National Measurement Institute (NMI)

Site

Impedance Standards Project

Accreditation No.

1

Site No.

14443

Date of Accreditation

08 Sep 2000

Address

36 Bradfield Road
Lindfield West, NSW 2070
Australia

measurement.gov.au

Contact

Ms Charmaine Kwan P: +61 (02) 84673836

Availability

Services conditionally available to external clients

Scope of Accreditation

Impedance Standards Project

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
DC and low frequency electrical metrology - Electrical standards Conductance boxes; Precision resistors; Resistance boxes Resistance Comparison with a primary realisation; Comparison with a reference standard

from 10 μΩ to 100 TΩ

Capability

With Calibration and Measurement Capability of -


Relative Expanded Uncertainty / (μΩ/Ω)Typical Current / VoltageStandard
10 μΩ1032 AStandard Resistors
>10 μΩ to <100 μΩ2010 A to 32 AStandard Resistors
100 μΩ
210 AStandard Resistors
>100 μΩ to <1 mΩ
23.2 A to 10 AStandard Resistors
1 mΩ1.53.2 AStandard Resistors
>1 mΩ to <10 mΩ
21 A to 3.2 AStandard Resistors
10 mΩ11 AStandard Resistors
>10 mΩ to <100 mΩ
1320 mA to 1 AStandard Resistors
100 mΩ0.3320 mAStandard Resistors
>100 mΩ to <1 Ω
1100 mA to 320 mAStandard Resistors
1 Ω0.15100 mAStandard Resistors
>1 Ω to <10 Ω
0.332 mA to 100 mAStandard Resistors
10 Ω
0.1532 mAStandard Resistors
>10 Ω to <100 Ω
0.310 mA to 32 mAStandard Resistors
100 Ω
0.1510 mAStandard Resistors
>100 Ω to <1 kΩ
0.33.2 mA to 10 mAStandard Resistors
1 kΩ
0.153.2 mAStandard Resistors
>1 kΩ to <10 kΩ
0.31 mA to 3.2 mAStandard Resistors
10 kΩ
0.151 mAStandard Resistors
>10 kΩ to 1 MΩ10.1 mA to 1 mAStandard Resistors
>1 MΩ to <10 MΩ
210 V to 100 VStandard Resistors
10 MΩ110 V to 100 VStandard Resistors
>10 MΩ to <100 MΩ
210 V to 100 V
Standard Resistors
10 MΩ to 100 MΩ
300>100 V to 1000 V
Standard Resistors
100 MΩ250 V to 100 VStandard Resistors
>100 MΩ to 1 GΩ
500>100 V to 1000 V
Standard Resistors
1 GΩ
5
100 V
Standard Resistors
>1 GΩ to 100 GΩ
1000100 V to 1000 V
Standard Resistors
>100 GΩ to <1 TΩ
3000100 V to 1000 V
Standard Resistors
1 TΩ
2000
100 V to 1000 V
Standard Resistors
>1 TΩ to 10 TΩ
3000
100 V to 1000 V
Standard Resistors
>10 TΩ to 100 TΩ
30000
100 V to 1000 V
Standard Resistors

Current shunts Resistance Comparison with a reference standard

10 μΩ to 1 mΩ

Capability

With Calibration and Measurement Capability of - 


Relative Expanded Uncertainty / (μΩ/Ω)CurrentStandard
10 μΩ to <100 μΩ200 to 20100 A to 2000 ADC Shunts
100 μΩ to 1 mΩ20 to 10100 A to 2000 ADC Shunts
Inductive voltage dividers Voltage ratio Comparison with a primary realisation; Comparison with a reference standard from 40 Hz to 63 Hz with input voltages from 100 V to 1000 V
from 400 Hz to 1 kHz with input voltages from 100 V to 1000 V
from 50 Hz to 5 kHz with input voltages from 5 V to 50 V

Capability

With Calibration and Measurement Capability of -

Expanded Uncertainty / (nV/V)

5 V to 50 V100 V to 1000 V
40 Hz to 50 Hz-1
> 50 Hz to 63 Hz101
> 63 Hz to 400 Hz1010
> 400 Hz to 1 kHz1050
> 1 kHz to 2 kHz10-
> 2 kHz to 5 kHz
100-

Inductors Inductance Comparison with a primary realisation; Comparison with a reference standard

from 10 μH to 11 H

Capability

With Calibration and Measurement Capability of -

Self inductors at 1000 Hz

5 nH from 10 μH to 1 mH

5 μH/H from 1 mH to 1.1 H

100 µH/H from 1.1 to 11 H

Precision capacitors Capacitance Comparison with a primary realisation; Comparison with a reference standard

from 1 pF to 1 μF

Capability

With Calibration and Measurement Capability of - 



Relative expanded uncertainty / (μF/F)


1000 Hz, 1592 Hz400 Hz to 5000 Hz
1 pF to 100 pFFused-silica standard capacitor0.08
100 pF to 10000 pFStandard capacitor, gas dielectric0.1
1 pF to 1000 pFStandard capacitor, gas dielectric

2
1 nF to 10 nFStandard capacitor

20
10 nF to 1μF
Standard capacitor

200

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.