National Association of Testing Authorities, Australia
Scope of Accreditation
Zinfra Pty Ltd
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Flow metrology - Flow measuring devices and systems | Gas meters | Volume (and nominal flow rate) | Comparison with a reference meter; Direct measurement by critical flow venturi nozzle | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Calibration in the laboratory of domestic gas flow meters using sonic nozzles with air at atmospheric conditions with Calibration and Measurement Capability of- 0.50% of volume at 0.01 m3/h to 0.03 m3/h, with a minimum 2.4L test volume. (Qmin Rig5) 0.30% of volume from 0.016 m3/h to 0.099 m3/h, with a minimum 3.6 L test volume. (Rig 1) 0.25% of volume from 0.1 m3/h to 12.6 m3/h, with a minimum 3.6 L test volume. (Rig 1) Calibration in the laboratory of commercial and industrial gas flow meters using sonic nozzles with air at atmospheric conditions with Calibration and Measurement Capability of- 0.30% of volume from 0.016 m3/h to 0.099 m3/h using sonic nozzles, with a minimum 300 L test volume (Rig 2) 0.25% of volume from 0.1 m3/h to 48 m3/h using sonic nozzles, with a minimum 300 L test volume (Rig 2) 0.25% of volume from 0.1 m3/h to 300 m3/h using sonic nozzles, with a minimum 300 L test volume (Rig 2 and 3) 0.25% of volume from 0.5 m3/h to 6500 m3/h, using reference meters, with a minimum 4000 pulses of reference meter (Rig 4) |
ISO/IEC 17025 (2017)
Manufactured Goods
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Pattern approval and physical performance evaluation of metering and trade measurement equipment | Gas measuring systems | Pattern approval tests - Leakage | Integrity | Leak testing on domestic and industrial gas meters between 5 kPa and 300 kPa with a detection limit of 0.00012m3/h or 0.0000333 L/s by in-house method of pressure decay as per PROC.4.10.6 | |
Water meters | Endurance | Durability | By in-house method of flowing tap water at up to 90ºC and periodically verifying registrations of water volume and condition of meters as per PROC.4.10.7 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
20726
Site No.
22851
Print date
28 Nov 2024