV to < 600 mV | 24 | 28 | 32 | 36 | 44 | 70 | 130 | 190 | 220 | 600 mV to < 2 V
| 8 | 8 | 8 | 12 | 16 | 26 | 34 | 40 | 50 | 2V
| 7 | 7 | 7 | 10 | 14 | 22 | 29 | 34 | 43 | > 2 V to 6 V
| 9 | 9 | 9 | 14 | 18 | 27 | 39 | 50 | 60 | > 6 V to 20 V
| 9 | 11 | 12 | 15 | 23 | 26 | 54 | - | - | > 20 V to 60 V
| 11 | 12 | 14 | 17 | - | - | - | - | - | > 60 V to 200 V
| 38 | 38 | 38 | 38 | - | - | - | - | - | > 200 V to 600 V | 22 | 27 | 35 | 45 | - | - | - | - | - | > 600 V to 1000 V | 28 | 33 | 40 | 57 | - | - | - | - | - |
DC Current (including Picoameters) from 10 pA up to 20 A
| Expanded Uncertainty / (μA/A) | 10 pA
| 1400
| 100 pA
| 700
| 1 nA
| 530
| 10 nA
| 530
| 100 nA
| 530
| 1 μA
| 210
| 10 μA
| 210
| 100 μA
| 120
| 100 μA to < 1 mA | 14 | 1 mA to < 10 mA | 12 | 10 mA to 100 mA | 7 | > 100 mA to 1 A | 13 | > 1 A to 10 A | 12 | > 10 A to 20 A | 26 |
AC Current from 10 μA to 20 A and 10 Hz to 30 kHz
| Expanded Uncertainty / (μA/A) |
| 10 Hz to < 57 Hz | 57 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | 10 μA to < 50 μA | 740 | 430 | 450 | 440 | - | - | - | - | 50 μA to < 100 μA | 470 | 140 | 140 | 140 | - | - | - | - | 0.1 mA to < 1 mA | 120 | 59 | 59 | 69 | 88 | 110 | 340 | 620 | 1 mA to < 10 mA | 120 | 59 | 59 | 59 | 73 | 87 | 100 | 150 | 10 mA to < 100 mA | 120 | 58 | 58 | 58 | 62 | 71 | 120 | 170 | 100 mA | 86 | 29 | 29 | 29 | 29 | 45 | 83 | 130 | > 100 mA to 1 A | 68 | 30 | 30 | 30 | 45 | 83 | 120 | 200 | > 1 A to 10 A | 110 | 72 | 72 | 95 | 300 | 500 | 700 | - | > 10 A to 20 A | 170 | 110 | 110 | 150 | 450 | 750 | 1100 | - |
|
DC and low frequency electrical metrology - Electrical standards | AC/DC transfer instruments | Current AC-DC difference | Comparison with a reference standard | | |
Capability from 0.0001 A to 100 A and 10 Hz to 200 kHz with Calibration and Measurement Capability of -
| Expanded Uncertainty / (μA/A) |
| 10 Hz to < 20 Hz | 20 Hz to < 30 Hz | 30 Hz to < 40 Hz | 40 Hz to < 57 Hz | 57 Hz to < 100 Hz | 100 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | 0.1 mA to < 1 mA | 99 | 52 | 40 | 36 | 36 | 36 | 44 | 51 | 1 mA to < 10 mA | 99 | 52 | 39 | 35 | 34 | 34 | 39 | 41 | 10 mA to < 30 mA | 6 | 5 | 5 | 4 | 4 | 4 | 4 | 4 | 30 mA to < 50 mA | 8 | 6 | 5 | 4 | 4 | 4 | 4 | 4 | 50 mA to < 100 mA | 9 | 7 | 5 | 4 | 4 | 4 | 4 | 4 | 100 mA | 9 | 7 | 6 | 5 | 4 | 4 | 4 | 4 | > 100 mA to 400 mA | 10 | 8 | 6 | 5 | 5 | 5 | 4 | 4 | > 400 mA to 1.5 A | 11 | 8 | 6 | 5 | 5 | 5 | 5 | 5 | > 1.5 A to 5 A | 11 | 9 | 7 | 5 | 5 | 5 | 5 | 5 | > 5 A to 20 A | 14 | 12 | 11 | 10 | 10 | 10 | 10 | 10 | > 20 A to 50 A | 19 | 16 | 15 | 14 | 14 | 14 | 14 | 13 | > 50 A to 100 A | 22 | 18 | 17 | 16 | 16 | 15 | 14 | 14 |
| Expanded Uncertainty / (μA/A)
|
| > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | > 30 kHz to 50 kHz | > 50 kHz to 100 kHz | > 100 kHz to 200 kHz | 0.1 mA to < 1 mA | 73 | 120 | - | - | - | - | - | 1 mA to < 10 mA | 49 | 65 | - | - | - | - | - | 10 mA to < 30 mA | 4 | 4 | 5 | 6 | 7 | 8 | 10 | 30 mA to < 50 mA | 4 | 5 | 5 | 6 | 7 | 9 | 11 | 50 mA to < 100 mA | 4 | 5 | 6 | 7 | 8 | 10 | 13 | 100 mA | 4 | 5 | 6 | 7 | 9 | 12 | 17 | > 100 mA to 400 mA | 5 | 6 | 6 | 8 | 10 | 12 | 18 | > 400 mA to 1.5 A | 5 | 6 | 7 | 8 | 10 | 13 | 19 | > 1.5 A to 5 A | 5 | 6 | 7 | 9 | 13 | 21 | - | > 5 A to 20 A | 10 | 14 | 16 | 21 | 26 | 32 | - | > 20 A to 50 A | 16 | 21 | 29 | 37 | 46 | 61 | - | > 50 A to 100 A | 18 | 26 | 38 | 52 | 65 | 84 | - |
|
| | Voltage AC-DC difference | Comparison with a reference standard | | |
Capability From 1 mV to 1000 V and 10 Hz to 1 MHz with Calibration and Measurement Capability of -
| Expanded Uncertainty / (μV/V) |
| 10 Hz to < 20 Hz | 20 Hz to < 30 Hz | 30 Hz to < 40 Hz | 40 Hz to < 57 Hz | 57 Hz to < 100 Hz | 100 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | 1 mV to < 2 mV | 260 | 220 | 200 | 170 | 170 | 170 | 170 | 170 | 180 | 180 | 2 mV to < 5 mV | 110 | 85 | 77 | 77 | 77 | 77 | 77 | 77 | 78 | 78 | 5 mV to < 10 mV | 77 | 58 | 53 | 53 | 53 | 53 | 53 | 53 | 54 | 54 | 10 mV to < 20 mV | 53 | 41 | 38 | 38 | 38 | 38 | 38 | 38 | 40 | 40 | 20 mV to < 30 mV | 46 | 33 | 31 | 31 | 31 | 31 | 31 | 31 | 33 | 33 | 30 mV to < 50 mV | 38 | 28 | 24 | 24 | 23 | 23 | 23 | 24 | 24 | 24 | 50 mV to < 100 mV | 30 | 20 | 16 | 14 | 14 | 14 | 14 | 14 | 15 | 16 | 100 mV to < 200 mV | 28 | 20 | 16 | 13 | 13 | 13 | 13 | 13 | 15 | 15 | 200 mV to < 300 mV | 28 | 20 | 16 | 13 | 12 | 12 | 12 | 12 | 14 | 14 | 300 mV to < 0.5 V | 28 | 20 | 16 | 13 | 12 | 12 | 12 | 12 | 14 | 14 | 0.5 V to < 1 V | 4.8 | 2.5 | 1.6 | 1.3 | 1.2 | 1.2 | 1.2 | 1.2 | 1.4 | 1.4 | 1 V to < 2 V | 5.2 | 2.6 | 1.7 | 1.3 | 1.3 | 1.2 | 1.2 | 1.2 | 1.4 | 1.4 | 2 V to < 3 V | 5.1 | 2.6 | 1.6 | 1.3 | 1.3 | 1.2 | 1.2 | 1.2 | 1.4 | 1.4 | 3 V to 4 V | 5.2 | 2.6 | 1.7 | 1.3 | 1.3 | 1.2 | 1.2 | 1.2 | 1.4 | 1.4 | > 4 V to 6 V | 7 | 5 | 5 | 5 | 5 | 5 | 5 | 4 | 4 | 4 | > 6 V to 10 V | 7 | 6 | 6 | 6 | 6 | 6 | 6 | 5 | 5 | 5 | > 10 V to 12 V | 8 | 7 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | > 12 V to 20 V | 8 | 7 | 7 | 7 | 7 | 7 | 7 | 6 | 6 | 6 | > 20 V to 30 V | 10 | 9 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | 8 | > 30 V to 40 V | 10 | 9 | 9 | 9 | 9 | 9 | 9 | 8 | 8 | 8 | > 40 V to 60 V | 11 | 10 | 9 | 9 | 9 | 9 | 10 | 9 | 9 | 9 | > 60 V to 100 V | 11 | 10 | 10 | 10 | 10 | 10 | 10 | 9 | 9 | 10 | > 100 V to 120 V | 12 | 11 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | > 120 V to 200 V | 12 | 11 | 11 | 11 | 11 | 11 | 11 | 10 | 10 | 11 | > 200 V to 300 V | 14 | 12 | 12 | 12 | 12 | 12 | 12 | 11 | 11 | 12 | > 300 V to 400 V | 14 | 13 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | > 400 V to 600 V | 15 | 14 | 14 | 14 | 14 | 14 | 14 | 13 | 13 | 16 | > 600 V to 1000 V | 17 | 16 | 16 | 16 | 16 | 16 | 16 | 15 | 16 | 18 |
| Expanded Uncertainty / (μV/V)
|
| > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | > 30 kHz to 50 kHz | > 50 kHz to 100 kHz | > 100 kHz to 200 kHz | > 200 kHz to 300 kHz | > 300 kHz to 500 kHz | > 500 kHz to 800 kHz | > 800 kHz to 1 MHz | 1 mV to 2 mV | 180 | 180 | 180 | 190 | 220 | 270 | 310 | 360 | 420 | 2 mV to < 5 mV | 78 | 79 | 80 | 88 | 100 | 160 | 220 | 280 | 310 | 5 mV to < 10 mV | 54 | 56 | 57 | 68 | 75 | 120 | 180 | 220 | 260 | 10 mV to < 20 mV | 40 | 41 | 43 | 56 | 63 | 70 | 110 | 170 | 200 | 20 mV to < 30 mV | 33 | 33 | 36 | 45 | 53 | 66 | 89 | 130 | 150 | 30 mV to < 50 mV | 25 | 25 | 26 | 31 | 41 | 57 | 80 | 140 | 150 | 50 mV to < 100 mV | 16 | 17 | 20 | 26 | 37 | 52 | 71 | 120 | 140 | 100 mV to < 200 mV | 15 | 16 | 20 | 26 | 38 | 54 | 73 | 129 | 140 | 200 mV to < 300 mV | 15 | 16 | 19 | 25 | 37 | 51 | 67 | 110 | 120 | 300 mV to < 0.5 V | 15 | 16 | 19 | 25 | 37 | 51 | 67 | 110 | 120 | 0.5 V to < 1 V | 1.6 | 2.2 | 3.0 | 5.2 | 6.8 | 8.3 | 10.6 | 13.8 | 17.6 | 1 V to < 2 V | 1.6 | 2.2 | 3.0 | 5.2 | 6.8 | 8.3 | 10.6 | 13.8 | 17.6 | 2 V to < 3 V | 1.6 | 2.1 | 2.8 | 5.0 | 6.8 | 9.2 | 11.6 | 15.8 | 19.5 | 3 V to 4 V | 1.6 | 2.1 | 2.8 | 5.0 | 6.8 | 9.2 | 11.6 | 15.8 | 19.5 | > 4 V to 6 V | 4 | 5 | 5 | 8 | 11 | 15 | 23 | 30 | 37 | > 6 V to 10 V | 5 | 5 | 6 | 9 | 12 | 17 | 25 | 32 | 39 | > 10 V to 12 V | 6 | 6 | 6 | 9 | 13 | 18 | 26 | 33 | 40 | > 12 V to 20 V | 6 | 7 | 7 | 10 | 14 | 19 | 27 | 34 | 42 | > 20 V to 30 V | 9 | 11 | 13 | 17 | 22 | 28 | 37 | - | - | > 30 V to 40 V | 9 | 11 | 14 | 18 | 23 | 30 | 39 | - | - | > 40 V to 60 V | 10 | 11 | 14 | 19 | 25 | 31 | - | - | - | > 60 V to 100 V | 10 | 12 | 15 | 20 | 26 | - | - | - | - | > 100 V to 120 V | 11 | 12 | 15 | 21 | - | - | - | - | - | > 120 V to 200 V | 11 | 13 | 16 | 21 | - | - | - | - | - | > 200 V to 300 V | 12 | 14 | 17 | 23 | - | - | - | - | - | > 300 V to 400 V | 12 | 15 | 18 | 25 | - | - | - | - | - | > 400 V to 600 V | 17 | 22 | 28 | 39 | - | - | - | - | - | > 600 V to 1000 V | 20 | 26 | 36 | 49 | - | - | - | - | - |
|
| Current shunts | Phase displacement | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability -0.1 rad to 0.1 rad Current 0.01 A to 100 A Frequency 40 Hz to 200 kHz with Calibration and Measurement Capability of -
| Expanded Uncertainty / (μrad) |
| 40 Hz to 1 kHz | 1 kHz to 10 kHz | > 10 kHz to 20 kHz | > 20 kHz to 50 kHz | > 50 kHz to 100 kHz | > 100 kHz to 200 kHz | 0.01 A to 0.5 A | 3 | 8 | 16 | 41 | 81 | 200 | > 0.5 A to 1 A | 4 | 10 | 19 | 47 | 95 | 230 | > 1 A to 2 A | 4 | 11 | 21 | 53 | 110 | 250 | > 2 A to 5 A | 5 | 12 | 23 | 58 | 120 | 280 | > 5 A to 10 A | 6 | 13 | 25 | 63 | 130 | 300 | > 10 A to 20 A | 6 | 16 | 30 | 75 | 150 |
| > 20 A to 40 A | 6 | 18 | 35 | 86 | 170 |
| > 40 A to 100 A | 7 | 20 | 39 | 96 | 190 |
|
|
| | Current AC-DC difference; Resistance; Voltage AC-DC difference | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability Resistance 0.0001 Ω to 1000 Ω Current 0.1 mA to 100 A Frequency 10 Hz to 200 kHz with Calibration and Measurement Capability of -
| Expanded Uncertainty / (μΩ/Ω) |
| 10 Hz to < 20 Hz | 20 Hz to < 30 Hz | 30 Hz to < 40 Hz | 40 Hz to < 57 Hz | 57 Hz to < 100 Hz | 100 Hz to < 400 Hz | 400 Hz to < 1 kHz | 1 kHz | 0.1 mA to < 1 mA | 99 | 52 | 40 | 36 | 36 | 36 | 44 | 51 | 1 mA to < 10 mA | 99 | 52 | 39 | 35 | 34 | 34 | 39 | 41 | 10 mA to < 30 mA | 6 | 5 | 5 | 4 | 4 | 4 | 4 | 4 | 30 mA to < 50 mA | 8 | 6 | 5 | 4 | 4 | 4 | 4 | 4 | 50 mA to < 100 mA | 9 | 7 | 5 | 4 | 4 | 4 | 4 | 4 | 100 mA | 9 | 7 | 6 | 5 | 4 | 4 | 4 | 4 | > 100 mA to 400 mA | 10 | 8 | 6 | 5 | 5 | 5 | 4 | 4 | > 400 mA to 1.5 A | 11 | 8 | 6 | 5 | 5 | 5 | 5 | 5 | > 1.5 A to 5 A | 11 | 9 | 7 | 5 | 5 | 5 | 5 | 5 | > 5 A to 20 A | 14 | 12 | 11 | 10 | 10 | 10 | 10 | 10 | > 20 A to 50 A | 19 | 16 | 15 | 14 | 14 | 14 | 14 | 13 | > 50 A to 100 A | 22 | 18 | 17 | 16 | 16 | 15 | 14 | 14 |
| Expanded Uncertainty / (μΩ/Ω) |
| > 1 kHz to 5 kHz | > 5 kHz to 10 kHz | > 10 kHz to 20 kHz | > 20 kHz to 30 kHz | > 30 kHz to 50 kHz | > 50 kHz to 100 kHz | > 100 kHz to 200 kHz | 0.1 mA to < 1 mA | 73 | 120 | - | - | - | - | - | 1 mA to < 10 mA | 49 | 65 | - | - | - | - | - | 10 mA to < 30 mA | 4 | 4 | 5 | 6 | 7 | 8 | 10 | 30 mA to < 50 mA | 4 | 5 | 5 | 6 | 7 | 9 | 11 | 50 mA to < 100 mA | 4 | 5 | 6 | 7 | 8 | 10 | 13 | 100 mA | 4 | 5 | 6 | 7 | 9 | 12 | 17 | > 100 mA to 400 mA | 5 | 6 | 6 | 8 | 10 | 12 | 18 | > 400 mA to 1.5 A | 5 | 6 | 7 | 8 | 10 | 13 | 19 | > 1.5 A to 5 A | 5 | 6 | 7 | 9 | 13 | 21 | - | > 5 A to 20 A | 10 | 14 | 16 | 21 | 26 | 32 | - | > 20 A to 50 A | 16 | 21 | 29 | 37 | 46 | 61 | - | > 50 A to 100 A | 18 | 26 | 38 | 52 | 65 | 84 | - |
|
| Potential dividers; Voltage ratio boxes | Voltage ratio | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability with Calibration and Measurement Capability of - 0.05 μV/V for voltage ratios from 0.01 to 0.1 at voltages from 10 V to 100 V 0.1 μV/V for voltage ratios from 0.01 to 0.1 at voltages between 100 V and 1000 V 0.2 μV/V for voltage ratios from 0.001 to 0.01 at voltages between 10 V and 1000 V |
| Voltage standards - E.M.F. reference devices | Voltage | Comparison with a primary realisation; Comparison with a reference standard | | |
Capability at 1 V, 1.018 V and 10 V with Calibration and Measurement Capability of -
| Expanded Uncertainty / (nV/V) | 1 V | 40 | 1.018 V | 40 | 10 V | 20 |
|
Magnetism - Magnetic instruments and equipment | Flux density meters | Magnetic flux density | Direct measurement against a reference standard | | |
Capability with Calibration and Measurement Capability of - 0 to 2 T on DC 0.02 µT + 1.5 % from 0 to 0.005 T at 50 Hz and 60 Hz (AC) Transverse Probes, Reference Magnets (DC) with Calibration and Measurement Capability of - Magnetic Flux Density (T) | Expanded Uncertainty | 0 to < 0.05 0.05 to 2 | 0.1 μT + 0.5 % 0.15 % |
Axial Probes (DC) with Calibration and Measurement Capability of -
Magnetic Flux Density (T)
| Expanded Uncertainty
| 0 to 0.095 | 0.1 μT + 0.5 % |
|
| Flux meters | Magnetic flux | Direct measurement against a reference standard | | |
Capability with Calibration and Measurement Capability of - 1% from 10 μWb to 10 mWb |
| Helmholtz coils; Magnets; Solenoids | Magnetic field strength | Direct measurement against a reference standard | | |
Capability with Calibration and Measurement Capability of - 0.1 µT + 0.3 % from 0 T to 2 T |
Time and frequency metrology - Frequency, time and waveform measuring equipment | Clocks and timers; Counters; Distortion analysers; Electrical power calibrators; Frequency analysers; Frequency meters; Oscilloscopes; Power analysers; Time interval meters | Attenuation; Current; Current harmonics; Distortion; Frequency; Phase angle; Voltage; Voltage flicker; Voltage harmonics | Measurement against reference standard | | |
Capability with Calibration and Measurement Capability of -
Distortion Voltage and current harmonic amplitude and total harmonic distortion up to 99ᵗʰ harmonic, for harmonic amplitudes of 5 % to 40 % of fundamental voltage or current from 0.01 V to 240 V, 0.005 A to 20 A and 47 Hz to 63 Hz
Fundamental Voltage / Current | Harmonic Number | Expanded Uncertainty (% of fundamental) | Reference Standard | 0.01 V to 240 V | 1 to 39 | 0.001 | Josephson Arbitraty Wavefom Synthesizer | 1 V to 240 V
| 1 to 99
| 0.005
| Digital Sampling Voltmeter
| 0.005 A to 20 A
| 1 to 99 | 0.005 | Digital Sampling Voltmeter |
Voltage Harmonics (Meter and Source) up to 39ᵗʰ, phase, from 1 V to 240 V and 47 Hz to 63 Hz Distortion | Expanded Uncertainty (degrees) | Harmonic Number | 1 - 9 | 11 - 19 | 21 - 39 | 5 % to < 10 % | 0.0016 | 0.0021 | 0.0099 | 10 % to < 20 % | 0.0014 | 0.0019 | 0.0097 | 20 % to < 30 % | 0.0012 | 0.0018 | 0.0095 | 30 % to < 40 % | 0.0011 | 0.0018 | 0.0094 | Current Harmonics (Meter and Source) up to 39ᵗʰ, phase, from 0.005 A to 20 A and 47 Hz to 63 HzDistortion | Expanded Uncertainty (degrees)
| Harmonic Number
| 1 - 9 | 11 - 19 | 21 - 39 | 5 % to < 10 % | 0.0026 | 0.0031 | 0.0109 | 10 % to < 20 % | 0.0024 | 0.0029 | 0.0107 | 20 % to < 30 % | 0.0022 | 0.0028 | 0.0105 | 30 % to < 40 % | 0.0021 | 0.0028 | 0.0104 | Flicker Meters 0.01% of fundamental for Voltage Flicker from 1 V to 240 V and 47 Hz to 53 HzPhasor Measurement Units 0.007 % for Total Vector Error from 1 V to 240 V and 47 Hz to 53 Hz 0.007 % for Total Vector Error, 0.005 A to 20 A and 47 Hz to 53 Hz 0.0022 º for phase angle from 0 º to 90 º, 1 V to 240 V, 0.005 A to 20 A and 47 Hz to 53 Hz 0.0034 % for frequency from 47 Hz to 53 Hz, 1 V to 240 V 0.0016 Hz/s for rate of change of frequency from -5 Hz/s to 5 Hz/s, 47 Hz to 53 Hz and 1 V to 240 V Phase Sources, Phase Standards, Power Calibrators from -180 º to 180 º, 0.01 V to 1000 V, 0.005 A to 200 A and 40 Hz to 1000 Hz
Frequency | Expanded Uncertainty | (Hz) | (Degrees) | 40 to 60 | 0.0005 | > 60 to 400 | 0.001 | > 400 to 1000 | 0.002 |
Phase Meters, Power Analysers
Frequency | Expanded Uncertainty | (Hz) | (Degrees) | 40 to 60 | 0.0005 | > 60 to 400 | 0.001 | > 400 to 1000 | 0.002 |
|