Scientific Devices Australia Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement

Capability

with Calibration and Measurement Capability of -

DC Voltage
5 μV/V at 1 V and 10 V
10 μV/V + 0.25 μV from 0 to 1 kV

AC Voltage
130 µV/V + 4 µV from 1 mV to 200 mV and 20 Hz to 40 Hz
110 µV/V + 4 µV from 1 mV to 200 mV and 40 Hz to 100 Hz
105 µV/V + 2 µV from 1 mV to 200 mV and 100 Hz to 2 kHz
105 µV/V + 4 µV from 1 mV to 200 mV and 2 kHz to 10 kHz
305 µV/V + 8 µV from 1 mV to 200 mV and 10 kHz to 30 kHz
705 µV/V + 20 µV from 1 mV to 200 mV and 30 kHz to 100 kHz
105 µV/V + 20 µV from 200 mV to 2 V and 20 Hz to 40 Hz
85 µV/V + 20 µV from 200 mV to 2 V and 40 Hz to 100 Hz
65 µV/V + 20 µV from 200 mV to 2 V and 100 Hz to 2 kHz
85 µV/V + 20 µV from 200 mV to 2 V and 2 kHz to 10 kHz
205 µV/V + 40 µV from 200 mV to 2 V and 10 kHz to 30 kHz
505 µV/V + 200 µV from 200 mV to 2 V and 30 kHz to 100 kHz
0.3% + 2 mV from 200 mV to 2 V and 100 kHz to 300 kHz
1% + 20 mV from 200 mV to 2 V and 300 kHz to 1 MHz
105 µV/V + 200 µV from 2 V to 20 V and 20 Hz to 40 Hz
85 µV/V + 200 µV from 2 V to 20 V and 40 Hz to 100 Hz
65 µV/V + 200 µV from 2 V to 20 V and 100 Hz to 2 kHz
85 µV/V + 200 µV from 2 V to 20 V and 2 kHz to 10 kHz
205 µV/V + 400 µV from 2 V to 20 V and 10 kHz to 30 kHz
505 µV/V + 2 mV from 2 V to 20 V and 30 kHz to 100 kHz
0.3% + 20 mV from 2 V to 10 V and 100 kHz to 300 kHz
1% + 200 mV from 2 V to 10 V and 300 kHz to 1 MHz
105 µV/V + 2 mV from 20 V to 200 V and 20 Hz to 40 Hz
85 µV/V + 2 mV from 20 V to 200 V and 40 Hz to 100 Hz
65 µV/V + 2 mV from 20 V to 200 V and 100 Hz to 2 kHz
85 µV/V + 2 mV from 20 V to 200 V and 2 kHz to 10 kHz
205 µV/V + 4 mV from 20 V to 200 V and 10 kHz to 30 kHz
505 µV/V + 20 mV from 20 V to 200 V and 30 kHz to 100 kHz
95 µV/V + 40 mV from 200 V to 1 kV and 40 Hz to 10 kHz
from 300 V to 1 kV, + 0.000 4 x (reading – 300)2 µV/V
205 µV/V + 80 mV from 200 V to 1 kV and 10 kHz to 30 kHz
from 300 V to 1 kV, + (0.000 4 + (Frequency – 10 000) x 107) x (reading – 300)2 µV/V
105 µV/V + 2 mV from 20 V to 200 V and 20 Hz to 40 Hz
85 µV/V + 2 mV from 20 V to 200 V and 40 Hz to 100 Hz
65 µV/V + 2 mV from 20 V to 200 V and 100 Hz to 2 kHz
85 µV/V + 2 mV from 20 V to 200 V and 2 kHz to 10 kHz
205 µV/V + 4 mV from 20 V to 200 V and 10 kHz to 30 kHz
505 µV/V + 20 mV from 20 V to 200 V and 30 kHz to 100 kHz
95 µV/V + 40 mV from 200 V to 300 V and 40 Hz to 10 kHz
95 µV/V + (0.000 4 x (reading – 300)2 µV/V) + 40 mV from 300 V to 1 kV and 40 Hz to 10 kHz
205 µV/V + 80 mV from 200 V to 300 V and 10 kHz to 30 kHz
(0.000 4 x (reading – 300)2 µV/V) from 300 V to 1 kV and 10 kHz to 30 kHz
205 µV/V + (0.000 4 x (reading – 300)2 µV/V) + 80 mV from 300 V to 1 kV and 10 kHz to 30 kHz

DC Current
0.05% + 0.1 pA from zero to 2 nA
0.02% + 0.6 pA from 2 nA to 20 nA
0.02% + 1 pA from 20 nA to 200 nA
0.02% + 10 pA from 200 nA to 2 μA
50 µA/A + 0.4 nA from 2 µA to 200 µA
50 µA/A + 2 nA from 200 µA to 2 mA
50 µA/A + 20 nA from 2 mA to 20 mA
50 µA/A + 0.2 µA from 20 mA to 200 mA
100 µA/A + 4 µA from 200 mA to 2 A
100 µA/A + 20 µA from 2 A to 10 A

AC Current
475 µA/A +20 nA from 2 µA to 200 µA and 20 Hz to 5 kHz
280 µA/A + 200 nA from 200 µA to 2 mA and 20 Hz to 5 kHz
280 µA/A + 2 µA from 2 mA to 20 mA and 20 Hz to 5 kHz
250 µA/A + 20 µA from 20 mA to 200 mA and 20 Hz to 5 kHz
0.015% from 200 mA to 2 A and 20 Hz to 1 kHz
0.11% from 200 mA to 2 A and 1 kHz to 5 kHz
0.025% from 2 A to 10 A and 10 Hz to 1 kHz
0.05% from 2 A to 10 A and 1 kHz to 5 kHz

Resistance
10 μΩ/Ω + 25 μΩ from 0 Ω  to 20 kΩ
15 μΩ/Ω from 20 kΩ to 200 kΩ
20 μΩ/Ω from 200 kΩ to 2 MΩ
30 μΩ/Ω from 2 MΩ to 20 MΩ
60 µΩ/Ω + 10 kΩ from 20 MΩ to 100 MΩ
0.2% from 100 MΩ to 10 GΩ


DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Data recorders; Digital multimeters (DMM); Ohm meters; Voltmeters AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

DC Voltage
5 μV/V at 1 V and 10 V
10 μV/V + 0.25 μV from zero to 1 kV

AC Voltage 
130 µV/V + 4 µV from 1 mV to 200 mV and 20 Hz to 40 Hz
110 µV/V + 4 µV from 1 mV to 200 mV and 40 Hz to 100 Hz
105 µV/V + 2 µV from 1 mV to 200 mV and 100 Hz to 2 kHz
105 µV/V + 4 µV from 1 mV to 200 mV and 2 kHz to 10 kHz
305 µV/V + 8 µV from 1 mV to 200 mV and 10 kHz to 30 kHz
705 µV/V + 20 µV from 1 mV to 200 mV and 30 kHz to 100 kHz
105 µV/V + 20 µV from 200 mV to 2 V and 20 Hz to 40 Hz
85 µV/V + 20 µV from 200 mV to 2 V and 40 Hz to 100 Hz
65 µV/V + 20 µV from 200 mV to 2 V and 100 Hz to 2 kHz
85 µV/V + 20 µV from 200 mV to 2 V and 2 kHz to 10 kHz
205 µV/V + 40 µV from 200 mV to 2 V and 10 kHz to 30 kHz
505 µV/V + 200 µV from 200 mV to 2 V and 30 kHz to 100 kHz
0.3% + 2 mV from 200 mV to 2 V and 100 kHz to 300 kHz
1% + 20 mV from 200 mV to 2 V and 300 kHz to 1 MHz
105 µV/V + 200 µV from 2 V to 20 V and 20 Hz to 40 Hz
85 µV/V + 200 µV from 2 V to 20 V and 40 Hz to 100 Hz
65 µV/V + 200 µV from 2 V to 20 V and 100 Hz to 2 kHz
85 µV/V + 200 µV from 2 V to 20 V and 2 kHz to 10 kHz
205 µV/V + 400 µV from 2 V to 20 V and 10 kHz to 30 kHz
505 µV/V + 2 mV from 2 V to 20 V and 30 kHz to 100 kHz
0.3% + 20 mV from 2 V to 10 V and 100 kHz to 300 kHz
1% + 200 mV from 2 V to 10 V and 300 kHz to 1 MHz
 
105 µV/V + 2 mV from 20 V to 200 V and 20 Hz to 40 Hz
85 µV/V + 2 mV from 20 V to 200 V and 40 Hz to 100 Hz
65 µV/V + 2 mV from 20 V to 200 V and 100 Hz to 2 kHz
85 µV/V + 2 mV from 20 V to 200 V and 2 kHz to 10 kHz
205 µV/V + 4 mV from 20 V to 200 V and 10 kHz to 30 kHz
505 µV/V + 20 mV from 20 V to 200 V and 30 kHz to 100 kHz
95 µV/V + 40 mV from 200 V to 1 kV and 40 Hz to 10 kHz
at >300 V, + 0.000 4 x (reading – 300)2 µV/V
205 µV/V + 80 mV from 200 V to 1 kV and 10 kHz to 30 kHz
at >300 V, + (0.000 4 + (Frequency – 10 000) x 107) x (reading – 300)2 µV/V

DC Current
0.05% + 0.1 pA from zero to 2 nA
0.02% + 0.6 pA from 2 nA to 20 nA
0.02% + 1 pA from 20 nA to 200 nA
0.02% + 10 pA from 200 nA to 2 μA
50 µA/A + 0.4 nA from 2 µA to 200 µA
50 µA/A + 2 nA from 200 µA to 2 mA
50 µA/A + 20 nA from 2 mA to 20 mA
50 µA/A + 0.2 µA from 20 mA to 200 mA
100 µA/A + 4 µA from 200 mA to 2 A
100 µA/A + 20 µA from 2 A to 10 A

AC Current
475 µA/A +20 nA from 2 µA to 200 µA and 20 Hz to 5 kHz
280 µA/A + 200 nA from 200 µA to 2 mA and 20 Hz to 5 kHz
280 µA/A + 2 µA from 2 mA to 20 mA and 20 Hz to 5 kHz
250 µA/A + 20 µA from 20 mA to 200 mA and 20 Hz to 5 kHz
0.015% from 200 mA to 2 A and 20 Hz to 1 kHz
0.11% from 200 mA to 2 A and 1 kHz to 5 kHz
0.025% from 2 A to 10 A and 10 Hz to 1 kHz
0.05% from 2 A to 10 A and 1 kHz to 5 kHz

Resistance
10 μΩ/Ω + 25 μΩ from 0 Ω to 20 kΩ
15 μΩ/Ω from 20 kΩ to 200 kΩ
20 μΩ/Ω from 200 kΩ to 2 MΩ
30 μΩ/Ω from 2 MΩ to 20 MΩ
60 µΩ/Ω + 10 kΩ from 20 MΩ to 100 MΩ
0.2% from 100 MΩ to 10 GΩ


DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Direct measurement

Capability

with Calibration and Measurement Capability of -

10 μΩ/Ω + 25 μΩ from 0 Ω to 20 kΩ
15 μΩ/Ω from 20 kΩ to 200 kΩ
20 μΩ/Ω from 200 kΩ to 2 MΩ
30 μΩ/Ω from 2 MΩ to 20 MΩ
60 µΩ/Ω + 10 kΩ from 20 MΩ to 100 MΩ
0.2% from 100 MΩ to 10 GΩ

Time and frequency metrology - Frequency and time standards Signal sources Frequency Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

1 part in 108  from 0.1 Hz  to 25 GHz

Time and frequency metrology - Frequency, time and waveform measuring equipment Counters; Frequency meters; Time interval meters Frequency; Time interval Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

Time interval
0.3 ms from 20 ms to 5 s

Frequency
1 part in 108 from 0;.1 o 25 Ghz

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.