SGS Australia Pty Ltd

Site

trading as SGS Radiation Services

Accreditation No.

2562

Site No.

18509

Date of Accreditation

16 Apr 2010

Address

10 / 585 Blackburn Road
Notting Hill, VIC 3168
Australia

sgs.com/en-au/service-groups/radiation-services

Contact

Mr Hasan Qasim P: +61 0400474572

Availability

Services available to external clients

Scope of Accreditation

trading as SGS Radiation Services

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Ionising radiation - Ionising radiation standards and equipment Contamination monitoring instruments Gamma activity Direct measurement against a reference standard In-house procedure CA 018

Capability

Calibration of contamination monitoring instruments using wide area reference sources with Calibration and Measurement Capability of -

7% at 8.1 Bq/cm2 for 238Pu (large area detectors only),

7% at 660 Bq/cm2 for 238Pu

7% at 570 Bq/cm2 for 241Am

7% at 510 Bq/cm2 for 14C

7% at 220 Bq/cm2 for 99Tc

7% at 250 Bq/cm2 for 36Cl

7% at 180 Bq/cm2 for 90Sr/90Y and

7% at 120 Bq/cm2 for 129I

Contamination monitoring instruments; Ionising radiation survey instruments Air kerma rate (mGy/h); Ambient dose equivalent (H*(10) rate (mSv/h)); Exposure rate (mR/h, C/kg/h) Direct measurement against a reference standard In-house procedure CA 001 and CA019

Calibration of ionising radiation survey instruments for exposure rate

(mR/h), air kerma rate (mGy/h) and ambient dose equivalent

(H*(10) rate (mSv/h)) and the corresponding integrated values

(exposure, air kerma and ambient dose equivalent) for survey

instruments with internal timers

Capability

Utilising 662 keV γ-rays in the air kerma rate range of

8 x 10-8 to 1.2 x 10-2 Gy/h using 137Cs sources of activity

with Calibration and Measurement Capability of -

33% using a 370 MBq source,

6% using a 3.7 GBq source,

5% using a 41 GBq source,

5% using a 145 GBq source

Dosimeters Personal dose equivalent and cumulative effective dose Direct measurement against a reference standard In-house procedure CA 010

Capability

Calibration of electronic personal dosemeters for personal dose equivalent Hp(10) (mSv)
 and quartz fibre electrometer dosemeters for exposure (mR)
 Utilising 662 keV γ-rays in the personal dose equivalent range of 2.5 x 10-5 to 1.6 x 10-4 Sv using 137Cs sources of activity 370 MBq, 3.7 GBq, 41 GBq and 145 GBq,
with Calibration and Measurement Capability of -
 8% for electronic dosimeters
 9% for quartz fibre electrometer dosimeters

ISO/IEC 17025 (2017)

Manufactured Goods

Service Product Determinant Technique Procedure Limitations
Ionising radiation - Dosimetry Radiation contamination monitoring instruments Response evaluation Not applicable In-house procedure CA 018
Ionising radiation - Source characterisation Isotopes and other sources of ionising radiation Radionuclide source identification 133Ba gamma ray point source; 137Cs gamma ray point source; 226Ra gamma ray point source; 232Th gamma ray point source; 241Am gamma ray point source; 60Co gamma ray point source In-house procedure CA 018

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.