National Association of Testing Authorities, Australia
Scope of Accreditation
JAF Australia Pty Ltd T/A Air Equipment Sales
Scope of Accreditation
Air Equipment Scientific Laboratory
ISO/IEC 17025 (2017)
Infrastructure and Asset Integrity
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Evaluation of air control equipment | Biological safety cabinets | Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1 | for compliance with: AS 2252.1 (2002), AS 2252.2 |
Air barrier containment | Cold DOP | AS 1807 clause 4.9 | for compliance with: AS 2252.1 (2002), AS 2252.2 | ||
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | for compliance with: AS 2252.1 (2002), AS 2252.2 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4 | for compliance with: AS 2252.1 (2002), AS 2252.2 | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | for compliance with: AS 2252.1 (2002), AS 2252.2 | ||
Sound level | Not applicable | AS 1807 clause 4.7 | for compliance with: AS 2252.1 (2002), AS 2252.2 | ||
Inward air velocity | Cold DOP | AS 1807 clause 4.8 | for compliance with: AS 2252.1 (2002), AS 2252.2 | ||
Clean rooms, pharmaceutical isolators and work stations | Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1, AS ISO 14644.3 clause 4.2.2, 4.2.3, Method B.2, B.3 | for compliance with: AS 2252.6, AS ISO 14644.1 | |
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Sound level | Not applicable | AS 1807.16 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Particle count in work zone | Automatic particle counter | AS ISO 14644.1 Appendix A, | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Work zone integrity | Cold DOP | AS 1807 clause 4.3 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Air pressure | Not applicable | AS ISO 14644.3 clause 4.2.1, Method B.1 | AS 2252.6, AS ISO 14644.1 | ||
Illuminance | Not applicable | AS 1807 clause 4.5 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4, AS 1807.7, AS ISO 14644.3 clause 4.2.7, Method B.7 | for compliance with: AS 2252.6, AS ISO 14644.1 | ||
Cytotoxic drug safety cabinets | Illuminance | Not applicable | AS 1807 clause 4.5 | for compliance with: AS 2252.5 | |
Ultraviolet intensity | Not applicable | AS 1807 clause 4.10 | for compliance with: AS 2252.5 | ||
Air barrier containment | Cold DOP | AS 1807 clause 4.9 | for compliance with: AS 2252.5 | ||
Inward air velocity | Cold DOP | AS 1807 clause 4.8 | for compliance with: AS 2252.5 | ||
Integrity of HEPA filter installations | Cold DOP | AS 1807 clause 4.4 | for compliance with: AS 2252.5 | ||
Sound level | Not applicable | AS 1807 clause 4.7 | for compliance with: AS 2252.5 | ||
Air velocity and uniformity of air velocity | Cold DOP | AS 1807 clause 4.1 | for compliance with: AS 2252.5 | ||
Fume cupboards and recirculating fume cabinets | Illuminance | Not applicable | AS 1807 clause 4.5 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | |
Extraction performance | Face velocity | AS/NZS 2243.8 Appendix B | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Sound level | Not applicable | AS 1807 clause 4.7 | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 | ||
Extraction performance | Smoke visualisation | AS/NZS 2243.8 Appendix A AS/NZS 2243.9 Appendix A | for compliance with: AS/NZS 2243.8, AS/NZS 2243.9 |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
14277
Site No.
14270
Print date
25 Nov 2024
END OF SCOPE