Thermo Fisher Scientific Australia Pty Ltd

Site

National Service Department

Accreditation No.

2145

Site No.

2138

Date of Accreditation

01 Feb 1983

Address

5 Caribbean Drive
Scoresby, VIC 3179
Australia

thermofisher.com.au

Contact

Mr Ian Walker P: +61 (08) 82088264

Availability

Services available to external clients

Scope of Accreditation

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -  

DC voltage
0.0013% + 3 μV up to 0.1 V
0.0007% above 0.1 V to 1 V 
0.0006% above 1 V to 10 V
0.0009% above 10 V to 100 V 
0.0025% above 100 V to 1000 V


AC voltage
0.04% + 1.1 μV from 0 V to 0.01 V and from 10 Hz to 1 kHz
0.06% + 1.1 μV from 0 V to 0.01 V and from 1 kHz to 20 kHz 
0.15% from 0.01 V to 0.1 V and from 10 Hz to 1 kHz 
0.03% from 0.01 V to 0.1 V and from 1 kHz to 20 kHz 

0.01% from 0.1 V to 1 V and from 10 Hz to 1 kHz 
0.02% from 0.1 V to 1 V and from 1 kHz to 10 kHz 
0.034% from 0.1 V to 1 V and from 10 kHz to 50 kHz 
0.09% from 0.1 V to 1 V and from 50 kHz to 100 kHz 
0.35% from 0.1 V to 1 V and from 100 kHz to 300 kHz 
0.01% from 1 V to 10.0 V and from 10 Hz to 1 kHz 
0.02% from 1 V to 10.0 V and from 1 kHz to 20 kHz 
0.033% from 1 V to 10.0 V and from 20 kHz to 50 kHz 
0.09% from 1 V to 10.0 V and from 50 kHz to 100 kHz 
0.022% from 10 V to 100.0 V and from 10 Hz to 1 kHz 
0.023% from 10 V to 100.0 V and from 1 kHz to 20 kHz 
0.045% from 100 V to 1000 V and from 10 Hz to 1 kHz 


DC current
0.007% + 0.3 nA from 0 μA to 100 μA
0.005% above 100 μA to 1 mA 
0.004% above 1 mA to 10 mA
0.007% above 10 mA to 100 mA
0.017% above 100 mA to 1 A


Resistance
0.003% + 60 μΩ from 0 Ω to 10 Ω
0.002% above 10 Ω to 100 Ω 
0.0013% above 100 Ω to 10 kΩ
0.0015% above 10 kΩ to 100 kΩ
0.003% above 100 kΩ to 1 MΩ
0.007% above 1 MΩ to 10 MΩ
DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Data recorders; Digital multimeters (DMM); Ohm meters; Voltmeters AC voltage; DC current; DC voltage; Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -  

DC voltage
0.0013% + 3μV from 0 V to 0.1 V
0.0013% + 5μV above 0.1 V to 1 V
0.0020% + 100μV above 1 V to 10 V
0.0030% + 2mV above 10 V to 100 V
0.0022% + 2mV above 100 V to 1000 V

AC voltage
0.040% + 15 μV from 0 V to 0.01 V (10 Hz to 1 kHz)
0.055% + 15 μV from 0 V to 0.01 V (1 kHz to 20 kHz)
0.10% + 5 μV from 0.01 V to 0.1 V (10 Hz to 1kHz)
0.050% + 25 μV from 0.01 V to 0.1 V (1 kHz to 20 kHz)
0.010% + 300 μV from 0.1 V to 1 V (10 Hz to 1 kHz)
0.015% + 300 μV from 0.1 V to 1 V (1 kHz to 10 kHz)
0.033% + 800 μV from 0.1 V to 1 V (10 kHz to 50 kHz)
0.085% + 3 mV from 0.1 V to 1 V (50 kHz to 100 kHz)
0.55% + 10 mV from 0.1 V to 1 V (100 kHz to 300 kHz)
0.013% + 3 mV from 1 V to 10 V (10 Hz to 1 kHz)
0.015% + 3 mV from 1 V to 10 V (1 kHz to 20 kHz)
0.033% + 8 mV from 1 V to 10 V (20 kHz to 50 kHz)
0.080% + 30 mV from 1 V to 10 V (50 kHz to 100 kHz)
0.022% + 20 mV from 10 V to 100 V (10 Hz to 1 kHz)
0.022% + 35 mV from 10 V to 100 V (1 kHz to 20 kHz)
0.040% + 0.6 V from 100 V to 1000 V (10 Hz to 1 kHz)


DC current
0.020% + 0.05 μA from 0 μA to 100 μA
0.020% + 0.05 μA above 100 μA to 1 mA
0.020% + 0.05 μA above 1 mA to 10 mA
0.020% + 0.5 μA above 10 mA to 100 mA
0.030% + 3 μA above 100mA to 1A

Resistance
0.0050% + 3 mΩ from 0 Ω to 10 Ω
0.0050% + 70 mΩ above 10 Ω to 100 Ω
0.0050% + 70 mΩ above 100 Ω to 10 kΩ
0.0050% above 10 kΩ to 100 kΩ
0.0050% above 100 kΩ to 1 MΩ
0.019% above 1MΩ to 10 MΩ

DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a reference standard

Capability

with Calibration and Measurement Capability of -  

0.003% + 60 μΩ from 0 Ω to 10 Ω 
0.002% above 10 Ω to 100 Ω 
0.0013% above 100 Ω to 10 kΩ
0.0015% above 10 kΩ to 100 kΩ
0.003% above 100 kΩ to 1 MΩ
0.007% above 1 MΩ to 10 MΩ
Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations Laboratory weighing devices; Precision laboratory balances Mass Gravimetric measurement against reference mass

Capability

Including on site calibrations

with Calibration and Measurement Capability of-


Precision laboratory balances
2.5 in 106 or 20 μg (whichever is greater) in the range up to 10 kg


Industrial balances
1 in 105 in the range from 10 kg to 200 kg

Pressure metrology - Pressure and vacuum measuring equipment Air data test sets; Altimeters Absolute pressure; Gauge pressure Comparison with dead weight tester; Comparison with reference instrument

Capability

With calibration and measurement capability of -  

Pneumatic Gauge
0.0025% or 0.002 kPa (whichever is greater) from -100 kPa to -2 kPa and 2 to 190 kPa
0.0035% or 0.002 kPa (whichever is greater) from 190 kPa to 7 MPa

Pneumatic Absolute
0.0025% or 0.002 kPa (whichever is greater) from 2 kPa to 190 kPa
0.0035% or 0.002 kPa (whichever is greater) from 190 kPa to 7 MPa
Barometers; Digital pressure gauges; Manometers; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gaugesNULL Absolute pressure; Gauge pressure By pressure calibrator AS 1349

Capability

Self-indicating types and transducers with voltage, current and frequency outputs Digital manometers

With Calibration and Measurement Capability of-


Pneumatic Gauge pressure

0.0025% or 0.002 kPa (whichever is greater) in the range from -100 kPa to -2 kPa and 2 kPa to 190 kPa

0.0035% or 0.002 kPa (whichever is greater) in the range from 190 kPa to 7 MPa


Pneumatic Absolute pressure

0.0025% or 0.002 kPa (whichever is greater) in the range from 2 kPa to 190 kPa

0.0035% or 0.002 kPa (whichever is greater) in the range from 190 kPa to 7 MPa


Hydraulic Gauge pressure

0.007% or 0.08 kPa (whichever is greater) in the range from 500 kPa to 70 MPa


Hydraulic Absolute pressure

0.007% or 0.15 kPa (whichever is greater) in the range from 600 kPa to 70 MPa

Temperature metrology - Ancillary temperature measuring equipment Digital voltmeters (DVM) and digital multimeters (DMM); Multi-channel thermocouple data recorders; Process calibrators Resistance to temperature conversion; Volt to temperature conversion Direct measurement by electrical input by the methods of Manual E0211

Capability

with Calibration and Measurement Capability of -  

0.1% of range from -200 °C to 2360 °C
Temperature metrology - Temperature measuring equipment Digital temperature measuring systems Temperature Measurement against reference standard by the methods of Manual E0211

Capability

with Calibration and Measurement Capability of -  

0.05 oC from -40 oC to 300 oC
0.25 oC from 300 oC to 650 oC
Metallic resistance thermometers Resistance over oC range; Temperature Direct temperature and electrical measurement against reference standard; Measurement against reference standard by the methods of Manual E0211

Capability

with Calibration and Measurement Capability of -  

0.05 oC from -40 oC to 300 oC
0.25 oC from 300 oC to 650 oC
Rare metal thermocouples Temperature; mV over oC range Direct temperature and electrical measurement against reference standard; Measurement against reference standard by the methods of Manual E0211

Capability

with Calibration and Measurement Capability of -  

0.5 oC from -40 oC to 650 oC
Temperature; mV over oC range Direct temperature and electrical measurement against reference standard; Measurement against reference standard methods of Manual E0211

Capability

with Calibration and Measurement Capability of -  

0.5 oC from -40 oC to 650 oC
Temperature metrology - Temperature standards and reference equipment Dry block calibrators Temperature To be determined EA 10/13

Capability

with Calibration and Measurement Capability of -  

0.5 oC In the range from -40 oC to 650 oC
Time and frequency metrology - Frequency, time and waveform measuring equipment Counters; Time interval meters Frequency Measurement against reference standard

Capability

with Calibration and Measurement Capability of -  

Counters
0.05% of reading from 1 Hz to 40 Hz 
0.01% of reading from 40 Hz to 10 MHz

Time interval meters
0.01% of reading from 100 ns to 1 s
Frequency meters Frequency Measurement against reference standard

Capability

with Calibration and Measurement Capability of -  

0.05% of reading from 1 Hz to 40 Hz 
0.01% of reading from 40 Hz to 10 MHz
Volume and density - Laboratory volumetric glassware and measures Pipettes; Piston operated volumetric apparatus (POVA) Volume Gravimetric ISO 8655-6

Capability

with Calibration and Measurement Capability of -  

0.070 μL from 0.20 μL to 10 μL

0.080 μL above 10 μL up to 50 μL

0.10 μL above 50 μL up to 100 μL

0.14 μL above 100 μL up to 200 μL

0.20 μL above 200 μL up to 300 μL

0.36 μL above 300 μL up to 500 μL

0.62 μL above 500 μL up to 1000 μL

3.9 μL above 1000 μL up to 5000 μL

6.1 μL above 5000 μL up to 10000 μL

18 μL above 10000 μL up to 30000 μL

35 μL above 30000 μL up to 60000 μL