AusNet Transmission Group Pty Ltd

Site

Mondo – Electrical Calibration Laboratory

Accreditation No.

754

Site No.

747

Date of Accreditation

14 Feb 1969

Address

308 Hyde Street
Yarraville, VIC 3013
Australia

mondo.com.au

Contact

Mr Deena Sri P: +61 (03) 93967552

Availability

Services available to external clients

Scope of Accreditation

Mondo – Electrical Calibration Laboratory

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

D.C. voltage

5 μV/V + 1 μV up to 100 V;
8 μV/V above 100 V to 1 kV;
50 μV/V above 1 kV to 1.5 kV;
0.05% of reading above 1.5 kV to 50 kV

 A.C. voltage
1 mV to 2.2 mV
1 µV from 10 Hz to 30 kHz
1.2 µV above 30 kHz to 50 kHz
6.8 µV above 50 kHz to 100 kHz
2.2 mV to 22 mV
115 µV/V from 10 Hz to 40Hz
(55 µV/V + 0.2 µV) above 40 Hz to 10 kHz
(60 µV/V + 0.4 µV) above 10 kHz to 20 kHz
(85 µV/V + 0.8 µV) above 20 kHz to 30 kHz
(100 µV/V + 1.1 µV) above 30 kHz to 50 kHz
(360 µV/V + 6 µV) above 50 kHz to 10 0 kHz
22 mV to 220 mV
(55 µV/V + 1 µV) from 10 Hz to 40 Hz
60 µV/V above 40 Hz to 10 kHz
65 µV/V above 10 kHz to 20 kHz
(75 µV/V + 1 µV) above 20 kHz to 30 kHz
(65 µV/V + 2 µV) above 30 kHz to 50 kHz
(175 µV/V + 10 µV) above 50 kHz to 100 kHz
220 mV to 2.2 V
55 µV/V from 10 Hz to 40 Hz
50 µV/V above 40 Hz to 50 kHz
80 µV/V above 10 kHz to 100 kHz
2.2 V to 22 V
55 µV/V from 10 Hz to 20 Hz
50 µV/V above 20 Hz to 50 kHz
75 µV/V above 50 kHz to 100 kHz
2 2 V to 220 V
55 µV/V from 10 Hz to 20 Hz
50 µV/V above 20 Hz to 50 kHz
85 µV/V above 50 kHz to 100 kHz
220 V to 1000 V
40µV/V from 40Hz to 1kHz
0.3% of reading from 0.25 V to 10 V and 100 kHz to 1 MHz

D.C. current

0.1% of reading + 5 pA from 100 pA to 100 μA
100 μA/A + 5 nA above 100 μA to 100 A
0.1% of reading above 100 A to 1,000 A 

A.C. current

0.05% of reading from 20 A to 4000 A at 50 Hz
0.05% of reading above 10 μA to 2.2 A from 10Hz to 10 kHz
0.05% of reading above 2.2 A to 20 A from 20 Hz to 5 kHz

Resistance

0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 mΩ to 10 mΩ
20 μΩ/Ω of reading above 10 mΩ to 1 Ω
5 μΩ/Ω of reading above 1 Ω to 10 kΩ
50 μΩ/Ω of reading above 10 kΩ to 1 MΩ
0.1% of reading above 1 MΩ to 10 MΩ
0.5% of reading above 10 MΩ to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above 10 GΩ to 1 TΩ
20% of reading above 1 TΩ to 100 TΩ

DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters AC current; DC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
D.C ammeters including clamp on ammeters

0.1% of reading + 5 pA from 100 pA to 100 μA
100 μA/A + 5 nA above 100 μA to 100 A
0.1% of reading above 100A to 1,000 A

A.C.  ammeters

0.05% of reading from 20 A to 4000 A at 50 Hz
0.05% of reading above 10 μA to 2.2 A from 10Hz to 10 kHz
0.05% of reading above 2.2 A to 20 A from 20 Hz to 5 kHz



Data recorders; Digital multimeters (DMM); Ohm meters Resistance Comparison with a reference standard

Capability

Including insulation resistance testers from 100 V to 10 kV and 50 GΩ

with Calibration and Measurement Capability of -

0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 mΩ to 10 mΩ
20 μΩ/Ω above 10 mΩ to 1 Ω
5 μΩ/Ω above 1 Ω to 10 kΩ
50 μΩ/Ω above 10 kΩ to 1 MΩ
0.1% of reading above 1 MΩ to 10 MΩ
0.5% of reading above 10 MΩ to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above 10 GΩ to 1 TΩ
20% of reading above 1 TΩ to 100 TΩ

Data recorders; Digital multimeters (DMM); Voltmeters AC voltage; DC voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

D.C. voltage
5 μV/V + 1 μV up to 100 V;
8 μV/V above 100 V to 1 kV;
50 μV/V above 1 kV to 1.5 kV;
0.05% of reading above 1.5 kV to 50 kV
 A.C. voltage
1 mV to 2.2 mV
1 µV from 10 Hz to 30 kHz
1.2 µV above 30 kHz to 50 kHz
6.8 µV above 50 kHz to 100 kHz
2.2 mV to 22 mV
115 µV/V from 10 Hz to 40Hz
(55 µV/V + 0.2 µV) above 40 Hz to 10 kHz
(60 µV/V + 0.4 µV) above 10 kHz to 20 kHz
(85 µV/V + 0.8 µV) above 20 kHz to 30 kHz
(100 µV/V + 1.1 µV) above 30 kHz to 50 kHz
(360 µV/V + 6 µV) above 50 kHz to 10 0 kHz
22 mV to 220 mV
(55 µV/V + 1 µV) from 10 Hz to 40 Hz
60 µV/V above 40 Hz to 10 kHz
65 µV/V above 10 kHz to 20 kHz
(75 µV/V + 1 µV) above 20 kHz to 30 kHz
(65 µV/V + 2 µV) above 30 kHz to 50 kHz
(175 µV/V + 10 µV) above 50 kHz to 100 kHz
220 mV to 2.2 V
55 µV/V from 10 Hz to 40 Hz
50 µV/V above 40 Hz to 50 kHz
80 µV/V above 10 kHz to 100 kHz
2.2 V to 22 V
55 µV/V from 10 Hz to 20 Hz
50 µV/V above 20 Hz to 50 kHz
75 µV/V above 50 kHz to 100 kHz
2 2 V to 220 V
55 µV/V from 10 Hz to 20 Hz
50 µV/V above 20 Hz to 50 kHz
85 µV/V above 50 kHz to 100 kHz
220 V to 1000 V
40µV/V from 40Hz to 1kHz
0.3% of reading from 0.25 V to 10 V and 100 kHz to 1 MHz

Electricity meters Active energy; Reactive energy By methods of NMI NITP 14 Calibration of single phase and poly phase energy meters including compliance with NMI NITP 14

Capability

Calibration in the laboratory:

Single and three phase active energy meters from 1 to 0.5 p.f. (leading and

lagging currents) at 50 Hz,

with Calibration and Measurement Capability of -

(0.03/cos Φ)% from 63.5 V to 415 V and 0.025 to 0.1 A,

(0.02/cos Φ)% from 63.5 V to 415 V and 0.1 to 12 A,

(0.02/cos Φ)% from 63.5 V to 240 V and 12 to 50 A,

(0.04/cos Φ)% from 63.5 V to 240 V and 50 to 100 A

Single and three phase reactive energy meters from 0 to 0.866 p.f (leading and

lagging currents) at 50 Hz,

with Calibration and Measurement Capability of -

(0.05/sin Φ)% from 63.5 to 415 V and 0.025 to 12 A,

(0.05/sin Φ)% from 63.5 to 240 V and 12 to 50 A,

(0.05/sin Φ)% from 63.5 to 240 V and 50 to 100 A

Field calibration at voltages from 63.5 V to 120 V, 50 Hz and currents up to 6A

with Calibration and Measurement Capability of -

(0.1/cos Φ)% for three phase Watt-hour meters

(0.2/sin Φ)% for three phase Var-hour meters

Galvanometers and null detectors DC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
 0.1% of reading + 5 pA from 100 pA to 100 μA
100 μA/A + 5 nA from 100 μA to 100 A
0.1% of reading over above 100 A to 1,000 A
LCR meters Capacitance; Inductance; Resistance Comparison with a reference standard

Capability

Excluding AC resistance
with Calibration and Measurement Capability of

Capacitance
Two terminal meters at 1 000 Hz
0.01% of reading + 1 pF from 100 pF to 0.1 μF
0.05% of reading above 0.1 μF to 1 μF
0.2% of reading above 1 μF to 11 μF
Three terminal meters at 1 000 Hz
0.01% of reading + 0.03 pF from 10 pF to 0.1 μF
0.04% of reading above 0.1 μF to 1 μF
0.1% of reading above 1 μF to 11 μF

Loss tangent at 1 000 Hz
0.1% of reading + 0.00005

Inductance
0.1% of reading in the range 10 μH to 100 mH

Resistance
0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 to 10 mΩ
20 μΩ/Ω from above 10 mΩ to 1 Ω
5 μΩ/Ω from above 1 Ω to 10 kΩ
50 μΩ/Ω from above 10 kΩ to 1 MΩ
0.1% of reading above 1 to 10 MΩ
0.5% of reading above 10 to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above 10 GΩ to 1 TΩ
20% of reading above 1 TΩ to 100 TΩ

Phase angle meters Phase angle Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

0.04° from 10 Hz to 2 kHz
0.08°above 2 kHz to 5 kHz
0.15° above  5 kHz to 10 kHz
0.29° above 10 kHz to 30 kHz

Power factor meters Power factor Comparison with a reference standard by null measurement

Capability

with Calibration and Measurement Capability of -

0.002

Varmeters; Watt meters Active power; Reactive power Comparison with a reference standard

Capability

 Wattmeters

At 50 Hz

with Calibration and Measurement Capability of -

0.04% from 63.5 V to 100 V and 50 mA to 10 A at unity p.f. (single phase)
0.06% up to 1,200 V and 100 A at unity p.f. (single phase);
0.05% from 63.5 V to 100 V and 50 mA to 10 A at 0.5 p.f. (single phase)
0.07% up to 1,200 V and 100 A at 0.5 p.f. (single phase)
0.15% up to 600 V and 50 A from unity to 0.5 p.f. (three phase)
0.03 crad for zero power factor error from 63.5 V to 100 V and 50 mA to 10 A
0.05 crad for zero power factor error up to 1,200 V and 100 A

Varmeters

0.2% from 63.5 V to 100 V and 50 mA to 10 A at unity p.f. (single phase)
0.2% up to 1,200 V and 100 A at unity p.f. (single phase);
0.2% from 63.5 V to 100 V and 50 mA to 10 A at 0.5 p.f. (single phase)
0.2% up to 1,200 V and 100 A at 0.5 p.f. (single phase)
0.2% up to 600 V and 50 A from unity to 0.5 p.f. (three phase)
0.2 crad for zero power factor error from 63.5 V to 100 V and 50 mA to 10 A
0.2 crad for zero power factor error up to 1,200 V and 100 A

DC and low frequency electrical metrology - Electrical standards AC bridges Capacitance; Inductance; Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Resistance
0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 mΩ to 10 mΩ
20 μΩ/Ω above 10 mΩ to 1 Ω
5 μΩ/Ω above 1 Ω to 10 kΩ
50 μΩ/Ω above 10 kΩ to 1 MΩ
0.1% of reading above 1 MΩ to 10 MΩ
0.5% of reading above 10 MΩ to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above 10 GΩ to 1 TΩ
20% of reading above 1 TΩ to 100 TΩ
Capacitance
Two terminal meters at 1 000 Hz
0.01% of reading + 1 pF from 100 pF to 0.1 μF
0.05% of reading above 0.1 μF to 1 μF
0.2% of reading above 1 μF to 11 μF
Three terminal meters at 1 000 Hz
0.01% + 0.03 pF from 10 pF to 0.1 μF
0.04% of reading above 0.1 μF to 1 μF
0.1% of reading above 1 μF to 11 μF
Loss tangent at 1 000 Hz
with least uncertainties of measurement of -
0.1% of reading + 0.00005
Inductance
0.1% of reading in the range 10 μH to 100 mH

Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of-

0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 mΩ to 10 mΩ
20 μΩ/Ω above 10 mΩ to 1 Ω
5 μΩ/Ω above1 Ω to 10 kΩ
50 μΩ/Ω from 10 kΩ to 1 MΩ
0.1% of reading above  1 MΩ to 10 MΩ
0.5% of reading above  10 MΩ to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above  10 GΩ to 1 TΩ
20% of reading above  1 TΩ to 100 TΩ


Current shunts Resistance Comparison with a reference standard

Capability

D.C. shunts

For currents up to 1 000 A

with Calibration and Measurement Capability of-

0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 mΩ to 10 mΩ
20 μΩ/Ω above 10 mΩ to 1 Ω
5 μΩ/Ω above 1 Ω to 10 kΩ
50 μΩ/Ω above 10 kΩ to 1 MΩ
0.1% of reading above 1 MΩ to 10 MΩ
0.5% of reading above 10 MΩ to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above 10 GΩ to 1 TΩ
20% of reading above 1 TΩ to 100 TΩ

DC bridges; DC potentiometers; Ratiometers Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

dc bridges
0.03% of reading or 0.1 μΩ (whichever is the greater) from 0.1 mΩ to 1 mΩ
0.02% of reading above 1 mΩ to 10 mΩ
20 μΩ/Ω above 10 mΩ to 1 Ω
5 μΩ/Ω above 1 Ω to 10 kΩ
50 μΩ/Ω above 10 kΩ to 1 MΩ
0.1% of reading above 1 MΩ to 10 MΩ

dc potentiometers
0.5% of reading above 10 MΩ to 100 MΩ
2.0% of reading above 100 MΩ to 10 GΩ
10% of reading above 10 GΩ to 1 TΩ
20% of reading above 1 TΩ to 100 TΩ


ratiometers
0.05% of reading

Inductors Inductance Comparison with a reference standard

Capability


Self inductors at 1 kHz
with Calibration and Measurement Capability of -
0.1% of reading in the range 10 μH to 100 mH


Potential dividers; Voltage ratio boxes Voltage ratio Comparison with a reference standard

Capability

On d.c

with Calibration and Measurement Capability of-

10 μV/V for volt ratio boxes from 1 mV to 1.5 kV
0.01% of reading above 1.5 kV to 2.5 kV
0.05% of reading above 2.5 kV to 50 kV


Precision capacitors Capacitance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Two terminal capacitors at 1 000 Hz

0.01% of reading + 1 pF from 100 pF to 0.1 μF
0.05% of reading above 0.1 μF to 1 μF
0.2% of reading above 1 μF to 11 μF

Three terminal capacitors at 1 000 Hz
0.01% of reading + 0.03 pF from 10 pF to 0.1 μF
0.04% of reading above  0.1 μF to 1 μF
0.1% of reading above 1 μF to 11 μF


Loss tangent at 1 000 Hz
0.1% of reading + 0.00005

Ratiometers Voltage Comparison with a reference standard

Capability

Primary and secondary voltages from 5 V to 125 V at 50 Hz

with Calibration and Measurement Capability of -

0.01% of input voltage

Voltage standards - E.M.F. reference devices; Voltage standards - Standard cells Voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Standard cells

10 μV

Electronic E.M.F.. reference devices

with Calibration and Measurement Capability of -

3 μV/V at 1 V, 1.018 V and 10 V

DC and low frequency electrical metrology - High-voltage/high-current standards and equipment Current transformer test sets; Voltage transformer test sets Current; Phase displacement; Voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Current transformer testing sets

0.005% of reading for current error

0.005 crad for phase error angle

Voltage transformer testing sets

0.005% of reading for voltage error

0.005 crad for phase error angle

High voltage measurement systems Capacitance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -
1% of reading plus 0.00001 tan delta
0.5% of capacitance reading
up to 400 kV at 50 Hz
Partial discharge test and measurement equipment Electric charge - Coulombs Comparison with a reference standard

Capability

Performance tests and checks of partial discharge calibrators to AS 60070 Annex A 

with Calibration and Measurement Capability of - 

3% of reading from 1 pC to 10,000 pC 

Voltage transformers Phase displacement; Voltage Comparison with a reference standard

Capability

Including on site calibrations
with Calibration and Measurement Capability of -

Voltage transformers and protective voltage transformers

Tests at 50 Hz on transformers with rated primary voltages from 63.5 V to 66 kV three phase
and 400 kV single phase and rated secondary voltages from 1 V to 1,100 V

0.01% of reading for voltage error
0.01 crad for phase angle

Field calibration - in High Voltage environment
Tests at 50 Hz on transformers with rated primary voltages from
63.5 V to 66 kV three phase and 400 kV single phase
0.06% for voltage error
0.06 crad for phase angle error


DC and low frequency electrical metrology - High-voltage/high-current standards and equipmentincluding on site calibrations Current transformers Current; Phase displacement Comparison with a reference standard

Capability

Current transformers and protective current transformers
Tests at 50 Hz on transformers with rated primary currents from 0.1 A to 5,000 A
and rated secondary currents from 0.25 A to 10 A
with Calibration and Measurement Capability of -

0.01% of reading for current error
0.015 crad for phase angle error


Field calibration - in High Voltage environment
Tests at 50 Hz on transformers with rated primary currents from 1 to 5 000 A
and rated secondary currents of 1 and 5 A,
with Calibration and Measurement Capability of -
0.05% of reading for current error 0.05 crad for phase angle error
Temperature metrology - Ancillary temperature measuring equipment Digital voltmeters (DVM) and digital multimeters (DMM) Volt to temperature conversion Direct measurement by electrical input

Capability

with Calibration and Measurement Capability of -
Type K Thermocouples
0.3 ºC + (0.1% T) at -200 ºC up to 0 ºC
0.1 ºC + (0.1% T) at 0 ºC to 1100 ºC
where T is magnitude of temperature in ºC
Time and frequency metrology - Frequency, time and waveform measuring equipment Counters; Frequency meters; Time interval meters Frequency; Time interval Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

Counters and frequency meters
2 in 1011  in the range 1 Hz to 10 MHz

Time interval meters

1 count +1 μs

Frequency analysers; Waveform measuring instruments Current; Distortion; Frequency; Time interval; Voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -

D.C. voltage
5 μV/V + 1 μV up to 100 V;
8 μV/V above 100 V to 1 kV;
50 μV/V above 1 kV to 1.5 kV;
0.05% of reading above 1.5 kV to 50 kV
 A.C. voltage
1 mV to 2.2 mV
1 µV from 10 Hz to 30 kHz
1.2 µV above 30 kHz to 50 kHz
6.8 µV above 50 kHz to 100 kHz
2.2 mV to 22 mV
115 µV/V from 10 Hz to 40Hz
(55 µV/V + 0.2 µV) above 40 Hz to 10 kHz
(60 µV/V + 0.4 µV) above 10 kHz to 20 kHz
(85 µV/V + 0.8 µV) above 20 kHz to 30 kHz
(100 µV/V + 1.1 µV) above 30 kHz to 50 kHz
(360 µV/V + 6 µV) above 50 kHz to 10 0 kHz
22 mV to 220 mV
(55 µV/V + 1 µV) from 10 Hz to 40 Hz
60 µV/V above 40 Hz to 10 kHz
65 µV/V above 10 kHz to 20 kHz
(75 µV/V + 1 µV) above 20 kHz to 30 kHz
(65 µV/V + 2 µV) above 30 kHz to 50 kHz
(175 µV/V + 10 µV) above 50 kHz to 100 kHz
220 mV to 2.2 V
55 µV/V from 10 Hz to 40 Hz
50 µV/V above 40 Hz to 50 kHz
80 µV/V above 10 kHz to 100 kHz
2.2 V to 22 V
55 µV/V from 10 Hz to 20 Hz
50 µV/V above 20 Hz to 50 kHz
75 µV/V above 50 kHz to 100 kHz
2 2 V to 220 V
55 µV/V from 10 Hz to 20 Hz
50 µV/V above 20 Hz to 50 kHz
85 µV/V above 50 kHz to 100 kHz
220 V to 1000 V
40µV/V from 40Hz to 1kHz
0.3% of reading from 0.25 V to 10 V and 100 kHz to 1 MHz

D.C.  current

0.1% of reading + 5 pA from 100 pA to 100 μA
100 μA/A + 5 nA from 100 μA to 100 A
0.1% of reading above 100A to 1,000 A

A.C. current

0.05% of reading from 20 A to 4000 A at 50 Hz
0.05% of reading above 10 μA to 2.2 A from 10 Hz to 10 kHz
0.05% of reading above 2.2 A to 20 A from 20 Hz to 5 kHz

Frequency

2 in 1011  in the range
1 Hz to 10 MHz

Timing characteristics

1 count +1μs

Distortion

Total harmonic distortion

0.021% of reading for Square wave 
0.034% of reading for Sawtooth wave 
0.018% of reading for Sine wave 
for 50 Hz source to the 60th harmonic 




Stroboscopes; Tachometers Rotational speed Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -
Stroboscopes

0.2% of reading + 1 LSD from 60 FPM to 10,000 FPM

Tachometers

0.01% of reading + 1 LSD from 10 RPM to 1000,000 RPM