Scientific Devices Australia Pty Ltd

Site

Scientific Devices Australia Pty Ltd

Accreditation No.

3686

Site No.

3679

Date of Accreditation

09 Jun 2023

Address

118 Atkinson Street
Oakleigh, VIC 3166
Australia

scientific-devices.com.au

Contact

Mr Laurence Motteram P: +61 (03) 95691366

Availability

Services available to external clients

Scope of Accreditation

Scientific Devices Australia Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical instrument calibrators Instrument calibrators AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement

Capability

with Calibration and Measurement Capability of -

DC Voltage
5 μV/V at 1 V and 10 V
10 μV/V + 0.25 μV from 0 mV to 1 kV

AC Voltage
130 µV/V + 4 µV from 1 mV to 200 mV and 20 Hz to 40 Hz
110 µV/V + 4 µV from 1 mV to 200 mV and 40 Hz to 100 Hz
110 µV/V + 2 µV from 1 mV to 200 mV and 100 Hz to 2 kHz
110 µV/V + 4 µV from 1 mV to 200 mV and 2 kHz to 10 kHz
310 µV/V + 8 µV from 1 mV to 200 mV and 10 kHz to 30 kHz
710 µV/V + 20 µV from 1 mV to 200 mV and 30 kHz to 100 kHz
110 µV/V + 20 µV above 200 mV to 2 V and 20 Hz to 40 Hz
85 µV/V + 20 µV above 200 mV to 2 V and 40 Hz to 100 Hz
65 µV/V + 20 µV above 200 mV to 2 V and 100 Hz to 2 kHz
85 µV/V + 20 µV above 200 mV to 2 V and 2 kHz to 10 kHz
210 µV/V + 40 µV above 200 mV to 2 V and 10 kHz to 30 kHz
510 µV/V + 200 µV above 200 mV to 2 V and 30 kHz to 100 kHz
0.3% + 2 mV above 200 mV to 2 V and 100 kHz to 300 kHz
1% + 20 mV above 200 mV to 2 V and 300 kHz to 1 MHz
110 µV/V + 200 µV above 2 V to 20 V and 20 Hz to 40 Hz
85 µV/V + 200 µV above 2 V to 20 V and 40 Hz to 100 Hz
65 µV/V + 200 µV above 2 V to 20 V and 100 Hz to 2 kHz
85 µV/V + 200 µV above 2 V to 20 V and 2 kHz to 10 kHz
210 µV/V + 400 µV above 2 V to 20 V and 10 kHz to 30 kHz
510 µV/V + 2 mV above 2 V to 20 V and 30 kHz to 100 kHz
0.3% + 20 mV above 2 V to 10 V and 100 kHz to 300 kHz
1% + 200 mV above 2 V to 10 V and 300 kHz to 1 MHz
110 µV/V + 2 mV above 20 V to 200 V and 20 Hz to 40 Hz
85 µV/V + 2 mV above 20 V to 200 V and 40 Hz to 100 Hz
65 µV/V + 2 mV above 20 V to 200 V and 100 Hz to 2 kHz
85 µV/V + 2 mV above 20 V to 200 V and 2 kHz to 10 kHz
210 µV/V + 4 mV above 20 V to 200 V and 10 kHz to 30 kHz
510 µV/V + 20 mV above 20 V to 200 V and 30 kHz to 100 kHz
95 µV/V + 40 mV above 200 V to 300 V and 40 Hz to 10 kHz
95 µV/V + (0.000 4 x (reading – 300)² µV/V) + 40 mV above 300 V to 1 kV and 40 Hz to 10 kHz
210 µV/V + 80 mV above 200 V to 300 V and 10 kHz to 30 kHz
210 µV/V + (0.000 4 + (Frequency – 10 000) x 10-⁷) x (reading – 300)²[DD1] [LM2] µV/V) + 80 mV above 300 V to 1 kV and 10 kHz to 30 kHz  

DC Current
0.05% + 0.1 pA from 0 nA to 2 nA
0.02% + 0.6 pA above 2 nA to 20 nA
0.02% + 1 pA above 20 nA to 200 nA
0.02% + 10 pA above 200 nA to 2 μA
50 µA/A + 0.4 nA above 2 µA to 200 µA
50 µA/A + 2 nA above 200 µA to 2 mA
50 µA/A + 20 nA above 2 mA to 20 mA
50 µA/A + 0.2 µA above 20 mA to 200 mA
100 µA/A + 4 µA above 200 mA to 2 A
100 µA/A + 20 µA above 2 A to 10 A  

AC Current
480 µA/A +20 nA from 2 µA to 200 µA and 20 Hz to 5 kHz
280 µA/A + 200 nA above 200 µA to 2 mA and 20 Hz to 5 kHz
280 µA/A + 2 µA above 2 mA to 20 mA and 20 Hz to 5 kHz
250 µA/A + 20 µA above 20 mA to 200 mA and 20 Hz to 5 kHz
0.015% above 200 mA to 2 A and 20 Hz to 1 kHz
0.11% above 200 mA to 2 A and 1 kHz to 5 kHz
0.025% above 2 A to 10 A and 10 Hz to 1 kHz
0.05% above 2 A to 10 A and 1 kHz to 5 kHz

Resistance
10 μΩ/Ω + 25 μΩ from 0 Ω to 20 kΩ
15 μΩ/Ω above 20 kΩ to 200 kΩ
20 μΩ/Ω above 200 kΩ to 2 MΩ
30 μΩ/Ω above 2 MΩ to 20 MΩ
60 µΩ/Ω + 10 kΩ above 20 MΩ to 100 MΩ
0.2% above 100 MΩ to 10 GΩ


DC and low frequency electrical metrology - Electrical measurement and test equipment Ammeters; Data recorders; Digital multimeters (DMM); Ohm meters; Voltmeters AC current; AC voltage; DC current; DC voltage; Resistance Direct measurement against a reference standard

Capability

with Calibration and Measurement Capability of -

DC Voltage
5 μV/V at 1 V and 10 V
10 μV/V + 0.25 μV from 0 mV to 1 kV

AC Voltage 
130 µV/V + 4 µV from 1 mV to 200 mV and 20 Hz to 40 Hz
110 µV/V + 4 µV from 1 mV to 200 mV and 40 Hz to 100 Hz
110 µV/V + 2 µV from 1 mV to 200 mV and 100 Hz to 2 kHz
110 µV/V + 4 µV from 1 mV to 200 mV and 2 kHz to 10 kHz
310 µV/V + 8 µV from 1 mV to 200 mV and 10 kHz to 30 kHz
705 µV/V + 20 µV from 1 mV to 200 mV and 30 kHz to 100 kHz
110 µV/V + 20 µV above 200 mV to 2 V and 20 Hz to 40 Hz
85 µV/V + 20 µV above 200 mV to 2 V and 40 Hz to 100 Hz
65 µV/V + 20 µV above 200 mV to 2 V and 100 Hz to 2 kHz
85 µV/V + 20 µV above 200 mV to 2 V and 2 kHz to 10 kHz
210 µV/V + 40 µV above 200 mV to 2 V and 10 kHz to 30 kHz
510 µV/V + 200 µV above 200 mV to 2 V and 30 kHz to 100 kHz
0.3% + 2 mV above 200 mV to 2 V and 100 kHz to 300 kHz
1% + 20 mV above 200 mV to 2 V and 300 kHz to 1 MHz
110 µV/V + 200 µV above 2 V to 20 V and 20 Hz to 40 Hz
85 µV/V + 200 µV above 2 V to 20 V and 40 Hz to 100 Hz
65 µV/V + 200 µV above 2 V to 20 V and 100 Hz to 2 kHz
85 µV/V + 200 µV above 2 V to 20 V and 2 kHz to 10 kHz
210 µV/V + 400 µV above 2 V to 20 V and 10 kHz to 30 kHz
510 µV/V + 2 mV above 2 V to 20 V and 30 kHz to 100 kHz
0.3% + 20 mV above 2 V to 10 V and 100 kHz to 300 kHz
1% + 200 mV above 2 V to 10 V and 300 kHz to 1 MHz
110 µV/V + 2 mV above 20 V to 200 V and 20 Hz to 40 Hz
85 µV/V + 2 mV above 20 V to 200 V and 40 Hz to 100 Hz
65 µV/V + 2 mV above 20 V to 200 V and 100 Hz to 2 kHz
85 µV/V + 2 mV above 20 V to 200 V and 2 kHz to 10 kHz
210 µV/V + 4 mV above 20 V to 200 V and 10 kHz to 30 kHz
510
µV/V + 20 mV above 20 V to 200 V and 30 kHz to 100 kHz
95 µV/V + 40 mV above 200 V to 300 V and 40 Hz to 10 kHz
95 µV/V + (0.000 4 x (reading – 300)² µV/V) + 40 mV above 300 V to 1 kV and 40 Hz to 10 kHz
210 µV/V + 80 mV above 200 V to 300 V and 10 kHz to 30 kHz
210 µV/V + (0.000 4 + (Frequency – 10 000) x 10-⁷) x (reading – 300)² µV/V) + 80 mV above 300 V to 1 kV and 10 kHz to 30 kHz

DC Current
0.05% + 0.1 pA from zero to 2 nA
0.02% + 0.6 pA above 2 nA to 20 nA
0.02% + 1 pA above 20 nA to 200 nA
0.02% + 10 pA above 200 nA to 2 μA
50 µA/A + 0.4 nA above 2 µA to 200 µA
50 µA/A + 2 nA above 200 µA to 2 mA
50 µA/A + 20 nA above 2 mA to 20 mA
50 µA/A + 0.2 µA above 20 mA to 200 mA
100 µA/A + 4 µA above 200 mA to 2 A
100 µA/A + 20 µA above 2 A to 10 A

AC Current
480µA/A +20 nA from 2 µA to 200 µA and 20 Hz to 5 kHz
280 µA/A + 200 nA above 200 µA to 2 mA and 20 Hz to 5 kHz
280 µA/A + 2 µA above 2 mA to 20 mA and 20 Hz to 5 kHz
250 µA/A + 20 µA above 20 mA to 200 mA and 20 Hz to 5 kHz
0.015% above 200 mA to 2 A and 20 Hz to 1 kHz
0.11% above 200 mA to 2 A and 1 kHz to 5 kHz
0.025% above 2 A to 10 A and 10 Hz to 1 kHz
0.05% above 2 A to 10 A and 1 kHz to 5 kHz

Resistance
10 μΩ/Ω + 25 μΩ from 0 Ω to 20 kΩ
15 μΩ/Ω above 20 kΩ to 200 kΩ
20 μΩ/Ω above 200 kΩ to 2 MΩ
30 μΩ/Ω above 2 MΩ to 20 MΩ
60 µΩ/Ω + 10 kΩ above 20 MΩ to 100 MΩ
0.2% above 100 MΩ to 10 GΩ


DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Direct measurement

Capability

with Calibration and Measurement Capability of -

10 μΩ/Ω + 25 μΩ from 0 Ω to 20 kΩ
15 μΩ/Ω above 20 kΩ to 200 kΩ
20 μΩ/Ω above 200 kΩ to 2 MΩ
30 μΩ/Ω above 2 MΩ to 20 MΩ
60 µΩ/Ω + 10 kΩ above 20 MΩ to 100 MΩ
0.2% above 100 MΩ to 10 GΩ

Time and frequency metrology - Frequency and time standards Signal sources Frequency Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

1 part in 108  from 0.1 Hz  to 25 GHz

Time and frequency metrology - Frequency, time and waveform measuring equipment Counters; Frequency meters; Time interval meters Frequency; Time interval Measurement against reference standard

Capability

with Calibration and Measurement Capability of -

Time interval
0.3 ms from 20 ms to 5 s

Frequency
1 part in 108 from 0.1 Hz to 25 Ghz

The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

The Product column may also describe an Item or Material.

The Limitations column may also describe the Range.