Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DC and low frequency electrical metrology - Electrical measurement and test equipment | Ammeters; Digital multimeters (DMM) | AC current; DC current | Comparison with a reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of -D.C. current A.C. current 1.30% + 0.25 μA from 25 μA to 100 μA and from 1 kHz to 2 kHz, 0.70% + 0.7 μA from 100 μA to 1 mA and from 1 kHz to 2 kHz, 0.46 % + 7 μA from 1 mA to 10 mA and from 1 kHz to 2 kHz, |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital multimeters (DMM); Ohm meters | Resistance | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of - 7 mΩ at 100 mΩ, 1 Ω, 10 Ω 0.011 Ω at 100 Ω 0.083 Ω at 1 kΩ 0.82 Ω at 10 kΩ 0.004 kΩ at 100 kΩ 0.00015 MΩ at 1 MΩ 0.0039 MΩ at 10 MΩ 0.52 MΩ at 100 MΩ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Digital multimeters (DMM); Voltmeters | AC voltage; DC voltage | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of -DC Voltage 0.004 % + 3 μV from 0 to 200 mV AC Voltage 0.032 % + 20 μV from 20 mV to 200 mV and from 45 Hz to 1 kHz, 0.033 % + 90 μV from 200 mV to 2 V and from 45 Hz to 1 kHz,
0.3 % + 70 μV from 20 mV to 200 mV and from 20 kHz to 100 kHz, 0.22 % + 2 mV from 200 mV to 2 V and from 20 kHz to 100 kHz. 0.18 % + 33 mV from 2 V to 20 V and from 20 kHz to 100 kHz. |