National Association of Testing Authorities, Australia
Scope of Accreditation
Clean Air Services and Products Pty Ltd
Site
Newcastle Branch
Accreditation No.
18380
Site No.
20296
Date of Accreditation
15 Aug 2023
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | For compliance with AS 2252.1, AS 2252.2 |
Air barrier containment | Not applicable | AS 1807 clause 4.9 | For compliance with AS 2252.2 | ||
Inward air velocity | Not applicable | AS 1807 clause 4.8 | For compliance with AS 2252.1 | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | For compliance with AS 2252.1, AS 2252.2 | ||
Particle count in work zone | Automatic particle counter | AS iso 14644.1 Annex A, section 5 clause A.5 | For compliance with AS 2252.2 | ||
Sound level | Not applicable | AS 1807 clause 4.7 | For compliance with AS 2252.1, AS 2252.2 | ||
Work zone integrity | Not applicable | AS 1807 clause 4.3 | For compliance with AS 2252.2 | ||
Air velocity and uniformity of air velocity | Not applicable | AS 1807 clause 4.1 | For compliance with AS 2252.2 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4 | For compliance with AS 2252.1, AS 2252.2 | ||
Clean rooms | Illuminance | Not applicable | AS 1807. clause 4.5 | ||
Integrity of HEPA filter installations | Cold DOP | AS ISO 14644.3 section 7, clauses B.7.2 & B.7.4, AS 1807.7 (2000) | |||
Air flow direction and visualization | Not applicable | AS ISO 14644.3 section 3, clause B.3.3.4 | |||
Leakage | Cold DOP | AS ISO 14644.3 section 8, clause B.8.2.2 | |||
Air velocity and uniformity of air velocity | Not applicable | AS ISO 14644.3 Section 2, clauses B.2.2.2, B.2.2.5 and B.2.3.4 | |||
Air pressure | Not applicable | AS ISO 14644.3 section 1, clause B.1.2 | |||
Sound level | Not applicable | AS 1807.16 (2000) | |||
Clean workstations | Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4, | For compliance with AS 2252.6 | |
Sound level | Not applicable | AS 1807 clause 4.7 | For compliance with AS 2252.6 | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | For compliance with AS 2252.6 | ||
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | For compliance with AS 2252.6 | ||
Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Annex A section 5, clause A.5 | For compliance with AS 2252.6 | ||
Work zone integrity | Cold DOP | AS 1807 clause 4.3 | For compliance with AS 2252.6 | ||
Air velocity and uniformity of air velocity | Not applicable | AS 1807 clause 4.1 | For compliance with AS 2252.6 | ||
Cytotoxic drug safety cabinets | Work zone integrity | Not applicable | AS 1807 clause 4.3 | For compliance with AS 2252.5 | |
Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4 | For compliance with:AS 2252.5 | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | For compliance with 2252.5 | ||
Air barrier containment | Cold DOP | AS 1807 clause 4.9 | For compliance with AS 2252.5 | ||
Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Annex A, section 5 clause A.5 | For compliance with AS 2252.5 | ||
Sound level | Not applicable | AS 1807 clause 4.7 | For compliance with:AS 2252.5 | ||
Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Annex A section 5 Clause A5 | For compliance with, AS 2252.5 | ||
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | For compliance with AS 2252.5 | ||
Air velocity and uniformity of air velocity | Not applicable | AS 1807 clause 4.1 | For compliance with AS 2252.5 | ||
Airflow smoke patterns | Not applicable | AS 1807 clause 4.13 | For compliance with AS 2252.5 | ||
Fume cupboards and recirculating fume cabinets | Extraction performance | Face velocity | AS/NZS 2243.8 Appendices A and B; AS/NZS 2243.9 Appendices A and B | For compliance with AS/NZS 2243.8, AS/NZS 2243.9 | |
Illuminance | Not applicable | AS 1807.15 (2000) | For compliance with AS/NZS 2243.8, AS/NZS 2243.9 | ||
Extraction performance | Smoke visualisation | AS/NZS 2243.8 Appendix A, AS/NZS 2243.9 Appendix A | For compliance with AS/NZS 2243.8, AS/NZS 2243.9 | ||
Sound level | Not applicable | AS 1807.20 (2000) | For compliance with AS/NZS 2243.8, AS/NZS 2243.9 | ||
Pharmaceutical isolators | Air velocity and uniformity of air velocity | Not applicable | AS 1807 clause 4.1 | For compliance with AS 4237 | |
Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Annex A, section 5, clause A.5 | For compliance with AS 4273 | ||
Air pressure | Not applicable | AS ISO 14644.1 section1, clause B.1.2, | For compliance with AS 4237 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4, AS 1807.7 (2000) | For compliance with AS 4237 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
18380
Site No.
20296
Print date
10 Sep 2024