Tex@Site Pty Ltd

Site

NATA Laboratory

Accreditation No.

15281

Site No.

15902

Date of Accreditation

16 Jul 2007

Address

17 Ceylon Street
Nunawading, VIC 3131
Australia

texatsite.com.au

Contact

Mr Salim Hellis P: +61 1300785935

Availability

Services available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations

Suspended

DC and low frequency electrical metrology - Electrical instrument calibrators
Instrument calibrators Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range up to 90 GΩ

20 μΩ/Ω + 5 μΩ up to 2 Ω; 

14 μΩ/Ω + 67 μΩ from 2 Ω to 20 Ω; 

12 μΩ/Ω + 570 μΩ from 20 Ω to 200 Ω; 

12 μΩ/Ω + 5.7 mΩ from 200 Ω to 2 kΩ; 

12 μΩ/Ω + 51 mΩ from 2 kΩ to 20 kΩ; 

16 μΩ/Ω + 1.4 Ω from 20 kΩ to 200 kΩ; 

17 μΩ/Ω + 14 Ω from 200 kΩ to 2 MΩ; 

49 μΩ/Ω + 590 Ω from 2 MΩ to 20 MΩ; 

170 μΩ/Ω + 19 kΩ from 20 MΩ to 200 MΩ; 

0.15% + 2.6 MΩ from 200 MΩ to 2 GΩ; 

0.35% + 47 MΩ from 2 GΩ to 20 GΩ;

0.75% + 0.056%/GΩ from 20 GΩ to 90 GΩ

AC voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range up to 1000 V

290 μV/V + 20 μV up to 200 mV and from 40 Hz to 100 Hz; 

220 μV/V + 160 μV from 200 mV to 2 V and from 40 Hz to 100 Hz;

220 μV/V + 1.6 mV from 2 V to 20 V and from 40 Hz to 100 Hz; 

220 μV/V + 16 mV from 20 V to 200 V and from 40 Hz to 100 Hz; 

260 μV/V + 140 mV from 200 V to 1000 V and from 40 Hz to 100 Hz;

280 μV/V + 17 μV up to 200 mV and from 100 Hz to 2 kHz; 

190 μV/V + 160 μV from 200 mV to 2 V and from 100 Hz to 2 kHZ; 

190 μV/V + 1.6 mV from 2 V to 20 V and from 100 Hz to 2 kHz; 

190 μV/V + 16 mV from 20 V to 200 V and from 100 Hz to 2 kHz; 

260 μV/V + 140 mV from 200 V to 1000 V and from 100 Hz to 2 kHz;

340 μV/V + 20 μV up to 200 mV and from 2 kHz to 10 kHz;

270 μV/V + 160 μV from 200 mV to 2 V and from 2 kHz to 10 kHz; 

220 μV/V + 1.6 mV from 2 V to 20 V and from 2 kHZ to 10 kHZ; 

220 μV/V + 16 mV from 20 V to 200 V and from 2 kHz to 10 kHz;

260 μV/V + 140 mV from 200 V to 1000 V and from 2 kHz to 10 kHz;

AC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range up to 20 A

780 μA/A + 68 nA up to 200 μA and from 10 Hz to 10 kHz; 

460 μA/A + 470 nA from 200 μA to 2 mA and from 10 Hz to 10 kHz; 

460 μA/A + 4.7 μA from 2 mA to 20 mA and from 10 Hz to 10 kHz; 

690 μA/A + 47 μA from 20 mA to 200 mA and from 10 Hz to 10 kHz; 

0.11% + 1.1mA from 200 mA to 2 A and from 10 Hz to 2 kHz; 

0.13% + 8.2mA from 2 A to 20 A and from 10 Hz to 2 kHz;

0.13% + 1.1 mA from 200 mA to 2 A and from 2 kHz to 10 kHz; 

0.36% + 13 mA from 2 A to 20 A and from 2 kHz to 10 kHz

DC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range up to 20 A

36 μA/A + 6.2 nA up to 200 μA; 

36 μA/A + 62 nA from 200 μA to 2 mA; 

41 μA/A + 620 nA from 2 mA to 20 mA; 

130 μA/A + 4.4 μA from 20 mA to 200 mA; 

460 μA/A + 96 μA from 200 mA to 2 A; 

0.11% + 3.9 mA from 2 A to 20 A

DC voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range up to 1000 V

15 μV/V + 1.7 μV up to 200 mV; 

10 μV/V + 15 μV from 200 mV to 2 V; 

10 μV/V + 120 μV from 2 V to 20 V; 

15 μV/V + 1.5 mV from 20 V to 200 V; 

13 μV/V + 12 mV from 200 V to 1000 V; 

Suspended

DC and low frequency electrical metrology - Electrical measurement and test equipment
Ammeters; Current clamps; Digital multimeters (DMM) AC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range from 69 nA to 20 A
clamp on ammeters only in the range from 20 A to 1000 A

710 μA/A + 69 nA up to 200 μA and from 20 Hz to 1 kHz; 

450 μA/A + 480 nA from 200 μA to 2 mA and from 20 Hz to 1 kHz; 

450 μA/A + 4.8 μA from 2 mA to 20 mA and from 20 Hz to 1 kHz; 

680 μA/A + 48 μA from 20 mA to 200 mA and from 20 Hz to 1 kHz; 

0.21% + 1.1 mA from 200 mA to 2 A and from 20 Hz to 500 Hz; 

0.22% + 8.3 mA from 2 A to 20 A and from 20 Hz to 500 Hz


For on site calibrations

Calibration and Measurement Capability of - 

0.18% + 550 nA from 20 μA to 200 μA and at 57 Hz; 

0.12% + 1.6 μA from 200 μA to 2 mA and at 57 Hz; 

0.12% + 16 μA from 2 mA to 20 mA and at 57 Hz; 

0.13% + 180 μA from 20 mA to 200 mA and at 57 Hz; 

0.26% + 3.5 mA from 200 mA to 2 A and at 57 Hz; 

0.38% + 35 mA from 2 A to 20 A and at 57 Hz


clamp on ammeters only both on site and at the laboratory

With Calibration and Measurement Capability of-

0.52% + 400 mA from 20 A to 100 A and at 57 Hz; 

0.63% + 2.4 A from 100 A to 1000 A and at 57 Hz

DC current Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

including onsite calibrations
in the range from 15 nA up to 20 A
Clamp on ammeters in the range from 20 A to 1000 A

37 μA/A + 15 nA up to 200 μA; 

38 μA/A + 150 nA from 200 μA to 2 mA; 

43 μA/A + 1.5 μA from 2 mA to 20 mA; 

122 μA/A + 14 μA from 20 mA to 200 mA; 

460 μA/A + 97 μA from 200 mA to 2 A; 

0.11% + 3.9 mA from 2 A to 20 A


On site calibrations

with Calibration and Measurement Capability of - 

230 μA/A + 31 nA from 20 μA to 200 μA; 

170 μA/A + 200 nA from 200 μA to 2 mA; 

150 μA/A + 1.9 μA from 2 mA to 20 mA; 

190 μA/A + 27 μA from 20 mA to 200 mA; 

610 μA/A + 710 μA from 200 mA to 2 A; 

0.16% + 20 mA from 2 A to 20 A


clamp on ammeters both on site and at the laboratory

0.48% + 180 mA from 20 A to 100 A;

0.53% + 1.3 A from 100 A to 1000 A



Digital multimeters (DMM); Ohm meters Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

including onsite calibrations
in the range from 7.2 μΩ up to 20 GΩ

20 μΩ/Ω + 7.2 μΩ up to 2 Ω; 

14 μΩ/Ω + 91 μΩ from 2 Ω to 20 Ω; 

12 μΩ/Ω + 810 μΩ from 20 Ω to 200 Ω; 

12 μΩ/Ω + 5.8 mΩ from 200 Ω to 2 kΩ; 

12 μΩ/Ω + 53 mΩ from 2 kΩ to 20 kΩ; 

16 μΩ/Ω + 1.4 Ω from 20 kΩ to 200 kΩ; 

17 μΩ/Ω + 14 Ω from 200 kΩ to 2 MΩ; 

49 μΩ/Ω + 590 Ω from 2 MΩ to 20 MΩ; 

170 μΩ/Ω + 20 kΩ from 19 MΩ to 200 MΩ; 

0.15% + 2.6 MΩ from 200 MΩ to 2 GΩ; 

0.35% + 47 MΩ from 2 GΩ to 20 GΩ;


For on site calibrations

Calibration and Measurement Capability of - 

1.2% from 1 Ω to 10 Ω;

0.6% from 10 Ω to 100 Ω;

0.12% + 50 mΩ from 100 Ω to 1 kΩ;

0.12% + 500 mΩ from 1 kΩ to 10 kΩ;

0.12% + 5 Ω from 10 kΩ to 100 kΩ;

0.12% + 50 Ω from 100 kΩ to 1MΩ;


Insulation resistance testers up to 1kV both on site and at the laboratory

1.9% from 10 MΩ to 100 MΩ;


Insulation resistance testers up to 5kV both on site and at the laboratory
2.6% from 100 MΩ to 200 MΩ;
4% from 200 MΩ to 300 MΩ;
3% from 300 MΩ to 400 MΩ;
2.7% from 400 MΩ to 500 MΩ;
2.5% from 500 MΩ to 1 GΩ;
3.5% from 1 GΩ to 2 GΩ;
5.5% from 2 GΩ to 3 GΩ;
3.7% from 3 GΩ to 4 GΩ;
3.5% from 4 GΩ to 7 GΩ;
3% from 7 GΩ to 10 GΩ;
10% from 10 GΩ to 50 GΩ;
15% from 50 GΩ to 99 GΩ

Digital multimeters (DMM); Voltmeters DC voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range from 2.3 μV up to 1000 V

15 μV/V + 2.3 μV up to 200 mV;

13 μV/V + 21 μV from 200 mV to 2 V;

13 μV/V + 47 μV from 2 V to 20 V;

17 μV/V + 580 μV from 20 V to 200 V;

18 μV/V + 8.7 mV from 200 V to 1000 V


For on site calibrations

Calibration and Measurement Capability of - 

230 μV/V + 7 μV from 2 mV to 20 mV;

43 μV/V + 16 μV from 20 mV to 200 mV;

19 μV/V + 130 μV from 200 mV to 2 V;

16 μV/V + 1.3 mV from 2 V to 20 V;

42 μV/V + 16 mV from 20 V to 200 V;

34 μV/V + 130 mV from 200 V to 1000 V


AC voltage Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

including onsite calibrations
in the range from 23 μV up to 20 V

280 μV/V + 23 μV up to 200 mV and from 40 Hz to 100 Hz;

260 μV/V + 200 μV from 200 mV to 2 V and from 40 Hz to 100 Hz;

260 μV/V + 2 mV from 2 V to 20 V and from 40 Hz to 100 Hz;

210 μV/V + 20 mV from 20 V to 200 V and from 40 Hz to 100 Hz;

240 μV/V + 180 mV from 200 V to 1000 V and from 40 Hz to 100 Hz;

260 μV/V + 23 μV up to 200 mV and from 100 Hz to 2 kHz;

230 μV/V + 200 μV from 200 mV to 2 V and from 100 Hz to 2 kHz;

230 μV/V + 2 mV from 2 V to 20 V and from 100 Hz to 2 kHz;

180 μV/V + 20 mV from 20 V to 200 V and from 100 Hz to 1 kHz;

240 μV/V + 180 mV from 200 V to 1000 V and from 100 Hz to 1 kHz;

330 μV/V + 23 μV up to 200mV and from 2 kHz to 10 kHz;

300 μV/V + 200 μV from 200mV to 2 V and from 2 kHz to 10 kHz;

300 μV/V + 2 mV from 2 V to 20 V and from 2 kHz to 10 kHz;


For onsite calibrations

Calibration and Measurement Capability of - 

0.2% + 150 μV from 2 mV to 20 mV and at 57 Hz;

750 μV/V + 180 μV from 20 mV to 200 mV and at 57 Hz;

570 μV/V + 650 μV from 200 mV to 2 V and at 57 Hz;

570 μV/V + 6.8 mV from 2 V to 20 V and at 57 Hz;

910 μV/V + 68 mV from 20 V to 200 V and at 57 Hz;

0.12% + 340 mV from 200 V to 1000 V and at 57 Hz


Suspended

DC and low frequency electrical metrology - Electrical standards
Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of - 

in the range up to 90 GΩ

20 μΩ/Ω + 5 μΩ up to 2 Ω; 

14 μΩ/Ω + 67 μΩ from 2 Ω to 20 Ω; 

12 μΩ/Ω + 570 μΩ from 20 Ω to 200 Ω; 

12 μΩ/Ω + 5.7 mΩ from 200 Ω to 2 kΩ; 

12 μΩ/Ω + 51 mΩ from 2 kΩ to 20 kΩ; 

16 μΩ/Ω + 1.4 Ω from 20 kΩ to 200 kΩ; 

17 μΩ/Ω + 14 Ω from 200 kΩ to 2 MΩ; 

49 μΩ/Ω + 590 Ω from 2 MΩ to 20 MΩ; 

170 μΩ/Ω + 19 kΩ from 20 MΩ to 200 MΩ;

0.15% + 2.6 MΩ from 200 MΩ to 2 GΩ; 

0.35% + 47 MΩ from 2 GΩ to 20 GΩ; 

0.75% + 0.056%/GΩ from 20 GΩ to 90 GΩ

Suspended

Time and frequency metrology - Frequency, time and waveform measuring equipment
Counters; Frequency meters Frequency Comparison with a reference standard

Capability

With Calibration and Measurement Capability of - 

in the laboratory calibration for Frequency
2.3 Hz/MHz from 10 Hz to 100 kHz;
onsite calibration of Frequency Meters

52 Hz/MHz from 10 Hz to 100 kHz


Tachometers Rotational speed Comparison with a reference standard

Capability

Including on site calibrations

With Calibration and Measurement Capability of-

52 rpm/1,000,000rpm + 1 rpm from 600 rpm to 60,000 rpm

Pressure metrology - Pressure and vacuum measuring equipment Digital pressure gauges; Mechanical pressure gauges; Vacuum gaugesNULL Gauge pressure By pressure calibrator

AS 1349 ;

MSA Test Method 1 ; 

MSA Test Method 2

Capability

Including on site calibrations

With Calibration and Measurement Capability of -

0.05% of reading or 0.01 kPa (whichever is greater) at -95 kPa to 70 MPa

Torque - Torque measuring and testing equipment Torque transducers Torque Gravimetric measurement with reference beam To the method of BS 7882

Capability

With Calibration and Measurement Capability of - 
0.4% of reading from 1.25 N.m to 25 N.m
0.1% of reading from 25 N.m to 1500 N.m
Torque wrenches Torque Comparison with a reference standard

Capability

Including on site calibrations

With Calibration and Measurement Capability of - 

1.0% of reading from 1.25 Nm to 1500 Nm