National Association of Testing Authorities, Australia
Scope of Accreditation
University of South Australia
Site
Australian Flow Management Group
Accreditation No.
17135
Site No.
18730
Date of Accreditation
29 Jun 2011
Address
Building L, Mawson Lakes Campus, Mawson Lakes Boulevard
Mawson Lakes, SA 5095
Australia
Contact
Availability
Services available to external clients
Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Flow metrology - Water flow measuring devices and systems | Liquid meters | Volume (and nominal flow rate) | Comparison with a reference meter; Gravimetric; Volumetric transfer | Error of measurement test on water meters with verification scales up to 40 L for compliance with NMI M10-2 clauses 6.3 and 9, NMI R49-2 clauses 7.4 and 10, AS 4747.2 clause 4.2 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability with Calibration and Measurement Capability of -0.27% from 0.00083337 L/s to 0.0222 L/s with a minimum volume of 4 L 0.28% from 0.0222 L/s to 0.083 L/s with a minimum volume of 50 L 0.20% from 0.083 L/s to 0.694 L/s with a minimum volume of 210 L 0.08% from 0.694 L/s to 5 L/s with a minimum volume of 380 L 0.08% from 5 L/s to 10 L/s with a minimum volume of 10 900 L 0.1% from 10 L/s to 20 L/s with a minimum volume of 14 000 L 0.11% from 20 L/s to 100 L/s with a minimum volume of 15 400 L 0.11% from 100 L/s to 250 L/s with a minimum volume of 16 600 L 0.14% from 250 L/s to 370 L/s with a minimum volume of 16 800 L 0.17% from 370 L/s to 520 L/s with a minimum volume of 18 500 L |
ISO/IEC 17025 (2017)
Manufactured Goods
Service | Product | Determinant | Technique | Procedure | Limitations |
---|---|---|---|---|---|
Electromagnetic compatibility (EMC) - Immunity - Commercial | Water meters | Magnetic effects | Not applicable | NMI R 49-2 clause 7.12 and 8.16 | |
Pattern approval and physical performance evaluation of metering and trade measurement equipment | Water meters | Water temperature | Not applicable | NMI R 49-2 clauses 7.5, 7.5a and 7.6 | |
Cold resistance | Not applicable | IEC 60068-2-1, NMI M 10-2 clause 7.3, NMI R 49-2 clause 8.3 | |||
Durability | Not applicable | NMI R 49-2 7.11 | |||
Water resistance - Immersion only | Not applicable | IEC 60068-2-18 , IEC 60512-14-7, IEC 60529, NMI M 10-2 clause 7.12 | |||
Absence of flow effects | Not applicable | NMI M 10-2 clause 6.6, NMI R 49-2 clause 7.8 | |||
Flow disturbance | Not applicable | NMI M 10-2 clause 6.8, NMI R 49-2 clause 7.10 | |||
Dry heat | Not applicable | IEC 60068-2-2, NMI M 10-2 Section 7.2, NMI R 49-2 Section 8.2, | |||
Impact resistance | Not applicable | IEC 60068-2-31, IEC 60068-2-47, NMI M 10-2 clause 7.7, NMI R 49-2 clause 8.7 | |||
Dust | Not applicable | IEC 60068-2-68, IEC 60512-11-8, IEC 60529, IEC 60721-2-5, NMI M 10-2 clause 7.13 | |||
Endurance | Not applicable | NMI M 10-2 clause 6.9 | |||
Water resistance - Immersion only | Not applicable | IEC 60068-2-18, IEC 60512-14-7, IEC 60529, NMI M 10-2 Sections 7.12 | |||
Water quality disturbance effects | Not applicable | NMI M 10-2 clause 6.10 | |||
Damp heat | Not applicable | IEC 60068-2-30, NMI M 10-2 clause 7.4, NMI R 49-2 clause 8.4 NMI R 49-2 clause 8.7. | |||
Hydrostatic pressure | Not applicable | NMI R 49-2 clause 7.7 | |||
Pattern approval tests | Not applicable | AS 3565.1 and .4 Non-urban water meters - NMI M10-2 clauses 5.3, 6.2, 6.3, 6.4, 6.6, 6.7, 6.8, 6.9, 6.12, 6.13, 6.14, 7.2, 7.3, 7.4, 7.5, 7.7, 7.8, 7.9, 7.10, 7.11, 7.12 and 7.13 Meters Intended for the Metering of Water in Full Flowing Pipes; Urban water meters to NMI R49-2 clauses 6.4, 7.3, 7.4, 7.8, 7.9, 7.10, 8.2, 8.3, 8.4, 8.5, 8.7 and 8.17 Water Meters Intended for the Metering of Cold Potable Water and Hot Water; | |||
Products and appliances connected to the water supply - Physical evaluation | Valves; Water service products - Connectors, filters, fittings, hoses and pipes | Head loss | Not applicable | In-house test method TM039, NMI M10-2 clause 6.7, NMI R49-2 clause 7.9 | up to 510L/s |
Hydrostatic pressure | Not applicable | AS 3565.1 clauses 2.11 NMI M10-2 clause 6.2, NMI R49-2 clause 7.3, AS 2638.2 clause 5.1.3 | up to 10MPa |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.
Accreditation No.
17135
Site No.
18730
Print date
22 Mar 2025