National Association of Testing Authorities, Australia
Scope of Accreditation
Norbar Torque Tools (Australia) Pty Ltd
Site
Torque Measurement Laboratory
Accreditation No.
3800
Site No.
3793
Date of Accreditation
25 Mar 1993
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Torque - Torque measuring and testing equipment | Recorders and controllers | Torque | Comparison with a reference standard | Tests on indicators, recorders and calibrators by electrical simulation and measurement Norbar Brand Only by in-house methods TCS-LAB-I-002 and TCS-LAB-I-023 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the range 0.5 mV DC to 5 V DC 0.088% of reading for an input voltage of 1 mV DC 0.073% of reading for an input voltage from 2 mV to 16.5 mV DC 0.0014% of reading at 5 V DC (excitation voltage) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Torque multiplying gearboxes | Torque | Comparison with a reference standard | Self-indicating types and ratio confirmation | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the range 500 N.m to 6800 N.m | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Torque transducers | Torque | Gravimetric measurement with reference beam | Calibration of torque transducers and torque calibrators to BS 7882 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - in the range 0.1 N.m to 6800 N.m 0.015% of reading above 1 N.m to 50 N.m 0.02% of reading above 50 N.m to 1500 N.m 0.14% of reading above 1500 N.m to 6800 N.m | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Torque wrenches | Torque | Comparison with a reference standard | Calibration of torque wrenches /drivers to ISO 6789-2 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of - Torque wrenches and torque drivers |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.