AG&G SERVICES

Site

AG&G SERVICES

Accreditation No.

15401

Site No.

16140

Date of Accreditation

25 Jul 2024

Address

Unit 1, 383 Dorset Road
Bayswater, VIC 3153
Australia

aggservices.com.au

Contact

Mr Gregory Fox P: +61 0417562891

Availability

Services available to external clients

Scope of Accreditation

AG&G SERVICES

ISO/IEC 17025 (2017)

Infrastructure and Asset Integrity

Service Product Determinant Technique Procedure Limitations
Evaluation of air control equipment Biological safety cabinets Air barrier containment; Air pressure; Air velocity and uniformity of air velocity; Illuminance; Integrity of HEPA filter installations; Particle count in work zone; Recovery times; Sound level; Ultraviolet intensity; Work zone integrity Not applicable

AS 1807.1, AS 1807.3, AS 1807.5, AS 1807.8, AS 1807.10, AS 1807.13, AS 1807.15, AS 1807.16, AS 1807.20, AS 1807.21, AS 1807.22, AS 1807.23, AS 1807.24, AS/NZS ISO 14644.3 Annex B Sections 4.2.2, 4.2.3, 5, 6.4 (leak test method only), B.7, B.8, B.9, B.12 and 13.2.2

For determining compliance to the performance requirements of AS 2252.1, AS 2252.2
Clean rooms, pharmaceutical isolators and work stations Air barrier containment; Air pressure; Air velocity and uniformity of air velocity; Illuminance; Particle count in work zone; Recovery times; Sound level; Ultraviolet intensity; Work zone integrity Automatic particle counter; Face velocity; Smoke visualisation

AS 1807.1, AS 1807.3, AS 1807.5, AS 1807.6, AS 1807.7, AS 1807.8, AS 1807.10, AS 1807.13, AS 1807.15, AS 1807.16, AS 1807.20, AS 1807.21, AS 1807.22, AS 1807.23, AS 1807.24, AS/NZS ISO 14644.3 Annex B Sections 4.2.2, 4.2.3, 5, 6.4 (leak test method only), B.7, B.8, B.9, B.12 (excluding Annex 12.3.2) and 13.2.2 

for the purpose of determining compliance with the requirements of AS ISO 14644.1 Annex B, AS 2252.6
Cytotoxic drug safety cabinets Air barrier containment; Air change rate; Air pressure; Air velocity and uniformity of air velocity; Illuminance; Integrity of HEPA filter installations; Sound level; Ultraviolet intensity; Work zone integrity Automatic particle counter; Cold DOP; Face velocity; Smoke visualisation AS 1807.1, AS 1807.3, AS 1807.5, AS 1807.6, AS 1807.7, AS 1807.8, AS 1807.10, AS 1807.13, AS 1807.15, AS 1807.16, AS 1807.20, AS 1807.21, AS 1807.22, AS 1807.23 For determining compliance to the performance requirements of AS 2252.5
Fume cupboards and recirculating fume cabinets Air velocity and uniformity of air velocity

AS/NZS 2243.8 Appendices A and B, and AS/NZS 2243.9 Appendices A and B.

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.