Cal Pacific Pty Ltd

Site

Cal Pacific Pty Ltd

Accreditation No.

16644

Site No.

17966

Date of Accreditation

28 Jan 2009

Address

Unit 19/13 Hartley Drive
Thornton, NSW 2322
Australia

calpacific.com.au

Contact

Mr Paul Gribbin P: +61 0407960533

Availability

Services available to external clients

Scope of Accreditation

Cal Pacific Pty Ltd

ISO/IEC 17025 (2017)

Calibration

  • The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Service Product Determinant Technique Procedure Limitations
Chemical metrology - Gas analysersincluding on site calibrations Gas analysers; Mine safety and confined space equipment Concentration Direct measurement by certified reference material In-house method 5-04-2  CH4 (catalytic combustion detector): 0-5% v/v
CH4 (thermal conductivity detector): 0-100% v/v
CH4 (nfra-red detector): 0-100% v/v
C3H8 (catalytic combustion detector): 0-100% LEL
CO2 (infra-red detector): 0-100%
CO (electrochemical detector): 0-2000ppm
CO (fuel cell detector): 0-100ppm
CO (infra-red detector): 0-20%
H2S (electrochemical detector): 0-1000ppm
NH3 (electrochemical detector): 0-5000ppm 
NO2 (chemiluminescent detector): 0-500ppm 
NO2 (electrochemical detector): 0-1000ppm
NO2 (ultraviolet detector): 0-500ppm
NO (chemiluminescent detector): 0-10000ppm 
NO (electrochemical detector): 0-5000ppm
NO (infrared detector): 0-100% 
NOX (oxides of nitrogen) (chemiluminescent detector): 0-10000ppm 
N2O (electrochemical detector): 0-5000ppm 
N2O (infrared detector): 0-60% 
O2 (electrochemical detector): 0-100%
O2 (polarographic cell detector): 0-100%
O2 (paramagnetic detector): 0-100%
O2 (fuel cell detector): 0-25%
O2 (magneto dynamic detector): 0-100%
SO2 (electrochemical detector): 0-150ppm 
SO2 (infrared detector): 0-80%

Capability

With Calibration and Measurement Capability of-

CH4 (catalytic combustion detector) - 2% of the measurement value in the range of 0-5% v/v
CH4 (thermal conductivity detector) - 2% of the measurement value in the range of 0-100% v/v
CH4 (nfra-red detector) - 2% of the measurement value in the range of 0-100% v/v
C3H8 (catalytic combustion detector) - 2% of the measurement value in the range of 0-100% LEL
CO2 (infra-red detector) - 2% of the measurement value in the range of 0-100%
CO (electrochemical detector) - 2% of the measurement value in the range of 0-2000ppm
CO (fuel cell detector) - 2% of the measurement value in the range of 0-100ppm
CO (infra-red detector) - 2% of the measurement value in the range of 0-20%
H2S (electrochemical detector) - 3% of the measurement value in the range of 0-1000ppm
NH3 (electrochemical detector) - 20.25% of the measurement value in the range of 0-5000ppm
NO2 (chemiluminescent detector) - 3.75% of the measurement value in the range of 0-500ppm
NO2 (electrochemical detector) - 5.10% of the measurement value in the range of 0-1000ppm
NO2 (ultraviolet detector) - 3.75% of the measurement value in the range of 0-500ppm
NO (chemiluminescent detector) - 3.55% of the measurement value in the range of 0-10000ppm
NO (electrochemical detector) - 4.95% of the measurement value in the range of 0-5000ppm
NO (infrared detector) - 3.55% of the measurement value in the range of 0-100%
NOX (oxides of nitrogen) (chemiluminescent detector) - 3.55% of the measurement value in the range of 0-10000ppm
N2O (electrochemical detector) - 5.10% of the measurement value in the range of 0-5000ppm
N2O (infrared detector) - 3.75% of the measurement value in the range of 0-60%
O2 (electrochemical detector) - 1% of the measurement value in the range of 0-100%
O2 (polarographic cell detector) - 1% of the measurement value in the range of 0-100%
O2 (paramagnetic detector) - 1% of the measurement value in the range of 0-100%
O2 (fuel cell detector) - 1% of the measurement value in the range of 0-25%
O2 (magneto dynamic detector) - 2% of the measurement value in the range of 0-100%
SO2 (electrochemical detector) - 10.35% of the measurement value in the range of 0-150ppm
SO2 (infrared detector) - 3.35% of the measurement value in the range of 0-80%

ISO/IEC 17025 (2017)

Environment

Service Product Determinant Technique Procedure
Analysis for engine emissions Diesel exhaust emissions Carbon monoxide; Nitrogen - Total oxidised; Nitrogen dioxide; Nitrogen oxide Chemiluminescence gas analyser; Nondispersive infrared (NDIR)

Mobile facility using in house QSP 5-04-4

Carbon dioxide (CO2); Carbon monoxide; Nitrogen - Total oxidised; Nitrogen dioxide; Nitrogen oxide Chemiluminescence gas analyser; Nondispersive infrared (NDIR) Mobile facility, in accordance with MDG29 Guideline for the management of diesel engine pollutants in under ground environment,New South Wales Department of Primary Industries using in-house QSP 5-04-4

The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.