National Association of Testing Authorities, Australia
Scope of Accreditation
National Measurement Institute (NMI)
Site
Temperature Standards Project
Accreditation No.
1
Site No.
14108
Date of Accreditation
03 Aug 1999
Contact
Availability
Services conditionally available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
| Service | Product | Determinant | Technique | Procedure | Limitations | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Temperature metrology - Ancillary temperature measuring equipment | Digital voltmeters (DVM) and digital multimeters (DMM); Multi-channel thermocouple data recorders; Process calibrators | Volt to temperature conversion | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of thermocouple indicators (simulation only)With Calibration and Measurement Capability of - Type T 0.09 °C from 0 °C to 400 °C Type K, J, and N 0.09 °C from 0 °C to 1100 °C Type S and R 0.12 °C from 0 °C to 1100 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Ancillary temperature measuring equipment | Resistance bridges | Resistance to temperature conversion | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of bridges for temperature measurement With Calibration and Measurement Capability of - 0.1 μΩ/Ω for linearity at 1 Ω, 10 Ω and 100 Ω 1 in 107 for ratio in the range 0 to 5 |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Ancillary temperature measuring equipment | Type R extension wire; Type S extension wire | Volt to temperature conversion | Direct measurement by electrical input | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - 0.22 μV from 0 °C to 50 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Humidity measuring equipment | Dew point meters; Hygrometers; Psychrometers | Dew point; Temperature | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of dew point meters |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Humidity measuring equipment | Relative humidity sensors | Relative humidity (RH); Temperature | Comparison with a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of relative humidity sensors
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Non-contact temperature measuring equipment | Clinical infrared thermometers; Radiation pyrometers | Temperature | Direct measurement against a black body source | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - Radiation pyrometers from 0.1 °C at −40 °C to 0.3 °C at 250 °C
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Non-contact temperature measuring equipment | Thermal imaging systems | Temperature | Direct measurement against a black body source | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of thermal imaging cameras With Calibration and Measurement Capability of - from 0.1 °C at −40 °C to 0.5 °C at 250 °C IEC 80601-2-59 clauses 201.101.3.1 Performance 201.101.5 Minimum resolvable temperature difference 201.101.6 Uniformity of Workable Target Plane 201.101.9 Spatial resolution SS 582-1 (Singapore) clauses 6.9.5, 6.9.6, 6.9.7, 6.9.8, 6.9.9 |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature measuring equipment | Base metal thermocouples | mV over oC range | Direct temperature and electrical measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - Type T 1.0 μV from −200 °C to 0 °C (1.0 + 0.01% E) μV from above 0 °C to 200 °C Type K (1.0 + 0.036% E) μV from −200 °C to 200 °C (1.0 + 0.055% E) μV from above 200 °C to 400 °C (4.14 + 0.06% E) μV from above 400 °C to 1100 °C where E is the thermocouple voltage |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature measuring equipment | Digital temperature measuring systems | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of -Thermocouple Measuring Systems Type T 1.0 μV from −200 °C to 0 °C (1.0 + 0.01% E) μV from above 0 °C to 200 °C Type K (1.0 + 0.036% E) μV from −200 °C to 200 °C (1.0 + 0.055% E) μV from above 200 °C to 400 °C (4.14 + 0.06% E) μV from above 400 °C to 1100 °C where E is the thermocouple voltage Metallic resistance Measuring Systems Calibration of standard PRTs using temperature fixed points from 1 mK at −189.34 °C to 0.6 mK at −38.83 °C 0.6 mK from −38.83 °C to 419.53 °C from 0.6 mK at 419.53 °C to 1.6 mK at 660.32 °C from 1.6 mK at 660.32 °C to 8 mK at 961.78 °C Calibration of PRTs by comparison against standard PRTs 5 mK at −200 °C 5 mK at −80 °C 3 mK at 0 °C 3 mK at 250 °C 6 mK at 550 °C 8 mK at 960 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature measuring equipment | Digital temperature measuring systems; Surface probes | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of surface temperature sensorsWith Calibration and Measurement Capability of - from 0.21 °C at 20 °C to 0.90 °C at 350 °C Calibration of surface temperature measuring system (calibrators) With Calibration and Measurement Capability of - from 0.21 °C at 20 °C to 0.90 °C at 350 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature measuring equipment | Liquid in glass (LIG) thermometers | Temperature | Measurement against reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - 0.3 °C from −200 °C to −50 °C25 mK from above −50 °C to −10 °C 5 mK from above −10 °C to 50 °C 3 mK at 0 °C 15 mK from above 50 °C to 100 °C 40 mK from above 100 °C to 250 °C 0.15 °C from above 250 °C to 300 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature standards and reference equipment | Black body sources | Temperature | Direct measurement using a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - from 0.4 °C at −40 °C to 0.3 °C at 50 °Cfrom 0.3 °C at 50 °C to 1.4 °C at 1000 °C from 1.4 °C at 1000 °C to 2.9 °C at 2000 °C Calibration of ear thermometer blackbodies 0.025 °C in the range 25 °C to 45 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature standards and reference equipment | Dry block calibrators | Temperature | Direct measurement using a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - from 0.15 °C at −50 °C to 0.05 °C at 50 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature standards and reference equipment | Platinum resistance thermometers (PRTs) reference standards | Temperature | Direct measurement using a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of standard PRTs using temperature fixed points With Calibration and Measurement Capability of -
5 mK at −200 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature standards and reference equipment | Platinum-gold (Pt-Au) thermocouples; Rare metal thermocouples | mV per oC | Direct measurement using a reference standard | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability Calibration of Rare Metal Thermocouple With Calibration and Measurement Capability of -(0.31 + 0.014% E) μV from 0 °C to 1560 °C where E is the thermocouple voltage
19 mK from 0 °C to 50 °C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Temperature metrology - Temperature standards and reference equipment | Temperature reference standards - Fixed points | Aluminium (Al) freezing point; Argon (Ar) triple point; Cadmium freezing point; Gallium melting point; Indium (In) freezing point; Mercury (Hg) triple point; Silver (Ag) freezing point; Tin (Sn) freezing point; Water triple point; Zinc freezing point | Direct measurement against H2O triple point; Direct measurement against a reference standard; Direct measurement against argon triple point; Direct measurement against mercury triple point | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Capability With Calibration and Measurement Capability of - 1 mK at Ar triple point 0.5 mK at Sn freezing point |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
The only data displayed is that deemed relevant and necessary for the clear description of the Activities and Services covered by the Scope of Accreditation (SoA).
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
The Product column may also describe an Item or Material.
The Limitations column may also describe the Range.