National Association of Testing Authorities, Australia
Scope of Accreditation
Fotometrix Pty Ltd
Site
Particle Counter and Aerosol Photometer Calibration Facility
Accreditation No.
2253
Site No.
2246
Date of Accreditation
11 Nov 1987
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Optical metrology - Particle measuring devices | Aerosol photometers | Flow rate; Linearity | Comparison with a reference meter | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Calibration of aerosol photometers with polystyrene latex (psl) particles with Calibration and Measurement Capability of - | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Concentration; Flow rate | Comparison with a reference meter; Gravimetric | In-house method including compliance with AS 1807 and ISO 14644.3 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability ofConcentration 2.1 μg/L for the range of 10 to 100 μg/L using polyalphaolefin (PA04) droplets generated to meet the requirements of AS 1807 and ISO 14644.3 Flow rate From 0.05 L/min at a flow rate of 2.83 L/min increasing to 0.20 L/min at a flow rate of 55 L/min, and From 1.11 L/min at a flow rate of 75 L/min increasing to 1.22 L/min at a flow rate of 120 L/min. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Particle counters - Air | Flow rate; Particle count; Particle size | Comparison with a reference meter; Comparison with a reference standard | ISO 21501-4 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Light Scattering Airborne Counters (LSAPCs)using polystyrene latex particles of nominal sizes 0.3 μm to 10 μm at flow rates of 2.83 L/min to 120 L/min with Calibration and Measurement Capability of - Particle size - 0.0008 µm + 0.019D for particles 0.30 µm to 0.99 µm diameter (where D is particle diameter in µm). 0.011 µm + 0.0083D for particles 1.00 µm to 6.00 µm diameter (where D is particle diameter in µm). 0.51 µm for particles 10.00 µm diameter. Flow rate - From 0.05 L/min at a flow rate of 2.83 L/min increasing to 0.20 L/min at a flow rate of 55 L/min, and From 1.11 L/min at a flow rate of 75 L/min increasing to 1.22 L/min at a flow rate of 120 L/min. |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
2253
Site No.
2246
Print date
27 Nov 2024