Scope of Accreditation

Training, Testing and Certification Centre

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysers Gas analysers Concentration Comparison measurement against a certified reference material; Direct measurement by certified reference material

in house method LP0022

Carbon monoxide : NDIR; 0.0001-10%

Carbon monoxide : Electrochemical; 0.0001-1%

Carbon dioxide : NDIR; 0.001-100%

Methane : NDIR; 0.01-100%

Methane : Pellistor; 0.01-5%

Methane : Methanometer; 0.01-100%

Oxygen : Electrochemical; 0.01-25%

Oxygen : Paramagnetic; 0.01-25%

Hydrogen Sulphide : Electrochemical; 0.1-20000ppm

Sulfur dioxide : Electrochemical; 0.1-20000ppm

Sulfur dioxide : NDIR; 0.00001-50%

Sulfur dioxide : UV fluorescence; 0.1-50ppm

Nitric oxide : Chemiluminescent; 0.1-5000ppm

Nitrogen dioxide : Chemiluminescent; 0.1-5000ppm

Nitrogen dioxide : Electrochemical; 0.1-10000ppm

Oxides of nitrogen : Chemiluminescent; 0.1-10000ppm

Ammonia : Electrochemical; 0.0001-1%

Propane : Pellistor; 0.01-3%

Hexane/Methane/Propane: Pellistor; 0.01-100% LEL

Methane : FID; 0.01-100%

Methane/Hydrogen : Thermal conductivity; 0.01 - 100% LEL

Capability

with least uncertainty of measurement of 5.0 % relative

Flow metrology - Environmental flow measuring devices and systems Air sampling pumps and meters Flow rate Comparison with a reference meter; Volumetric transfer EP0143 - Calibration of Flow Meters (Flow Meter Calibration using the Bubble Flow meter as reference)

Capability

Calibration of gas flowmeters using atmospheric air as the test medium with a minimum test volume of 500 mL
With a Calibration and Measurement Capability of -
6% of reading for flow rates from 0.2 L/min to less than 0.3 L/min
4% of reading for flow rates from 0.3 L/min to 1.5 L/min
3% of reading for flow rates from above 1.5 L/min to 5 L/min
Flow rate Comparison with a reference meter EP0140

Capability


Calibration of air sampling pumps using air at atmospheric conditions
with Calibration and Measurement Capability of -
Sampling pump flow rate: 1.2% of flow from 1 L/min to 3 L/min
Sampling pump timer: (0.05% reading + 0.5 minutes) over a minimum of 2 hours

Temperature metrology - Verification of controlled enclosures Baths; Furnaces; Ovens Spatial uniformity; Temperature; Temporal uniformity Direct temperature measurement

AS 2853- 1986

From -40 ºC to 850 ºC
 
 

Capability

Including on site calibrations.

With Calibration and Measurement Capability of -
1.3 ºC  from -40 ºC to 300 ºC
5.4 ºC  above 300 ºC to 850 ºC

Environmental chambers - Humidity Fluctuation; Gradient; Humidity By temperature and humidity measurement AS 60068.3.6; IEC 60068.3.6

Capability

Including on site calibrations.

With Calibration and Measurement Capability of -
Humidity and Relative humidity gradient tests
3.0 %RH in the range 10% to 95%RH  and from 23 °C to 55 °C
5.0 %RH in the range 10% to 93% RH and above 55 °C to 80 °C

Environmental chambers - Temperature Fluctuation; Gradient; Temperature Direct temperature measurement

AS 60068.3.5, IEC 60068.3.5

From -40 ºC to 300 ºC
 
 

Capability

Including on site calibrations.

With Calibration and Measurement Capability of -
1.3 ºC from -40 ºC to 300 ºC