Cal Pacific Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysersincluding on site calibrations Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material In-house method CPPRO024 CH4 (catalytic combustion detector): 0-5% v/v
CH4 (thermal conductivity detector): 0-100% v/v
CH4 (nfra-red detector): 0-100% v/v
C3H8 (catalytic combustion detector): 0-100% LEL
CO2 (infra-red detector): 0-100%
CO (electrochemical detector): 0-2000ppm
CO (fuel cell detector): 0-100ppm
CO (infra-red detector): 0-20%
H2S (electrochemical detector): 0-1000ppm
NH3 (electrochemical detector): 0-5000ppm 
NO2 (chemiluminescent detector): 0-500ppm 
NO2 (electrochemical detector): 0-1000ppm
NO2 (ultraviolet detector): 0-500ppm
NO (chemiluminescent detector): 0-10000ppm 
NO (electrochemical detector): 0-5000ppm
NO (infrared detector): 0-100% 
NOX (oxides of nitrogen) (chemiluminescent detector): 0-10000ppm 
N2O (electrochemical detector): 0-5000ppm 
N2O (infrared detector): 0-60% 
O2 (electrochemical detector): 0-100%
O2 (polarographic cell detector): 0-100%
O2 (paramagnetic detector): 0-100%
O2 (fuel cell detector): 0-25%
O2 (magneto dynamic detector): 0-100%
SO2 (electrochemical detector): 0-150ppm 
SO2 (infrared detector): 0-80%

Capability

With Calibration and Measurement Capability of-

CH4 (catalytic combustion detector) - 2% of the measurement value in the range of 0-5% v/v
CH4 (thermal conductivity detector) - 2% of the measurement value in the range of 0-100% v/v
CH4 (nfra-red detector) - 2% of the measurement value in the range of 0-100% v/v
C3H8 (catalytic combustion detector) - 2% of the measurement value in the range of 0-100% LEL
CO2 (infra-red detector) - 2% of the measurement value in the range of 0-100%
CO (electrochemical detector) - 2% of the measurement value in the range of 0-2000ppm
CO (fuel cell detector) - 2% of the measurement value in the range of 0-100ppm
CO (infra-red detector) - 2% of the measurement value in the range of 0-20%
H2S (electrochemical detector) - 3% of the measurement value in the range of 0-1000ppm
NH3 (electrochemical detector) - 20.25% of the measurement value in the range of 0-5000ppm
NO2 (chemiluminescent detector) - 3.75% of the measurement value in the range of 0-500ppm
NO2 (electrochemical detector) - 5.10% of the measurement value in the range of 0-1000ppm
NO2 (ultraviolet detector) - 3.75% of the measurement value in the range of 0-500ppm
NO (chemiluminescent detector) - 3.55% of the measurement value in the range of 0-10000ppm
NO (electrochemical detector) - 4.95% of the measurement value in the range of 0-5000ppm
NO (infrared detector) - 3.55% of the measurement value in the range of 0-100%
NOX (oxides of nitrogen) (chemiluminescent detector) - 3.55% of the measurement value in the range of 0-10000ppm
N2O (electrochemical detector) - 5.10% of the measurement value in the range of 0-5000ppm
N2O (infrared detector) - 3.75% of the measurement value in the range of 0-60%
O2 (electrochemical detector) - 1% of the measurement value in the range of 0-100%
O2 (polarographic cell detector) - 1% of the measurement value in the range of 0-100%
O2 (paramagnetic detector) - 1% of the measurement value in the range of 0-100%
O2 (fuel cell detector) - 1% of the measurement value in the range of 0-25%
O2 (magneto dynamic detector) - 2% of the measurement value in the range of 0-100%
SO2 (electrochemical detector) - 10.35% of the measurement value in the range of 0-150ppm
SO2 (infrared detector) - 3.35% of the measurement value in the range of 0-80%

ISO/IEC 17025 (2017)

Environment

Service Product Determinant Technique Procedure
Analysis for engine emissions Diesel exhaust emissions Carbon monoxide; Nitrogen - Total oxidised; Nitrogen dioxide; Nitrogen oxide Chemiluminescence gas analyser; Nondispersive infrared (NDIR)

Mobile facility using in house method CPPRO027

Carbon dioxide (CO2); Carbon monoxide; Nitrogen - Total oxidised; Nitrogen dioxide; Nitrogen oxide Chemiluminescence gas analyser; Nondispersive infrared (NDIR) Mobile facility, in accordance with MDG29 Guideline for the management of diesel engine pollutants in under ground environment, New South Wales Department of Primary Industries using in-house method CPPRO027

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.