S W Hart & Co

Site

S W Hart & Co

Accreditation No.

11238

Site No.

11231

Date of Accreditation

20 Dec 1995

Address

19 Garnet Way
Maddington, WA 6109
Australia

swhart.com.au

Contact

Availability

Services available to external clients

Scope of Accreditation

S W Hart & Co

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Work zone integrity Cold DOP AS 1807 clause 4.3 for compliance with AS 2252.2
Illuminance Not applicable

AS 1807 clause 4.5

for compliance to AS 2252.1, AS 2252.2
Air barrier containment Cold DOP AS 1807 clause 4.9 for compliance with AS 2252.2
Air velocity and uniformity of air velocity Not applicable AS 1807 clause 4.1 for compliance with AS 2252.2
Integrity of HEPA filter installations Cold DOP AS 1807.7, AS 1807 clause 4.4 for compliance with AS 2252.1, AS 2252.2
Ultraviolet intensity Not applicable AS 1807 clause 4.10 for compliance with AS 2252.1, AS 2252.2
Sound level Not applicable AS 1807 clause 4.7 for compliance with AS 2252.1, AS 2252.2
Inward air velocity Not applicable AS 1807 clause 4.8 for compliance with AS 2252.1
Clean rooms Air pressure Not applicable AS ISO 14644.3 method B1
Integrity of HEPA filter installations Cold DOP AS 1807.7, AS ISO 14644.3 method B7.2
Particle count in work zone Automatic particle counter AS ISO 14644.1 Appendix A
Sound level Not applicable AS 1807.16
Air velocity and uniformity of air velocity Not applicable AS ISO 14644.3 method B2
Clean workstations; Pharmaceutical isolators Illuminance Not applicable AS 1807 clause 4.5 for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Air velocity and uniformity of air velocity Not applicable AS 1807 clause 4.1 for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Work zone integrity Cold DOP AS 1807 clause 4.3 for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Ultraviolet intensity Not applicable AS 1807 clause 4.10 for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Air barrier containment Cold DOP

AS 1807 clause 4.9

for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Sound level Not applicable AS 1807 Clause 4.7 for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Particle count in work zone Automatic particle counter AS ISO 14644.1 Appendix A for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Integrity of HEPA filter installations Cold DOP AS 1807.7, AS 1807 clause 4.4 for compliance with AS 2252.6 (Clean workstations), AS 4273 (Pharmaceutical isolators)
Cytotoxic drug safety cabinets Air velocity and uniformity of air velocity Not applicable AS 1807 clause 4.1 for compliance with AS 2252.5
Work zone integrity Cold DOP AS 1807 clause 4.3 for compliance with AS 2252.5
Sound level Not applicable AS 1807 clause 4.7 for compliance with AS 2252.5
Illuminance Not applicable AS 1807 clause 4.5 for compliance with AS 2252.5
Integrity of HEPA filter installations Cold DOP AS 1807 clause 4.4 for compliance with AS 2252.5
Ultraviolet intensity Not applicable AS 1807 clause 4.10 for compliance with AS 2252.5
Air barrier containment Cold DOP AS 1807 clause 4.9 for compliance to AS 2252.5
Fume cupboards and recirculating fume cabinets Illuminance Not applicable AS 1807 clause 4.5 for compliance with AS/NZS 2243.8, AS/NZS 2243.9
Sound level Not applicable AS 1807 clause 4.7 for compliance with AS/NZS 2243.8, AS/NZS 2243.9
Extraction performance Face velocity AS/NZS 2243.8 Appendix B; AS/NZS 2243.9 Appendix B for compliance with AS/NZS 2243.8, AS/NZS 2243.9
Extraction performance Smoke visualisation

AS/NZS 2243.8 Appendix A; AS/NZS 2243.9 Appendix A 

for compliance with AS/NZS 2243.8, AS/NZS 2243.9

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.