Scope of Accreditation
ISO/IEC 17025 (2017)
Calibration
- The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service | Product | Determinant | Technique | Procedure | Limitations | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensional metrology - Engineering equipment and precision instruments | Centrifuges | Rotational speed | Comparison with a reference standard by differential measurement | In-house method CY 20821 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability including on-site calibration With Calibration and measurement capability of - 1.2 rpm from 10 to 900 rpm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devices | Mass standards | Mass | Gravimetric measurement against reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 3 µg from 1mg to 20 mg4 µg from above 20 mg up to 50 mg 5 µg from above 50 mg up to 500 mg 6 µg from above 500 mg up to 1 g 7 µg from above 1 g up to 2 g 10 µg from above 2 g up to 5 g 25 µg from above 5 g up to 10 g 30 µg above 10 g up to 20 g 40 µg above 20 g up to 50 g 60 µg from above 50 g up to 100 g 0.11 mg from above 100 g up to 200 g 0.3 mg from above 200 g up to 500 g 0.6 mg from above 500 g up to 1 kg 2.5 mg from above 1 kg up to 2 kg 20 mg from above 2 kg up to 10 kg 0.2 g from above 10 kg up to 20 kg 0.02 kg from above 20 kg up to 250 kg 0.05 kg from above 250 kg up to 500 kg 0.1 kg from above 500 kg up to 1000 kg | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mass - Determination of mass and calibration of weighing devicesincluding on site calibrations | Industrial weighing devices | Mass | Gravimetric measurement against reference mass | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and measurement capability of- 1 in 104 or 1 mg (whichever is greater) up to 3000 kg | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Laboratory weighing devices; Precision laboratory balances | Mass | Gravimetric measurement against reference mass | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability Including on site calibrations With Calibration and measurement capability of- 1 in 106 or 4 ug (whichever is greater) up to 0.5 g | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Ancillary temperature measuring equipment | Multi-channel thermocouple data recorders; Process calibrators | Volt to temperature conversion | Direct measurement by electrical input | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 0.2 °C from -100 °C to 500 °C 0.3 °C from 500 °C to 1100 °C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Temperature measuring equipment | Digital temperature measuring systems | Temperature | Measurement against reference standard | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of- 0.4 °C from -80 °C to 250 °C 0.6 °C from 250 °C to 500 °C 2.5 °C from 500 °C to 1100 °C | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature metrology - Verification of controlled enclosures | Autoclaves and sterilising ovens; Baths; Freezers; Furnaces; Incubators; Ovens | Gradient; Spatial uniformity | Direct temperature measurement | AS 2853 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of - 0.9 °C in the range from 0 °C to 250 °C for Ovens and Furnaces 2.6 °C in the range from 250 °C to 1100 °C for Ovens and Furnaces 0.9 °C in the range from -40 °C to 100 °C for Incubators 0.8 °C in the range from 100 °C to 130 °C for Autoclaves and sterilising ovens 1 °C in the range from 131 °C to 160 °C for Autoclaves and sterilising ovens 1 °C in the range from -80 °C to -16 °C for Industrial freezers 0.9 °C in the range from -15 °C to ambient for Industrial freezers 0.8 °C in the range from 0 °C to 130 °C for Baths | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Torque - Torque measuring and testing equipment | Torque wrenches | Torque | Comparison with a reference standard | ISO 6789.2 and in house method 5.23 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of -1.2% from 1.25 Nm to 1500 Nm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Volume and density - Determination of density | Alcohol meters; Apparent density (draft survey) hydrometers; Bark hydrometer; Baume hydrometers; Brix hydrometers; Density hydrometers; Soil (colloidal solution) hydrometers; Specific gravity hydrometers | Baume; Brix; Percentage (%) alcohol by volume; Specific gravity; g/L | Gravimetric | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and measurement capability of -1/3 of the smallest scale division in the range 0 to 2.0 g/ml for Density Hydrometers; 1/3 of the smallest scale division in the range 0.990 to 1.040 kg/L for Apparent Density Hydrometers; 1/3 of the smallest scale division in the range -5 to +60 g/l for Soil Hydrometers; 1/3 of the smallest scale division in the range 0 to 90º Bx for Brix Hydrometers; 1/3 of the smallest scale division in the range 0 to 80º Bk for Bark Hydrometers; 1/3 of the smallest scale division in the range 0 to 90º Be for Baume Hydrometers; 1/3 of the smallest scale division in the range 0.6 to 2.0 for Specific Gravity Hydrometers; 1/3 of the smallest scale division in the range 0 to 100% for Alcohol Hydrometers | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Volume and density - Laboratory volumetric glassware and measures | Piston operated volumetric apparatus (POVA) | Volume | Gravimetric | by the method of ISO 8655 and in house method 5.19 | from 1 μL to 10000 μL | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Capability With Calibration and Measurement Capability of -0.034uL from 1uL to 10uL 0.06uL above 10uL to 20uL 0.07uL above 20uL to 50uL 0.13uL above 50uL to 100uL 0.15uL above 100uL to 200uL 0.43uL above 200uL to 500uL 1.08uL above 500uL to 1000uL 3.68uL above 1000uL to 2500uL 6.53uL above 2500uL to 5000uL 10.17uL above 5000uL to 10000uL |
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.