National Measurement Institute (NMI)

Site

Low Frequency Standards, Lindfield

Accreditation No.

1

Site No.

14538

Date of Accreditation

03 Jul 2001

Address

36 Bradfield Road
Lindfield West, NSW 2070
Australia

measurement.gov.au

Contact

Ms Charmaine Kwan P: +61 (02) 84673836

Availability

Services conditionally available to external clients

Scope of Accreditation

Low Frequency Standards, Lindfield

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators Resistance Comparison with a primary realisation; Comparison with a reference standard

Capability

With Calibration and Measurement Capability of -


 Expanded Uncertainty / (µΩ/Ω)

≤ 1 Ω to ≤ 10 kΩ

1.0

< 10 kΩ to ≤ 100 kΩ

1.5

< 100 kΩ to ≤ 1 MΩ

2

< 1 MΩ to ≤ 10 MΩ

4.0

< 10 MΩ to ≤ 100 MΩ

10

< 100 MΩ to ≤ 1 GΩ

100


AC current; AC voltage; DC current; DC voltage Comparison with a primary realisation; Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

DC Voltage 0 V to 1100 V
Measurements with Reference standard JVS presently not available


Expanded Uncertainty
VoltageReference standard: NanovoltmeterReference standard: CalibratorReference standard: JVS
0 V to 10 nV0.1 µV-0.05 µV
> 10 nV to 100 nV0.2 µV-0.05 µV
> 100 nV to < 100 mV0.3 µV-0.05 µV
100 mV to 220 mV2.2 µV11 µV/V1.4 µV/V
220 mV to 10 V-2.9 µV/V0.8 µV/V



Expanded Uncertainty / (µV/V)
VoltageReference standard: CalibratorReference standard: JAVS
10 V to 22 V2.31.1
22 V to 220 V2.11.2
220 V to 1100 V2.81.2

AC Voltage from 0.6 mV to 1000 V and 1 Hz to 1 MHz

Expanded Uncertainty / (µV/V)

Voltage Measurement Standard
Frequency10 Hz to
< 20 Hz
20 Hz to
< 30 Hz
30 Hz to
< 40 Hz
40 Hz to
< 57 Hz
57 Hz to
< 100 Hz
100 Hz to
< 400 Hz
400 Hz to
< 1 kHz
1 kHz> 1 kHz to
5 kHz
0.6 mV to < 1 mV250200210190180190170180170
1 mV to < 2.2 mV120909011090110908090
2.2 mV to < 6 mV404070444040404041
6 mV to < 22 mV595038813952423838
22 mV to < 60 mV443431272729322727
60 mV to < 220 mV13011098979696969696
0.22 V to < 0.6 V462920131211111115
0.6 V to 1 V33201391010889
> 1 V to 6 V332113101010101010
> 6V to 22 V332114131213131212
> 22 V to 60 V412420181718172019
> 60 V to 100 V362318181818181718
> 100 V to 250 V352318181818181818
> 250 V to 300 V---202020202020
> 300 V to 750 V---202020202024
> 750 V to 1000 V---212121212126


Expanded Uncertainty / (µV/V)

Voltage Measurement Standard
Frequency> 5 kHz to
10 kHz
> 10 kHz to
20 kHz
> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz> 200 kHz to 300 kHz> 300 kHz to 500 kHz> 500 kHz to 800 kHz> 800 kHz to 1 MHz
0.6 mV to < 1 mV17016015025069011001300140019002000
1 mV to < 2.2 mV
808090120140310310490710690
2.2 mV to < 6 mV4040404153110140230270300
6 mV to < 22 mV38403845476984150210230
22 mV to < 60 mV2734343444546789160180
60 mV to < 220 mV98989899110110120140200220
0.22 V to < 0.6 V131418202226335882110
0.6 V to 1 V9101214182227417499
> 1 V to 6 V101212141926314995140
> 6 V to 22 V1214141624323451110160
> 22 V to 60 V19302643354354---
> 60 V to 100 V181818212842----
> 100 V to 250 V1818182129-----
> 250 V to 300 V2020232333-----
> 300 V to 750 V2424243466-----
> 750 V to 1000 V262632-------

DC Current from 0 A up to 10 A

Expanded Uncertainty
0 A to 50 µA1.5 nA
50 µA to < 100 µA30 µA/A
100 µA to < 1 mA14 µA/A
1 mA to < 10 mA12 µA/A
10 mA to < 100 mA7 µA/A
100 mA to < 1A13 µA/A
1 A to 10 A12 µA/A

AC Current from 10 μA to 100 A and 10 Hz to 30 kHz

Expanded Uncertainty / (µA/A)

10 Hz to < 57 Hz57 Hz> 57 Hz to 400 Hz> 400 Hz to 1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 30 kHz
10 µA to < 0.1 mA230909092140600-
0.1 mA to < 1 mA12045454767300620
1 mA to < 10 mA12044444663300150
10 mA to < 100 mA12044444563300170
100 mA6526262640180130
> 100 mA to 1 A1003956200250300-
> 1 A to 10 A20078120400---
> 10 A to 20 A250100150500---
> 20 A to 100 A
-40-----

DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Digital multimeters (DMM); Electricity meters; Ohm meters; Power factor meters; Varmeters; Voltmeters; Watt meters AC current; AC voltage; Apparent power; DC current; DC voltage; Power factor; Resistance Comparison with a primary realisation; Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
DC Voltage from 1 mV up to 1100 V

Measurements with Reference standard JVS presently available only at voltages from 1 mV to 10 V

Expanded Uncertainty 
VoltageReference standard: CalibratorReference standard: JVS
0 V to 1 mV
5 μV
0.5 μV
1 mV to 10 mV500 μV/V50 μV/V
10 mV to 100 mV50 μV/V8.6 μV/V
100 mV to 1 V8 μV/V0.5 μV/V
1 V to 10 V3 μV/V0.1 μV/V
10 V to 100 V2 μV/V0.1 μV/V
100 V to 1000 V5 μV/V1.0 μV/V


AC Voltage from 3 mV to 1000 V and 1 Hz to 1 MHz

For Digital Multimeter

 Expanded Uncertainty / (μV/V)

Digital
Sampling
Voltmeter
Digital
Sampling
Voltmeter
Voltage Measurement Standard
Frequency1 Hz to < 5 Hz5 Hz to < 10 Hz10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz
3 mV to < 10 mV6030790690670670670620620620
10 mV to < 30 mV6020670560540530530460460460
30 mV to < 100 mV6020440430420420420380380380
100 mV to < 0.3 V6020130100808080808080
0.3 V to < 1 V602010070505050505050
1 V to 3 V60208060404040404040
> 3 V to 10 V60208060404040404040
> 10 V to 30 V60208060404040404040
> 30 V to 100 V60208060404040404040
> 100 V to 700 V60208060606040404040
> 700 V to 1000 V--10080707070707070


Expanded Uncertainty / (μV/V)

Voltage Measurement Standard
Frequency> 1 kHz to 5 kHz> 5 kHz to 10 kHz
> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz> 200 kHz to 300 kHz> 300 kHz to 500 kHz> 500 kHz to 1 MHz
3 mV to < 10 mV
620620620620720750850170024005000
10 mV to < 30 mV460460470480600640750160024005000
30 mV to < 100 mV380380380390470500620110012002500
100 mV to < 0.3 V8080901001902704504806701500
0.3 V to < 1 V505060801102104204506501400
1 V to 3 V405060701102104104506501400
> 3 V to 10 V405060701102104104506501400
> 10 V to 30 V405060701102104104506501400
> 30 V to 100 V40506070110210410450640-
> 100 V to 700 V40506080110210410450--
> 700 V to 1000 V707080130160290----

 DC Current (including Picoameters) from 10 pA up to 20 A


Expanded Uncertainty / (μA/A)
10 pA
1400
100 pA
700
1 nA
530
10 nA
530
100 nA
530
1 μA
210
10 μA
210
100 μA
120
100 μA to < 1 mA14
1 mA to < 10 mA12
10 mA to 100 mA7
> 100 mA to 1 A13
> 1 A to 10 A12
> 10 A to 20 A26

AC Current from 10 μA to 20 A and 10 Hz to 30 kHz

Expanded Uncertainty / (μA/A)

10 Hz to < 57 Hz57 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz
10 μA to < 50 μA740430450440----
50 μA to < 100 μA470140140140----
0.1 mA to < 1 mA12059596988110340620
1 mA to < 10 mA1205959597387100150
10 mA to < 100 mA1205858586271120170
100 mA86292929294583130
> 100 mA to 1 A683030304583120200
> 1 A to 10 A110727295300500700-
> 10 A to 20 A1701101101504507501100-

Ohmmeters from 0 mΩ to 1 GΩ

Expanded Uncertainty 
0 mΩ to <100 mΩ
0.2 μΩ
100 mΩ to < 10 Ω2 µΩ/Ω
10 Ω to 10 kΩ1 µΩ/Ω
> 10 kΩ to 100 kΩ2 µΩ/Ω
> 100 kΩ to 1 MΩ3 µΩ/Ω
> 1 MΩ to 100 MΩ10 µΩ/Ω
> 100 MΩ to 1 GΩ100 µΩ/Ω

Wattmeters and Electricity Meters (sinewave from 1 V to 1000 V and 0.01 A to 200 A and 47 Hz to 60 Hz)
Mains Frequency Power

Expanded Uncertainty / (μW/(VA))VoltageFrequencyPower Factor
0.01 A to 0.1 A201 V to 1000 V47 Hz to 60 Hz1 to 0 L/C
> 0.1 A to 100 A
> 100 A to 200 A
10
30
1 V to 1000 V
1 V to 1000 V
47 Hz to 60 Hz
47 Hz to 60 Hz
1 to 0 L/C
1 to 0 L/C

Power with Distorted Voltage or Current
Distorted voltage or current from 60 V to 240 V and 0.001 A to 10 A and 47 Hz to 60 Hz
Fundamental frequency and voltage or current harmonics up to 200 kHz


Expanded Uncertainty / (μW/(VA))VoltageFrequencyVoltage or Current Harmonics
maximum frequency / kHz
Power Factor

0.001 A to 0.01 A10060 V to 240 V47 Hz to 60 Hz2001 to 0 L/C

> 0.01 A to 10 A
5060 V to 240 V47 Hz to 60 Hz
2001 to 0 L/C

Varmeters
0.004% of Apparent Power from 60 V to 240 V and 0.1A to 120 A at 47 Hz to 60 Hz

Power Factor Meters
0.00005 power factor from 60 V to 480 V and 0.1 A to 100 A and 47 Hz to 60 Hz
Multi-purpose AC measurement and transfer standard AC current; AC voltage; DC current; DC voltage Comparison with a primary realisation; Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

DC Voltage from 0 mV up to 1100 V

Measurements with Reference standard JVS presently available only at voltages from 1 mV to 10 V

Expanded Uncertainty 
VoltageReference standard: CalibratorReference standard: JVS
0 V to <1 mV
0.5 μV
0.05 μV
1 mV to 10 mV500 μV/V50 μV/V
10 mV to 100 mV50 μV/V8.6 μV/V
100 mV to 1 V8 μV/V0.5 μV/V
1 V to 10 V3 μV/V0.1 μV/V
10 V to 100 V2 μV/V0.1 μV/V
100 V to 1000 V5 μV/V1.0 μV/V


AC Voltage from 1 mV to 1000 V and 10 Hz to 1 MHz

For AC Measurement Standard


Expanded Uncertainty / (μV/V)
Frequency10 Hz to
< 20 Hz
20 Hz to
< 30 Hz
30 Hz to
< 40 Hz
40 Hz to
< 57 Hz
57 Hz to
< 100 Hz
100 Hz to
< 400 Hz
400 Hz to
< 1 kHz
1 kHz> 1 kHz
to 5 kHz
> 5 kHz
to 10 kHz
1 mV to < 2 mV200150120100100100100100100100
2 mV to < 6 mV1301007070505050505050
6 mV to < 20 mV90704535353535353535
20 mV to < 60 mV70504035353535353535
60 mV to < 200 mV65453535353535353535
200 mV to < 600 mV60363020202020202020
600 mV to < 2 V
10988888888
2V
9877777777
> 2 V to 6 V
11999999999
> 6 V to 20 V
111099999999
> 20 V to 60 V
15121111111111111111
> 60 V to 200 V
20151515151515151515
> 200 V to 600 V20181717171717172121
> 600 V to 1000 V41272323232323232325

 


Expanded Uncertainty / (μV/V)
Frequency> 10 kHz to
20 kHz
> 20 kHz to
30 kHz
> 30 kHz to
50 kHz
> 50 kHz to
100 kHz
> 100 kHz to
200 kHz
> 200 kHz to
300 kHz
> 300 kHz to
500 kHz
> 500 kHz to
800 kHz
> 800 kHz to
1 MHz
1 mV to < 2 mV1001001002008001100120015001600
2 mV to < 6 mV505050130180200220550800
6 mV to < 20 mV3535357080110200400600
20 mV to < 60 mV353535556590140260300
60 mV to < 200 mV353535506080130200240
200 mV to < 600 mV242832364470130190220
600 mV to < 2 V
888121626344050
2V
777101422293443
> 2 V to 6 V
999141827395060
> 6 V to 20 V
9111215232654--
> 20 V to 60 V
11121417-----
> 60 V to 200 V
38383838-----
> 200 V to 600 V22273545-----
> 600 V to 1000 V28334057-----

 

DC Current (including Picoameters) from 10 pA up to 20 A


Expanded Uncertainty / (μA/A)
10 pA
1400
100 pA
700
1 nA
530
10 nA
530
100 nA
530
1 μA
210
10 μA
210
100 μA
120
100 μA to < 1 mA14
1 mA to < 10 mA12
10 mA to 100 mA7
> 100 mA to 1 A13
> 1 A to 10 A12
> 10 A to 20 A26

AC Current from 10 μA to 20 A and 10 Hz to 30 kHz

Expanded Uncertainty / (μA/A)

10 Hz to < 57 Hz57 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz
10 μA to < 50 μA740430450440----
50 μA to < 100 μA470140140140----
0.1 mA to < 1 mA12059596988110340620
1 mA to < 10 mA1205959597387100150
10 mA to < 100 mA1205858586271120170
100 mA86292929294583130
> 100 mA to 1 A683030304583120200
> 1 A to 10 A110727295300500700-
> 10 A to 20 A1701101101504507501100-
DC and low frequency electrical metrology - Electrical standards AC/DC transfer instruments Current AC-DC difference Comparison with a reference standard

Capability

from 0.0001 A to 100 A and 10 Hz to 200 kHz
with Calibration and Measurement Capability of -

Expanded Uncertainty / (μA/A)

10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz
0.1 mA to < 1 mA9952403636364451
1 mA to < 10 mA9952393534343941
10 mA to < 30 mA65544444
30 mA to < 50 mA86544444
50 mA to < 100 mA97544444
100 mA97654444
> 100 mA to 400 mA108655544
> 400 mA to 1.5 A118655555
> 1.5 A to 5 A119755555
> 5 A to 20 A1412111010101010
> 20 A to 50 A1916151414141413
> 50 A to 100 A2218171616151414


Expanded Uncertainty / (μA/A)

> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz
0.1 mA to < 1 mA73120-----
1 mA to < 10 mA4965-----
10 mA to < 30 mA44567810
30 mA to < 50 mA45567911
50 mA to < 100 mA456781013
100 mA456791217
> 100 mA to 400 mA5668101218
> 400 mA to 1.5 A5678101319
> 1.5 A to 5 A56791321-
> 5 A to 20 A101416212632-
> 20 A to 50 A162129374661-
> 50 A to 100 A182638526584-

Voltage AC-DC difference Comparison with a reference standard

Capability

From 1 mV to 1000 V and 10 Hz to 1 MHz
with Calibration and Measurement Capability of -

Expanded Uncertainty / (μV/V)

10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz
1 mV to < 2 mV260220200170170170170170180180
2 mV to < 5 mV110857777777777777878
5 mV to < 10 mV77585353535353535454
10 mV to < 20 mV53413838383838384040
20 mV to < 30 mV46333131313131313333
30 mV to < 50 mV38282424232323242424
50 mV to < 100 mV30201614141414141516
100 mV to < 200 mV28201613131313131515
200 mV to < 300 mV28201613121212121414
300 mV to < 0.5 V28201613121212121414
0.5 V to < 1 V4.82.51.61.31.21.21.21.21.41.4
1 V to < 2 V5.22.61.71.31.31.21.21.21.41.4
2 V to < 3 V5.12.61.61.31.31.21.21.21.41.4
3 V to 4 V5.22.61.71.31.31.21.21.21.41.4
> 4 V to 6 V7555555444
> 6 V to 10 V7666666555
> 10 V to 12 V8766666666
> 12 V to 20 V8777777666
> 20 V to 30 V10988888888
> 30 V to 40 V10999999888
> 40 V to 60 V1110999910999
> 60 V to 100 V111010101010109910
> 100 V to 120 V12111010101010101010
> 120 V to 200 V12111111111111101011
> 200 V to 300 V14121212121212111112
> 300 V to 400 V14131212121212121212
> 400 V to 600 V15141414141414131316
> 600 V to 1000 V17161616161616151618



Expanded Uncertainty / (μV/V)

> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz> 200 kHz to 300 kHz> 300 kHz to 500 kHz> 500 kHz to 800 kHz> 800 kHz to 1 MHz
1 mV to 2 mV180180180190220270310360420
2 mV to < 5 mV78798088100160220280310
5 mV to < 10 mV5456576875120180220260
10 mV to < 20 mV404143566370110170200
20 mV to < 30 mV33333645536689130150
30 mV to < 50 mV25252631415780140150
50 mV to < 100 mV16172026375271120140
100 mV to < 200 mV15162026385473129140
200 mV to < 300 mV15161925375167110120
300 mV to < 0.5 V15161925375167110120
0.5 V to < 1 V1.62.23.05.26.88.310.613.817.6
1 V to < 2 V1.62.23.05.26.88.310.613.817.6
2 V to < 3 V1.62.12.85.06.89.211.615.819.5
3 V to 4 V1.62.12.85.06.89.211.615.819.5
> 4 V to 6 V45581115233037
> 6 V to 10 V55691217253239
> 10 V to 12 V66691318263340
> 12 V to 20 V677101419273442
> 20 V to 30 V9111317222837--
> 30 V to 40 V9111418233039--
> 40 V to 60 V101114192531---
> 60 V to 100 V1012152026----
> 100 V to 120 V11121521-----
> 120 V to 200 V11131621-----
> 200 V to 300 V12141723-----
> 300 V to 400 V12151825-----
> 400 V to 600 V17222839-----
> 600 V to 1000 V20263649-----

Current shunts Phase displacement Comparison with a primary realisation; Comparison with a reference standard

Capability

-0.1 rad to 0.1 rad
Current 0.01 A to 100 A
Frequency 40 Hz to 200 kHz 
with Calibration and Measurement Capability of -

Expanded Uncertainty / (μrad)

40 Hz to 1 kHz1 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz
0.01 A to 0.5 A38164181200
> 0.5 A to 1 A410194795230
> 1 A to 2 A4112153110250
> 2 A to 5 A5122358120280
> 5 A to 10 A6132563130300
> 10 A to 20 A6163075150
> 20 A to 40 A6183586170
> 40 A to 100 A7203996190

Current AC-DC difference; Resistance; Voltage AC-DC difference Comparison with a primary realisation; Comparison with a reference standard

Capability

Resistance 0.0001 Ω to 1000 Ω
Current 0.1 mA to 100 A
Frequency 10 Hz to 200 kHz
with Calibration and Measurement Capability of -


 Expanded Uncertainty / (μΩ/Ω)

10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz
0.1 mA to < 1 mA9952403636364451
1 mA to < 10 mA9952393534343941
10 mA to < 30 mA65544444
30 mA to < 50 mA86544444
50 mA to < 100 mA97544444
100 mA97654444
> 100 mA to 400 mA108655544
> 400 mA to 1.5 A118655555
> 1.5 A to 5 A119755555
> 5 A to 20 A1412111010101010
> 20 A to 50 A1916151414141413
> 50 A to 100 A2218171616151414



Expanded Uncertainty / (μΩ/Ω)

> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz
0.1 mA to < 1 mA73120-----
1 mA to < 10 mA4965-----
10 mA to < 30 mA44567810
30 mA to < 50 mA45567911
50 mA to < 100 mA456781013
100 mA456791217
> 100 mA to 400 mA5668101218
> 400 mA to 1.5 A5678101319
> 1.5 A to 5 A56791321-
> 5 A to 20 A101416212632-
> 20 A to 50 A162129374661-
> 50 A to 100 A182638526584-

Potential dividers; Voltage ratio boxes Voltage ratio Comparison with a primary realisation; Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

0.05 μV/V for voltage ratios from 0.01 to 0.1 at voltages from 10 V to 100 V
0.1 μV/V for voltage ratios from 0.01 to 0.1 at voltages between 100 V and 1000 V
0.2 μV/V for voltage ratios from 0.001 to 0.01 at voltages between 10 V and 1000 V

Voltage standards - E.M.F. reference devices Voltage Comparison with a primary realisation; Comparison with a reference standard

Capability

at 1 V, 1.018 V and 10 V
with Calibration and Measurement Capability of -

Expanded Uncertainty / (nV/V)
1 V40
1.018 V40
10 V20

Magnetism - Magnetic instruments and equipment Flux density meters Magnetic flux density Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

0 to 2 T on DC
0.02 µT + 1.5 % from 0 to 0.005 T at 50 Hz and 60 Hz (AC)

Transverse Probes, Reference Magnets (DC)

with Calibration and Measurement Capability of -

Magnetic Flux Density
(T)
Expanded Uncertainty
0 to < 0.05
0.05 to 2
0.1 μT + 0.5 %
0.15 %


Axial Probes (DC)

with Calibration and Measurement Capability of -
Magnetic Flux Density
(T)
Expanded Uncertainty
0 to 0.0950.1 μT + 0.5 %

Flux meters Magnetic flux Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

1% from 10 μWb to 10 mWb

Helmholtz coils; Magnets; Solenoids Magnetic field strength Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

0.1 µT + 0.3 % from 0 T  to 2 T

Time and frequency metrology - Frequency, time and waveform measuring equipment Clocks and timers; Counters; Distortion analysers; Electrical power calibrators; Frequency analysers; Frequency meters; Oscilloscopes; Power analysers; Time interval meters Attenuation; Current; Current harmonics; Distortion; Frequency; Phase angle; Voltage; Voltage flicker; Voltage harmonics Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

Distortion

Voltage and current harmonic amplitude and total harmonic distortion up to 99ᵗʰ harmonic, for harmonic amplitudes of 5 % to 40 % of fundamental voltage or current
from 0.01 V to 240 V, 0.005 A to 20 A and 47 Hz to 63 Hz
Fundamental Voltage / CurrentHarmonic NumberExpanded Uncertainty
(% of fundamental)
Reference Standard
0.01 V to 240 V1 to 390.001Josephson Arbitraty Wavefom Synthesizer
1 V to 240 V
1 to 99
0.005
Digital Sampling Voltmeter
0.005 A to 20 A
1 to 990.005Digital Sampling Voltmeter

Voltage Harmonics  (Meter and Source) up to 39ᵗʰ, phase, from 1 V to 240 V and 47 Hz to 63 Hz

DistortionExpanded Uncertainty (degrees)
Harmonic Number
1 - 911 - 1921 - 39
5 % to < 10 %0.00160.00210.0099
10 % to < 20 %0.00140.00190.0097
20 % to < 30 %0.00120.00180.0095
30 % to < 40 %0.00110.00180.0094
Current Harmonics (Meter and Source) up to 39ᵗʰ, phase, from 0.005 A to 20 A and 47 Hz to 63 Hz
DistortionExpanded Uncertainty (degrees)
Harmonic Number
1 - 911 - 1921 - 39
5 % to < 10 %0.00260.00310.0109
10 % to < 20 %0.00240.00290.0107
20 % to < 30 %0.00220.00280.0105
30 % to < 40 %0.00210.00280.0104
Flicker Meters
0.01% of fundamental for Voltage Flicker from 1 V to 240 V and 47 Hz to 53 Hz

Phasor Measurement Units
0.007 % for Total Vector Error from 1 V to 240 V and 47 Hz to 53 Hz
0.007 % for Total Vector Error, 0.005 A to 20 A and 47 Hz to 53 Hz
0.0022 º for phase angle from 0 º to 90 º, 1 V to 240 V, 0.005 A to 20 A and 47 Hz to 53 Hz
0.0034 % for frequency from 47 Hz to 53 Hz, 1 V to 240 V
0.0016 Hz/s for rate of change of frequency from -5 Hz/s to 5 Hz/s, 47 Hz to 53 Hz and 1 V to 240 V

Phase Sources, Phase Standards, Power Calibrators
from -180 º to 180 º, 0.01 V to 1000 V, 0.005 A to 200 A and 40 Hz to 1000 Hz
FrequencyExpanded Uncertainty
(Hz)(Degrees)
40 to 600.0005
> 60 to 4000.001
> 400 to 10000.002

Phase Meters, Power Analysers
FrequencyExpanded Uncertainty
(Hz)(Degrees)
40 to 600.0005
> 60 to 4000.001
> 400 to 10000.002

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.