National Association of Testing Authorities, Australia
Scope of Accreditation
EMC Technologies Pty Ltd
Site
Melbourne Laboratory – EMC Test Laboratory
Accreditation No.
5292
Site No.
5285
Date of Accreditation
28 Apr 1994
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
High frequency electrical metrology - Communications, electromagnetic field strength and EMC test equipment | Active rod antennas; Biconical antennas; Horn antennas; Log periodic antennas | Antenna factor; Frequency | Direct measurement by electrical input | SAE ARP 958, version D | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Narrowband transducers and indicators with Calibration and Measurement Capability of- Antenna Factor from -5 dB(1/m) up to 50 dB(1/m)± 1.0 dB from 9kHz to 40 GHz | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Artificial mains networks (AMN); Line impedance stabilisation networks (LISN) | Voltage division factor | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the frequency range 9 kHz to 108 MHz± 0.6 dB for voltage division factor ± 2.1% for voltage drop ± 10.9% for impedance± 3.8º for phase | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Coupling/decoupling networks (CDN) | Insertion loss | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the frequency range 150 kHz to 230 MHz± 10.9% for impedance | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
E&H field probes and meters; Electromagnetic field hazard meters and warning indicators | Frequency | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of-
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electrical fast transients (EFT) and surge generators | Pulse characteristics | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Surge Generators with Calibration and Measurement Capability of - in the range ± 0.1 kV to 7 kV± 5.60% for open circuit voltage ± 10.53% for open circuit front time± 4.4% for open circuit duration ± 4.4% for short circuit current ± 4.29% for short circuit front time ± 4.3% for short circuit duration | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pulse characteristics | Direct measurement by electrical input | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Electrical fast transients (EFT) with Calibration and Measurement Capability of - in the range ± 0.5 kV to 20 kV± 5.80% for peak voltage ± 4.2% for pulse width± 1.2% for rise time ± 3.07% for repetition rate | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electrostatic discharge (ESD) generators | Pulse characteristics | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the range ± 2 kV to ± 30 kV± 14.2% for rise time ± 8.10% for peak current± 1.2% for peak voltage | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Harmonic emission compliance test systems | Harmonic currents | Comparison with a reference standard by differential measurement | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.15% of the fundamental of current for Harmonic currents - from 0.7 A to 16 A total rms per phase - from 1st to 40th harmonic with fundamental frequency between 50 to 60 Hz and voltage between 120 to 230 V | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Line impedance stabilisation networks (LISN) | Voltage division factor | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Balanced Telecom ISN (TLISN) with Calibration and Measurement Capability of - in the frequency range 150 kHz to 30 MHz± 4.3 Ω for common mode impedance ± 2.4º for common mode phase± 0.97 dB for voltage division factor ± 1.12 dB for isolation / DCMD ± 2.28 dB for Longitudinal conversion loss (LCL) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Voltage fluctuation and flicker compliance test systems | Internal reference impedance resistance; Short term flicker; Voltage deviation | Comparison with a reference standard by substitution measurement | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - 0.5% of Pst in the range from 0.5 Ps to 1.5 Ps for Short term flicker perceptibility 0.07% of fundamental of the voltage in the range from 0.5 to 2.0% for Voltage deviation, Dmax, with fundamental frequency between 50 Hz to 60 Hz and voltage between 1 V to 230 V 1.8 MΩ for Internal reference impedance resistance |
ISO/IEC 17025 (2017)
Manufactured Goods
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Electromagnetic compatibility (EMC) - Emissions - Commercial | Airborne equipment | Conducted emissions; Radiated emissions | Not applicable | RTCA DO160 F/G section 21 | Conducted RF emissions 150 kHz to 150 MHz |
Electrical and electronic products; Power tools - Electric | Conducted emissions; Radiated emissions | Not applicable | AS/NZS CISPR 14.1; CISPR 14-1; EN 55014-1; CNS 13783-1; | Radiated measurements performed in the 10 metre iOATS Semi Anechoic Chamber Room 12 - Test Site 1 3 metre iOATS Semi Anechoic Chamber Room 13 - Test Site 2 | |
Electrical equipment for measurement, control and laboratory use | Conducted emissions; Radiated emissions | Not applicable | AS/NZS 2064; AS/NZS CISPR 11; CISPR 11; EN 55011; CNS 13803; | ||
Electrical lighting and similar equipment | Conducted emissions; Radiated emissions | Not applicable | AS/NZS CISPR 15; CISPR 15; EN 55015 | ||
Information technology equipment | Conducted emissions; Conducted telecommunications port emissions; Radiated emissions | AMN and voltage probe; ISN and current clamp | Excluding capacitive voltage probe tests for telecommunications ports AS/NZS CISPR 22; CISPR 22; CAN/CSA-CEI/IEC CISPR 22; VCCI; CNS 13438; EN 55022; AS/NZS CISPR 32; CISPR 32 and EN 55032 | ||
Machine tools | Conducted emissions; Discontinuous disturbances; Mains harmonics and flicker; Radiated emissions | Not applicable | AS/NZS 2064; AS/NZS CISPR 11; CISPR 11; EN 55011; CNS 13803; | ||
Maritime navigation equipment | Conducted emissions; Radiated emissions | Not applicable | EN 60945 | Excluding measurements below 2 kHz | |
Medical electrical equipment | Radiated emissions | Not applicable | AS/NZS 2064; AS/NZS CISPR 11; CISPR 11; EN 55011; CNS 13803; | ||
Radio equipment | Conducted emissions; Radiated emissions | Not applicable | ETSI EN 301 489-1 | ||
Railway vehicles and equipment including railway signalling and telecommunications equipment | Conducted emissions; Radiated emissions | Not applicable | EN 50121-2; EN 50121-3-1; EN 50121-3-2; EN 50121-4; EN 50121-5 | ||
Sound and television broadcast receivers and associated equipment | Conducted emissions; Radiated emissions | Not applicable | AS/NZS CISPR 13; CISPR 13; EN 55013; CNS 13439 | ||
Telecommunication network equipment | Conducted emissions; Radiated emissions | Not applicable | ETSI EN 300 386 | ||
Vehicles, boats and internal combustion engine driven equipment | Conducted emissions; Radiated emissions | Not applicable | AS/NZS CISPR 25; CISPR 25; EN 55025; | Testing on components only | |
Electromagnetic compatibility (EMC) - Emissions - Military | Aerospace and military grade equipment | Radiated emissions | Not applicable | MIL-STD-461 B/C: RE01; MIL-STD-461 B/C: RE02; MIL-STD-461 D/E/F/G: RE101; MIL-STD-461 D/E/F/G: RE102; | MIL-STD-461 B/C: RE01 - 30 Hz to 50 kHz MIL-STD-461 B/C: RE02 - 14 kHz to 10 GHz MIL-STD-461 D/E/F/G: RE101 - 30 Hz to 100 kHz MIL-STD-461 D/E/F/G: RE102 - 10 kHz to 18 GHz |
Conducted emissions | Not applicable | MIL-STD-461 B/C: CE01 - 0 to 15 kHz; | |||
Power leads | Conducted emissions | Not applicable | MIL-STD-461 D/E/F/G: CE101 (30 Hz to 10 kHz); | ||
Electromagnetic compatibility (EMC) - Immunity - Commercial | Airborne equipment | Conducted disturbance; Electrostatic discharge; Induced signal susceptibility; Magnetic effects; Radiated radio-frequency; Voltage spikes | Not applicable | RTCA DO160 F/G Sections 15, 17, 18, 19, 20, 25; | |
Components of fire, intruder and social alarm systems | Common mode low frequency; Conducted disturbance; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | EN 50130-4; | ||
Electrical and electronic products; Power tools - Electric | Common mode low frequency; Conducted disturbance; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | AS/NZS CISPR 14.2; CISPR 14-2; EN 55014-2; | ||
Electrical equipment for measurement, control and laboratory use | Common mode low frequency; Conducted disturbance; Conducted susceptibility; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Radiated susceptibility; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | IIEC 61326-1; EN 61326-1; | ||
Electrical lighting and similar equipment | Common mode low frequency; Conducted disturbance; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | EN 61547; ESD - IEC 801-2; AS/NZS 61000.4.2; IEC 61000-4-2; EN 61000-4-2; IEC 255-22-2; ISO 10605; SAE J1113-13; AS/NZS 61000.4.3; IEC 61000-4-3; EN 61000-4-3; IEC 60255-22-3; ISO 11452-2; ISO 11452-4; IEC 801-4; AS/NZS 61000.4.4; IEC 61000-4-4; EN 61000-4-4; IEC 60255-22-4; AS/NZS 61000.4.5; IEC 61000-4-5; EN 61000-4-5; IEC 255-5; AS/NZS 61000.4.6; IEC 61000-4-6; EN 61000-4-6; AS/NZS 61000.4.8; IEC 61000-4-8; EN 61000-4-8; IEC 61000-4-9; EN 61000-4-9; IEC 61000-4-10; EN 61000-4-10; AS/NZS 61000.4.11; IEC 61000-4-11; EN 61000-4-11 IEC 61000-4-12; AS/NZS 61000.4.13; IEC 61000-4-13; EN 61000-4-13; AS/NZS 61000.4.16; IEC 61000-4-16; EN 61000-4-16; IEC 61000-4-18; EN 61000-4-18 | ||
Electricity meters and associated transformers | Electrical fast transient; Radiated radio-frequency; Surge immunity | Not applicable | NSC publication for Pattern Approval and Initial verification of electricity meters and associated transformers | ||
Gaming machines | Electrical fast transient; Voltage short dips and interruptions | Not applicable | Australian/New Zealand Gaming Machine National Standard clauses 6.28 to 6.35 only | ||
Information technology equipment | Common mode low frequency; Conducted disturbance; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | AS/NZS CISPR 24; CISPR 24; EN 55024; ESD - IEC 801-2; AS/NZS 61000.4.2; IEC 61000-4-2; EN 61000-4-2; IEC 255-22-2; ISO 10605; SAE J1113-13; AS/NZS 61000.4.3; IEC 61000-4-3; EN 61000-4-3; IEC 60255-22-3; ISO 11452-2; ISO 11452-4 IEC 801-4; AS/NZS 61000.4.4; IEC 61000-4-4; EN 61000-4-4; IEC 60255-22-4; AS/NZS 61000.4.5; IEC 61000-4-5; EN 61000-4-5; IEC 255-5; AS/NZS 61000.4.6; IEC 61000-4-6; EN 61000-4-6; AS/NZS 61000.4.8; IEC 61000-4-8; EN 61000-4-8; IEC 61000-4-9; EN 61000-4-9; IEC 61000-4-10; EN 61000-4-10; AS/NZS 61000.4.11; IEC 61000-4-11; EN 61000-4-11 IEC 61000-4-12; AS/NZS 61000.4.13; IEC 61000-4-13; EN 61000-4-13; AS/NZS 61000.4.16; IEC 61000-4-16; EN 61000-4-16; IEC 61000-4-18; EN 61000-4-18 | ||
Low voltage switchgear and control equipment | Common mode low frequency; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | EN 60947; | ||
Machine tools | Conducted disturbance; Electrical fast transient; Electrostatic discharge; Radiated radio-frequency; Surge immunity; Voltage short dips and interruptions | Not applicable | |||
Medical electrical equipment | Common mode low frequency; Conducted disturbance; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | AS/NZS 3200.1.2; IEC 60601-1-2; EN 60601-1-2; | ||
Multimedia equipment | Common mode low frequency; Conducted disturbance; Conducted susceptibility; DC input power ripple; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Induced signal susceptibility; Magnetic effects; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Radiated susceptibility; Ring wave; Surge immunity; Voltage fluctuations; Voltage short dips and interruptions; Voltage spikes | Not applicable | EN 55035 excluding clause 4.2.7; AS/NZS CISPR 35 excluding clause 4.2.7; CISPR 35 excluding clause 4.2.7 | ||
Radio equipment | Common mode low frequency; Conducted disturbance; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | ETSI EN 301 489-1; | ||
Railway vehicles and equipment including railway signalling and telecommunications equipment | Common mode low frequency; Conducted disturbance; Conducted susceptibility; Damped oscillatory wave; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Radiated susceptibility; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | EN 50121-3-2; EN 50121-4; EN 51021-5 | ||
Telecommunication network equipment | Common mode low frequency; Conducted disturbance; Conducted susceptibility; Damped oscillatory wave; Electrical fast transient; Electrostatic discharge; Harmonic and inter harmonics low frequency; Power frequency magnetic field; Pulsed magnetic field; Radiated radio-frequency; Radiated susceptibility; Ring wave; Surge immunity; Voltage short dips and interruptions | Not applicable | ETSI EN 300 386; | ||
Electromagnetic compatibility (EMC) - Immunity - Military | Aerospace and military grade equipment | Conducted susceptibility; Radiated radio-frequency; Radiated susceptibility | Bulk cable injection; Damped sinusoidal transients; Impulse excitation; Magnetic field | MIL-STD-461 B/C: RS01, RS02, RS03; | MIL-STD-461 B/C RS01: 30 Hz to 50 kHz; MIL-STD-461 B/C RS03: 14 kHz to 18 GHz ; MIL-STD-461 D/E/F/G RS101: 30 Hz to 100 kHz; MIL-STD-461 D/E/F/G RS103: 2 MHz to 18 GHz; MIL-STD-461 B/C CS01: 30 Hz to 50 kHz; MIL-STD-461 B/C CS02: 0.5 MHz to 400 MHz; MIL-STD-461 D/E/F/G CS101: 30 Hz to 150 kHz; MIL-STD-461 D/E/F/G CS114: 10 kHz to 400 MHz; |
Electromagnetic radiation exposure - Extremely low frequency (ELF) fields | Electricity grid installations; High voltage transformer equipment | Broadband measurements of H-fields; Narrowband measurements of H-fields | Not applicable | AS/NZS 2772.2 and ARPANSA Radiation Protection (RPS S-1) and NZS 2772.1; | Broadband measurements of E-fields in the range 100 kHz to 45.5 GHz; Broadband measurements of H-fields in the range 100 kHz to 1 GHz; Narrowband measurements of E-fields in the range 9 kHz to 40 GHz; Narrowband measurements of H-fields in the range 9 kHz to 300 MHz |
Electromagnetic radiation exposure - Radiofrequency (RF) fields | Antennas; Microwave ovens | Broadband measurements of E-fields; Broadband measurements of H-fields; Measurement of electromagnetic fields; Narrowband measurements of E-fields; Narrowband measurements of H-fields | Not applicable | AS/NZS 2772.2 and ARPANSA Radiation Protection (RPS S-1) and NZS 2772.1; | Broadband measurements of E-fields in the range 100 kHz to 45.5 GHz; |
Shielded enclosures | Attenuation performance | Not applicable | MIL-STD-285 Rev. 1.8 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
5292
Site No.
5285
Print date
11 Nov 2024